Bone Marrow Metastases: T2-weighted Dixon Spin-Echo Fat Images Can Replace T1-weighted Spin-Echo Images

被引:80
|
作者
Maeder, Yael [1 ]
Dunet, Vincent [1 ]
Richard, Raphael [1 ]
Becce, Fabio [1 ]
Omoumi, Patrick [1 ]
机构
[1] Lausanne Univ Hosp, Dept Diagnost & Intervent Radiol, Rue Bugnon 46, CH-1011 Lausanne, Switzerland
关键词
WHOLE-BODY MRI; PROSTATE-CANCER; AXIAL SKELETON; IN-PHASE; PET-CT; SUPPRESSION; MALIGNANCY; DIAGNOSIS; SEQUENCE; SYSTEM;
D O I
10.1148/radiol.2017170325
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To test the potential of Dixon T2-weighted fat-only sequences to replace T1-weighted sequences for the detection of bone metastases, with the hypothesis that diagnostic performance with an alternative magnetic resonance (MR) imaging protocol (sagittal spin-echo Dixon T2-weighted fat-only and water-only imaging) would not be inferior to that with the standard protocol (sagittal spin-echo T1-weighted and spin-echo Dixon T2-weighted water-only imaging). Materials and Methods: A total of 121 consecutive whole-spine MR imaging examinations (63 men; mean age 6 standard deviation, 61.4 years 6 +/- 11.8) performed for suspected vertebral bone metastases were included in this retrospective, institutional review board-approved study. Quantitative image analysis was performed for 30 randomly selected spine levels. Qualitative analysis was performed separately by two musculoskeletal radiologists, who registered the number of metastases for each spine level. Areas under the curve with the protocols were compared on the basis of nonparametric receiver operating characteristic curve estimations by using a noninferiority test on paired data, with a best valuable comparator as a reference. Interobserver and interprotocol agreement was assessed by using k statistics. Results: Contrast-to-noise ratio was significantly higher on the alternative protocol images than on the standard protocol images (181.1 [95% confidence interval: 140.4, 221.7] vs 84.7 [95% confidence interval: 66.3, 103.1] respectively; P < .001). Diagnostic performance was not significantly inferior with the alternative protocol than with the standard protocol for both readers in a per-patient analysis (sensitivity, 97.9%-98.9% vs 93.6%-97.9%; specificity, 85.2%-92.6% vs 92.6%-96.3%; area under the curve, 0.92-0.96 vs 0.95, respectively; all P <= .02) and a per-spine level analysis (all P < .01). Interobserver and interprotocol agreement was good to very good (k = 0.70-0.81). Conclusion: Dixon T2-weighted fat-only and water-only imaging provide, in one sequence, diagnostic performance similar to that of the standard combination of morphologic sequences for the detection of probable spinal bone metastases, thereby providing an opportunity to reduce imaging time by eliminating the need to perform T1 sequences. (C) RSNA, 2017
引用
收藏
页码:853 / 864
页数:12
相关论文
共 50 条
  • [31] T2-weighted fast spin-echo MR imaging of liver.
    Mirowitz, SA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 171 (01) : 263 - 264
  • [32] ARE SAGITTAL T2-WEIGHTED FAST SPIN-ECHO MR-IMAGES SUFFICIENT FOR ROUTINE SPINE IMAGING
    MIAUX, Y
    FELDMAN, L
    DUCERVEAU, MN
    MARTINDUVERNEUIL, N
    COGNARD, C
    SAVIN, D
    CHIRAS, J
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1995, 16 (07) : 1558 - 1559
  • [33] THE PELVIS - T2-WEIGHTED FAST SPIN-ECHO MR-IMAGING
    NGHIEM, HV
    HERFKENS, RJ
    FRANCIS, IR
    SOMMER, FG
    JEFFREY, RB
    LI, KCP
    STEINER, RM
    GLOVER, GH
    RADIOLOGY, 1992, 185 (01) : 213 - 217
  • [34] Contrast enhancement of intracranial lesions: conventional T1-weighted spin-echo versus fast spin-echo MR imaging techniques
    Sugahara, T
    Korogi, Y
    Ge, YL
    Shigematsu, Y
    Liang, LX
    Yoshizumi, K
    Kitajima, M
    Takahashi, M
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1999, 20 (08) : 1554 - 1559
  • [35] Reduced metallic artefacts in 3 T knee MRI using fast spin-echo multi-point Dixon compared to fast spin-echo T2-weighted sequences
    Park, H. J.
    Lee, S. Y.
    Choi, S. H.
    Hong, H. P.
    Choi, Y. J.
    Kim, M. S.
    CLINICAL RADIOLOGY, 2017, 72 (11) : 996.e1 - 996.e6
  • [36] Are T1- and T2-weighted spin-echo MR images sufficient for the diagnosis of central nervous system vasculitis? Reply
    Wasserman, BA
    Pomper, MG
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (01) : 273 - 274
  • [37] Comparison of T1-weighted fast spin-echo and T1-weighted fluid-attenuated inversion recovery images of the lumbar spine at 3.0 Tesla
    Lavdas, Eleftherios
    Vlychou, Marianna
    Arikidis, Nikos
    Kapsalaki, Eftychia
    Roka, Violetta
    Fezoulidis, Ioannis V.
    ACTA RADIOLOGICA, 2010, 51 (03) : 290 - 295
  • [38] Whole-body MRI in children and adolescents: Can T2-weighted Dixon fat-only images replace standard T1-weighted images in the assessment of bone marrow?
    de Horatio, Laura Tanturri
    Zadig, Pia K.
    von Brandis, Elisabeth
    Rosendahl, Karen
    Avenarius, Derk F. M.
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 166
  • [39] UNSUPPRESSED FAT IN THE RIGHT ANTERIOR DIAPHRAGMATIC REGION ON FAT-SUPPRESSED T2-WEIGHTED FAST SPIN-ECHO MR-IMAGES
    YOSHIMITSU, K
    VARMA, DGK
    JACKSON, EF
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (02): : 145 - 149
  • [40] Fat-suppressed fast spin-echo mid-TE (TE[effective]=34) MR images: comparison with fast spin-echo T2-weighted images for the diagnosis of tears and anatomic variants of the glenoid labrum
    Tuite, MJ
    Shinners, TJ
    Hollister, MC
    Orwin, JF
    SKELETAL RADIOLOGY, 1999, 28 (12) : 685 - 690