Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device

被引:33
|
作者
Zeidan, OA [1 ]
Li, JG [1 ]
Ranade, M [1 ]
Stell, AM [1 ]
Dempsey, JF [1 ]
机构
[1] Univ Florida, Dept Radiat Oncol, Gainesville, FL 32610 USA
关键词
EPID; IMRT; MLC; fluence; CCD camera; IMRT treatment plan verification;
D O I
10.1118/1.1644518
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present an investigation into the use of a fast video-based electronic portal-imaging device (EPID) to study intensity modulated radiation therapy (IMRT) delivery. The aim of this study is to test the feasibility of using an EPID system to independently measure the orchestration of collimator leaf motion and beam fluence; simultaneously measuring both the delivered field fluence and shape as it exits the accelerator head during IMRT delivery. A fast EPID that consists of a terbium-doped gadolinium oxysulphide (GdO2S:Tb) scintillator coupled with an inexpensive commercial 30 frames-per-second (FPS) CCD-video recorder (16.7 ms shutter time) was employed for imaging IMRT delivery. The measurements were performed on a Varian 2100 C/D linear accelerator equipped with a 120-leaf multileaf-collimator (MLQ. A characterization of the EPID was performed that included measurements of spatial resolution, linac pulse-rate dependence, linear output response, signal uniformity, and imaging artifacts. The average pixel intensity for fields imaged with the EPID was found to be linear in the delivered monitor units of static non-lMRT fields between 3 X 3 and 15 X 15 cm(2). A systematic increase of the average pixel intensity was observed with increasing field size, leading to a maximum variation of 8%. Deliveries of a clinical step-andshoot mode leaf sequence were imaged at 600 MU/min. Measurements from this IMRT delivery were compared with experimentally validated MLC controller log files and were found to agree to within 5%. An analysis of the EPID image data allowed identification of three types of errors: (1) 5 out of 35 segments were undelivered; (2) redistributing all of the delivered segment MUs; and (3) leaf movement during segment delivery. Measurements with the EPID at lower dose rates showed poor agreement with log files due to an aliasing artifact. The study was extended to use a highspeed camera (1-1000 FPS and 10 mus shutter time) with our EPID to image the same delivery to demonstrate the feasibility of imaging without aliasing artifacts. High-speed imaging was shown to be a promising direction toward validating IMRT deliveries with reasonable image resolution and noise. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:463 / 476
页数:14
相关论文
共 50 条
  • [1] IMRT verification of step-and-shoot delivery using a fast scintillation camera
    Zeidan, O
    Li, J
    Ranade, M
    Stell, A
    Dempsey, J
    MEDICAL PHYSICS, 2003, 30 (06) : 1418 - 1418
  • [2] Automatic verification of step-and-shoot IMRT field segments using portal imaging
    Woo, MK
    Lightstone, AW
    Shan, G
    Kumaraswamy, L
    Li, Y
    MEDICAL PHYSICS, 2003, 30 (03) : 348 - 351
  • [3] Dosimetric verification of step-and-shoot and tomotherapy delivery methods for IMRT. (WIP)
    Kapulsky, A
    Rebo, I
    Ndlovu, A
    Gejerman, G
    Pipman, Y
    Hanley, J
    MEDICAL PHYSICS, 2003, 30 (06) : 1496 - 1496
  • [4] Portal dose measurements with a video-based electronic portal imaging device using a deconvolution algorithm
    Pasma, KL
    Kroonwijk, M
    Visser, AG
    Heijmen, BJM
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 282 - 284
  • [5] Monte Carlo phase space calculations and experimental verification of step-and-shoot IMRT delivery
    Tyagi, N
    Chetty, I
    Litzenberg, D
    Moran, J
    Bielajew, A
    Fraass, B
    MEDICAL PHYSICS, 2003, 30 (06) : 1421 - 1421
  • [6] A simple video-based EPID technique for pre-treatment verification of step-and-shoot intensity modulation
    Coughlin, M
    Leszczynski, K
    Provost, D
    MEDICAL PHYSICS, 2003, 30 (07) : 1940 - 1940
  • [7] Dosimetric Verification of IMRT Treatment Plans Using an Electronic Portal Imaging Device
    Kruszyna, Marta
    NUCLEAR PHYSICS METHODS AND ACCELERATORS IN BIOLOGY AND MEDICINE, 2009, 1204 : 198 - 199
  • [8] Fast, daily linac verification for segmented IMRT using electronic portal imaging
    Vieira, Sandra C.
    Bolt, Rene A.
    Dirkx, Maarten L. P.
    Visser, Andries G.
    Heijmen, Ben J. M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 80 (01) : 86 - 92
  • [9] Verification of patient position and delivery of IMRT by electronic portal imaging
    Fielding, AL
    Evans, PM
    Clark, CH
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 (03) : 339 - 347
  • [10] Verification of patient position and delivery of IMRT by electronic portal Imaging
    Fielding, AL
    Clark, CH
    Evans, PM
    MEDICAL PHYSICS, 2003, 30 (06) : 1352 - 1352