A new concept of pencil beam dose calculation for 40-200 keV photons using analytical dose kernels

被引:20
|
作者
Bartzsch, Stefan [1 ,2 ]
Oelfke, Uwe [1 ,2 ]
机构
[1] Deutsch Krebsforschungszentrum, D-69120 Heidelberg, Germany
[2] Inst Canc Res, Sutton SM2 5NG, Surrey, England
关键词
dose calculation; point dose kernels; microbeam radiation therapy; MONTE-CARLO SIMULATIONS; MICROBEAM RADIATION-THERAPY; X-RAYS; DISTRIBUTIONS; TRANSPORT; DENSITY; ARRAYS; TISSUE;
D O I
10.1118/1.4824150
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The advent of widespread kV-cone beam computer tomography in image guided radiation therapy and special therapeutic application of keV photons, e.g., in microbeam radiation therapy (MRT) require accurate and fast dose calculations for photon beams with energies between 40 and 200 keV. Multiple photon scattering originating from Compton scattering and the strong dependence of the photoelectric cross section on the atomic number of the interacting tissue render these dose calculations by far more challenging than the ones established for corresponding MeV beams. That is why so far developed analytical models of kV photon dose calculations fail to provide the required accuracy and one has to rely on time consuming Monte Carlo simulation techniques. Methods: In this paper, the authors introduce a novel analytical approach for kV photon dose calculations with an accuracy that is almost comparable to the one of Monte Carlo simulations. First, analytical point dose and pencil beam kernels are derived for homogeneous media and compared to Monte Carlo simulations performed with the Geant4 toolkit. The dose contributions are systematically separated into contributions from the relevant orders of multiple photon scattering. Moreover, approximate scaling laws for the extension of the algorithm to inhomogeneous media are derived. Results: The comparison of the analytically derived dose kernels in water showed an excellent agreement with the Monte Carlo method. Calculated values deviate less than 5% from Monte Carlo derived dose values, for doses above 1% of the maximum dose. The analytical structure of the kernels allows adaption to arbitrary materials and photon spectra in the given energy range of 40-200 keV. Conclusions: The presented analytical methods can be employed in a fast treatment planning system for MRT. In convolution based algorithms dose calculation times can be reduced to a few minutes. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:15
相关论文
共 40 条
  • [31] Very high-energy electron dose calculation using the Fermi-Eyges theory of multiple scattering and a simplified pencil beam model
    Ronga, Maria Grazia
    Deut, Umberto
    Bonfrate, Anthony
    De Marzi, Ludovic
    MEDICAL PHYSICS, 2023, 50 (12) : 8009 - 8022
  • [32] Calculation of Forward Scatter Dose Distribution at the skin entrance from the patient table for fluoroscopically guided interventions using a pencil beam convolution kernel
    Vijayan, Sarath
    Xiong, Zhenyu
    Guo, Chao
    Troville, Jonathan
    Islam, Naveed
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [33] 4D dose calculation for pencil beam scanning proton therapy of pancreatic cancer using repeated 4DMRI datasets
    Dolde, Kai
    Naumann, Patrick
    David, Christian
    Gnirs, Regula
    Kachelriess, Marc
    Lomax, Antony John
    Saito, Nami
    Weber, Damien Charles
    Pfaffenberger, Asja
    Zhang, Ye
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (16):
  • [34] A new convolution/superposition dose calculation method for external photon beam radiation therapy using beam modifiers and/or independent collimators
    Liu, HH
    MEDICAL PHYSICS, 1997, 24 (10) : 1647 - 1647
  • [35] Streamlined Four-Dimensional CT Dose Calculation Using a GPU-Based Monte Carlo Simulation for Pencil Beam Scanning Proton Therapy
    Shin, J.
    Paganetti, H.
    MEDICAL PHYSICS, 2020, 47 (06) : E483 - E483
  • [36] A setup and analytical dose calculation model for solid-state sample irradiation using a microtron type e-beam accelerator
    Khasanov, Shakhboz
    Su, Youwu
    Suleymanov, Renat
    Safarov, Akmal
    Safarov, Askar
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2024, : 1568 - 1577
  • [37] A new add-on electron multi-leaf collimation (EMLC): Dosimetry and dose calculation using a new beam model with Voxel Monte Carlo
    Vatanen, T.
    Traneus, E.
    Lahtinen, T.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S166 - S166
  • [38] Changes in Dose-Volume Parameters of Stereotactic Body Radiation Therapy (SBRT) for Stage I Lung Cancer by Calculation Algorithms and Relationship With Outcome: Pencil Beam Convolution With Batho Power Law Versus Anisotropic Analytical Algorithm
    Otsuka, S.
    Baba, F.
    Miyakawa, A.
    Ishikura, S.
    Murai, T.
    Tamura, T.
    Shibamoto, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S618 - S618
  • [39] The Investigation of The Anisotropic Analytical Algorithm (AAA) and the Acuros XB (AXB) Dose Calculation Algorithms Accuracy in Surface and Buildup Region for 6 MV Photon Beam Using Gafchromic EBT3 Film
    Donmez Kesen, Nazmiye
    Koksal Akbas, Canan
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2021, 36 (03): : 365 - 372
  • [40] The treatment results of 40 patients with localized endobronchial cancer with external beam irradiation and intraluminal irradiation using low dose rate 192Ir thin wires with a new catheter
    Fuwa, N
    Ito, Y
    Matsumoto, A
    Morita, K
    RADIOTHERAPY AND ONCOLOGY, 2000, 56 (02) : 189 - 195