Comparison of Axial Performance of Cone-Beam Reconstruction Algorithms for Off-Center Flat-Panel Imaging with a SPECT System

被引:0
|
作者
Konate, Souleymane [1 ]
Pretorius, P. Hendrik [1 ]
Mukherjee, Joyeeta Mitra [1 ]
Dey, Joyoni [1 ]
Ritacco, Alan [1 ]
Glick, Stephen [1 ]
O'Connor, J. Michael [1 ]
Shao, Lingxiong [2 ]
Wang, Jiong [2 ]
Feng, Bing
King, Michael A. [1 ]
机构
[1] Univ Massachusetts, Sch Med, Worcester, MA 01605 USA
[2] Philips Med Syst, San Jose, CA USA
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
a In this paper, using Defrise Phantoms, we present our investigations of the axial limitations of an analytical cone-beam reconstruction algorithm (FDK), and the iterative ordered-subset transmission reconstruction (OSTR) iterative algorithm, for an axially extended version of the Philips Brightview XCT cone beam CT (CBCT) geometry. Simulations were preformed for head size and body size Defrise Phantoms of different axial dimensions to investigate limitations on axial extent as a function of these. OSTR yielded overall better axial performance than FDK. It may be possible to reconstruct a larger axial extent for brain studies; however, patients with larger body sizes pose more of a problem.
引用
收藏
页码:3483 / 3491
页数:9
相关论文
共 50 条
  • [41] Optimization of helical orbit cone-beam SPECT imaging performance for brain imaging
    Vogelsang, Levon
    Krol, Andrzej
    Lipson, Edward
    Feiglin, David
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [42] Flat panel detector-based cone-beam volume CT angiography imaging: System evaluation
    Ning, R
    Chen, B
    Yu, RF
    Conover, D
    Tang, XY
    Ning, Y
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) : 949 - 963
  • [43] Investigation of the Lag Effect in X-Ray Flat-Panel Detector for Cone-Beam Computed Tomography
    A. K. Avakyan
    I. L. Dergacheva
    A. A. Elanchik
    T. A. Krylova
    T. K. Lobzhanidze
    S. A. Polikhov
    V. P. Smirnov
    [J]. Biomedical Engineering, 2020, 54 : 179 - 183
  • [44] What Is the Diagnostic Accuracy of Flat-panel Cone-beam CT Arthrography for Diagnosis of Scapholunate Ligament Tears?
    Dornberger, Jenny E.
    Rademacher, Grit
    Stengel, Dirk
    Hoenning, Alexander
    Schueler, Gabriele
    Eisenschenk, Andreas
    Mutze, Sven
    Goelz, Leonie
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2021, 479 (01) : 151 - 160
  • [46] Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2001, 28 (02) : 220 - 231
  • [47] Investigation of the Lag Effect in X-Ray Flat-Panel Detector for Cone-Beam Computed Tomography
    Avakyan, A. K.
    Dergacheva, I. L.
    Elanchik, A. A.
    Krylova, T. A.
    Lobzhanidze, T. K.
    Polikhov, S. A.
    Smirnov, V. P.
    [J]. BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2020, 54 (03): : 179 - 183
  • [48] Model-based iterative reconstruction for flat-panel cone-beam CT with focal spot blur, detector blur, and correlated noise
    Tilley, Steven, II
    Siewerdsen, Jeffrey H.
    Stayman, J. Webster
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01): : 296 - 319
  • [49] Cone-beam CT with a flat-panel detector: From image science to image-guided surgery
    Siewerdsen, Jeffrey H.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S241 - S250
  • [50] A volumetric cone-beam CT system based on a 41x41 cm2 flat-panel imager
    Jaffray, DA
    Siewerdsen, JH
    [J]. MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 800 - 807