Objective: To report the incidence of intraoperative complications of coiling of ruptured cerebral aneurysms leading to permanent disability or death in a consecutive series of 341 patients and to identify risk factors for these events. Methods: Between January 2007 and May 2011, 341 consecutive patients with ruptured intracranial aneurysms were treated with detachable coils. Procedural complications of coils embolization leading to death or neurologic disability at the time of hospital discharge were recorded. For patients with procedural complications, odds ratios (OR) with corresponding 95% confidence intervals (CI) were calculated for the following patient and aneurysm characteristics: patient age and sex, use of a supporting balloon, aneurysm location, timing of treatment, clinical condition at the time of treatment, and aneurysm size. Results: Procedural complications occurred in 21 of 341 patients (6.16%; 95% CI, 4.2% to 8.1%), leading to death in 9 patients (procedural mortality, 2.6%; 95% CI, 1.6% to 4.2%) and to disability in 12 patients (procedural morbidity, 3.5%; 95% CI, 2.0% to 5.3%). There were 4 procedural ruptures and 17 thrombotic complications. The significant risk factors for the occurrence of procedural complications were use of supporting balloon (OR, 5.3; 95% CI, 2.3 to 16.1%) and stent (OR, 8.4; 95% CI, 4.1 to 20.5%). Conclusion: Procedural complication rate of coiling of ruptured aneurysms leading to disability or death is 6.16%. In this series, the use of a temporary supporting balloon and stent in the treatment of wide-necked aneurysms was the risk factors for the occurrence of complications. [Analysis of intraoperative complication of ruptured cerebral aneurysm with detachable coils embolization. Life Science Journal. 2011; 8(3):543-546] (ISSN: 1097-8135). http://www.lifesciencesite.com.
引用
收藏
页码:543 / 546
页数:4
相关论文
共 50 条
  • [41] Embolization of a large-neck aneurysm of the internal iliac artery with interlocking detachable coils: case report
    Soulez, G
    Oliva, VL
    Therasse, E
    Provencher, D
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 1998, 49 (01): : 39 - 41
  • [42] Splenic Artery Aneurysm of the Anomalous Splenomesenteric Trunk: Successful Treatment by Transcatheter Embolization Using Detachable Coils
    Motohiro Sato
    Izumi Anno
    Masayuki Yamaguchi
    Hiroyuki Iida
    Kazuo Orii
    CardioVascular and Interventional Radiology, 2006, 29 : 432 - 434
  • [43] Peripheral use of detachable coils: expanding the boundaries of embolization
    Orgera, Gianluigi
    Tipaldi, Marcello A.
    Zaman, Iftikhar
    Ali, Tariq
    Rebonato, Alberto
    Hatzidakis, Adam
    Rossi, Michele
    Krokidis, Miltiadis E.
    MINERVA CARDIOANGIOLOGICA, 2017, 65 (01): : 52 - 60
  • [44] Splenic artery aneurysm of the anomalous splenomesenteric trunk: Successful treatment by transcatheter embolization using detachable coils
    Sato, M
    Anno, I
    Yamaguchi, M
    Iida, H
    Orii, K
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (03) : 432 - 434
  • [45] Aneurysmal rupture during embolization with Guglielmi detachable coils
    Peltier, J
    Nowtash, A
    Toussaint, P
    Desenclos, C
    Deramond, H
    Le Gars, D
    NEUROCHIRURGIE, 2004, 50 (04) : 454 - 460
  • [46] Endovascular treatment of ruptured, peripheral cerebral aneurysms: Parent artery occlusion with short Guglielmi detachable coils
    Cloft, HJ
    Kallmes, DF
    Jensen, ME
    Lanzino, G
    Dion, JE
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1999, 20 (02) : 308 - 310
  • [47] Predictors of angiographic changes in neck remnants of ruptured cerebral aneurysms treated with Guglielmi detachable coils - Commentary
    Awad, IA
    NEUROLOGIA MEDICO-CHIRURGICA, 2006, 46 (01) : 9 - 10
  • [48] Radio aneurysm coils for noninvasive detection of cerebral embolization failures: A preliminary study
    Mohammadi, Abdolreza Rashidi
    Chen, Keqin
    Ali, Mohamed Sultan Mohamed
    Takahata, Kenichi
    BIOSENSORS & BIOELECTRONICS, 2011, 30 (01): : 300 - 305
  • [49] New detachable coils for treating cerebral aneurysms
    Monsein, LH
    NATURE MEDICINE, 1996, 2 (02) : 160 - 161
  • [50] Late rebleeding of ruptured intracranial aneurysms treated with detachable coils
    Sluzewski, M
    van Rooij, WJ
    Beute, GN
    Nijssen, PC
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2005, 26 (10) : 2542 - 2549