ESTABLISHMENT OF DIAGNOSTIC REFERENCE LEVELS FOR DENTAL PANORAMIC RADIOGRAPHY IN GREECE

被引:6
|
作者
Manousaridis, G. [1 ,2 ]
Koukorava, C. [1 ,2 ]
Hourdakis, C. J. [1 ]
Kamenopoulou, V. [1 ]
Yakoumakis, E. [2 ]
Tsiklakis, K. [2 ]
机构
[1] Greek Atom Energy Commiss, Athens, Greece
[2] Univ Athens, Athens 11528, Greece
关键词
PRODUCT;
D O I
10.1093/rpd/ncv088
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of the present study was to present the national diagnostic reference levels (DRL) established for panoramic dental examinations in Greece. The establishment of DRL, as a tool for the optimisation of radiological procedures, is a requirement of national regulations. Measurements performed by the Greek Atomic Energy Commission on 90 panoramic systems have been used for the derivation of DRL values. DRL values have been proposed for exposure settings of different patient types (child, small adult and standard adult), both for film and digital imaging. The DRLs for different patient types are grouped in three categories: children, small adults (corresponding to female) and average adults (corresponding to male). Proposed DRLs for these groups are 2.2, 3.3 and 4.1 mGy, respectively. In order to investigate the correlation of DRLs with the available imaging modalities (CR, DR and film), this parameter was taken into account. DR imaging DRL is the lowest at 3.5 mGy, CR imaging the highest at 4.2 mGy and film imaging at 3.7 mGy. In order to facilitate comparison with other studies, kerma-width product values were calculated from K-i, air and field size.
引用
收藏
页码:111 / 114
页数:4
相关论文
共 50 条
  • [41] Proposed procedure for the establishment of diagnostic reference levels in Germany
    Veit, R
    Bauer, B
    Bernhardt, HJ
    Lechel, U
    RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) : 117 - 120
  • [42] Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging
    Tonkopi, E.
    O'Brien, K.
    MEDICAL PHYSICS, 2016, 43 (06) : 3716 - 3716
  • [43] PROPOSAL FOR LOCAL DIAGNOSTIC REFERENCE LEVELS IN GENERAL RADIOGRAPHY IN JAPAN
    Asada, Yasuki
    Ono, Koji
    Kondo, Yuya
    Sugita, Kazuma
    Ichikawa, Takuma
    Shibata, Hideki
    RADIATION PROTECTION DOSIMETRY, 2019, 187 (03) : 338 - 344
  • [44] Proposed diagnostic reference levels for general radiography and mammography in Japan
    Asada, Yasuki
    Kondo, Yuya
    Kobayashi, Masanao
    Kobayashi, Kenichi
    Ichikawa, Takuma
    Matsunaga, Yuta
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (03) : 867 - 876
  • [45] DETERMINATION OF DIAGNOSTIC REFERENCE LEVELS IN GENERAL RADIOGRAPHY IN LATIN AMERICA
    Blanco, Susana
    Mora, Patricia
    Almonte, Narkiss
    Benavente, Tony
    Benson, Nadja
    Blanco, Daniel
    Cardenas, Juan
    Defaz Gomez, Yolanda
    Edding, Oscar
    Escobar, Carolina
    Fonseca, Maria
    Gamarra, Mirtha
    Garcia Aguilar, Juan
    Khoury, Helen Jamil
    Rosa Quintero, Ana
    Roas Zuniga, Norma
    Zaire, Edgar
    Nader, Alejandro
    RADIATION PROTECTION DOSIMETRY, 2013, 156 (03) : 303 - 309
  • [46] DIAGNOSTIC-IMPORTANCE OF PANORAMIC RADIOGRAPHY IN ENT
    BUNGER, B
    LARYNGO-RHINO-OTOLOGIE, 1990, 69 (04) : 196 - 200
  • [47] Measuring the dose–width product and proposing the local diagnostic reference level in panoramic dental radiography: a multi-center study from Iran
    Akbar Aliasgharzadeh
    Davood Khezerloo
    Bagher Farhood
    Mehran Mohseni
    Habiballah Moradi
    Tohid Mortezazadeh
    Oral Radiology, 2021, 37 : 80 - 85
  • [48] PROPOSALS FOR THE ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS FOR RADIOGRAPHY FOR ADULT PATIENTS BASED ON REGIONAL DOSE SURVEYS IN RUSSIAN FEDERATION
    Vodovatov, A. V.
    Balonov, Mikhail I.
    Golikov, V. Yu.
    Shatsky, I. G.
    Chipiga, L. A.
    Bernhardsson, C.
    RADIATION PROTECTION DOSIMETRY, 2017, 173 (1-3) : 223 - 232
  • [49] Reference doses for dental radiography
    Napier, ID
    BRITISH DENTAL JOURNAL, 1999, 186 (08) : 392 - 396
  • [50] Reference doses for dental radiography
    I D Napier
    British Dental Journal, 1999, 186 (8) : 392 - 396