Evaluation of the wear of the pivot bearing in the Gyro C1E3 pump

被引:31
|
作者
Makinouchi, K
Nakazawa, T
Takami, Y
Takatani, S
Nose, Y
机构
[1] Baylor College of Medicine, Department of Surgery, Houston, TX
[2] Department of Surgery, Baylor College of Medicine, Houston, TX 77030, One Baylor Plaza
关键词
pivot bearing; ultra high molecular weight polyethylene; ceramics; cumulative rotation; loosening distance; impeller position;
D O I
10.1111/j.1525-1594.1996.tb04474.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To estimate the lifetime of the pivot bearing system of the sealless centrifugal Gyro C1E3 pump, pivot bearing wear phenomena of the C1E3 were studied. The pivot bearing system consisted of a male and female pivot made of ceramics and ultrahigh molecular weight polyethylene (UHMWPE), respectively. First, many pumping tests were performed with the C1E3 under various pumping conditions, and the effects of impeller position and fluid on wear were analyzed. Through these preliminary tests, it was found that the wear progress of the pivot bearing consisted of initial wear and stationary wear. Most of this initial wear is caused by the plastic deformation of the polyethylene female pivot. It also was observed that bovine blood was almost comparable to water in its effect on the stationary wear rate at the same rotational speed. Based on these results, a long-term pumping test was performed with the C1E3, and initial and stationary wear rates were determined. At the same time, the maximal loosening distance (LDmax) (permissible total wear) of the C1E3 was determined experimentally from hemolytic and hydraulic performance perspectives. By using experimentally determined parameters the lifetime of the pivot bearing system of the C1E3 pump was estimated for various pumping conditions. The lifetime of the pivot bearing system of the C1E3 was typically 10 years for right ventricular assist, 8 years for left ventricular assist, and 5 years for cardiopulmonary bypass.
引用
收藏
页码:523 / 528
页数:6
相关论文
共 50 条
  • [1] Material of the double pivot bearing system in the Gyro C1E3 centrifugal pump
    Takami, Y
    Nakazawa, T
    Makinouchi, K
    Benkowski, R
    Glueck, J
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1997, 21 (02) : 143 - 147
  • [3] Effect of surface roughness on hemolysis in a pivot bearing supported gyro centrifugal pump (C1E3)
    Takami, Y
    Makinouchi, K
    Nakazawa, T
    Glueck, J
    Benkowski, R
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1996, 20 (11) : 1155 - 1161
  • [4] Pump power loss and heat generation in a pivot bearing-supported gyro centrifugal pump (C1E3)
    Takami, Y
    Nakazawa, T
    Makinouchi, K
    Glueck, J
    Ohara, Y
    Benkowski, RJ
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1996, 20 (07) : 794 - 797
  • [5] Hemolytic characteristics of a pivot bearing supported gyro centrifugal pump (C1E3) simulating various clinical applications
    Takami, Y
    Makinouchi, K
    Nakazawa, T
    Benkowski, R
    Glueck, J
    Ohara, Y
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1996, 20 (09) : 1042 - 1049
  • [6] Hemolytic characteristics of a pivot bearing supported Gyro centrifugal pump (C1E3) simulating various clinical applications
    Department of Surgery, Baylor College of Medicine, Houston, TX, United States
    不详
    [J]. ARTIF. ORGANS, 9 (1042-1049):
  • [7] Clinical evaluation of the Gyro Pump C1e3 as a cardiopulmonary bypass pump
    Orime, Y
    Shiono, M
    Yagi, S
    Yamamoto, T
    Okumura, H
    Kimura, S
    Hata, M
    Sezai, A
    Kashiwazaki, S
    Choh, S
    Negishi, N
    Sezai, Y
    Matsui, T
    Suzuki, M
    [J]. ASAIO JOURNAL, 2000, 46 (01) : 128 - 133
  • [8] Morphological changes of the pivot bearings in the Gyro Pump C1E3 after clinical use
    Takami Y.
    Ina H.
    Ohara Y.
    [J]. Journal of Artificial Organs, 2001, 4 (2) : 156 - 160
  • [9] Safety margin of magnetic coupling distance in decoupling of a pivot bearing-supported gyro centrifugal pump (C1E3)
    Takami, Y
    Nakazawa, T
    Makinouchi, K
    Glueck, J
    Benkowski, R
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1996, 20 (07) : 817 - 819
  • [10] In vitro evaluation of the heparin-coated Gyro C1E3 blood pump
    Muramatsu, K
    Masuoka, T
    Fujisawa, A
    [J]. ARTIFICIAL ORGANS, 2001, 25 (07) : 585 - 590