Evaluation of the wear of the pivot bearing in the Gyro C1E3 pump

被引:31
|
作者
Makinouchi, K
Nakazawa, T
Takami, Y
Takatani, S
Nose, Y
机构
[1] Baylor College of Medicine, Department of Surgery, Houston, TX
[2] Department of Surgery, Baylor College of Medicine, Houston, TX 77030, One Baylor Plaza
关键词
pivot bearing; ultra high molecular weight polyethylene; ceramics; cumulative rotation; loosening distance; impeller position;
D O I
10.1111/j.1525-1594.1996.tb04474.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To estimate the lifetime of the pivot bearing system of the sealless centrifugal Gyro C1E3 pump, pivot bearing wear phenomena of the C1E3 were studied. The pivot bearing system consisted of a male and female pivot made of ceramics and ultrahigh molecular weight polyethylene (UHMWPE), respectively. First, many pumping tests were performed with the C1E3 under various pumping conditions, and the effects of impeller position and fluid on wear were analyzed. Through these preliminary tests, it was found that the wear progress of the pivot bearing consisted of initial wear and stationary wear. Most of this initial wear is caused by the plastic deformation of the polyethylene female pivot. It also was observed that bovine blood was almost comparable to water in its effect on the stationary wear rate at the same rotational speed. Based on these results, a long-term pumping test was performed with the C1E3, and initial and stationary wear rates were determined. At the same time, the maximal loosening distance (LDmax) (permissible total wear) of the C1E3 was determined experimentally from hemolytic and hydraulic performance perspectives. By using experimentally determined parameters the lifetime of the pivot bearing system of the C1E3 pump was estimated for various pumping conditions. The lifetime of the pivot bearing system of the C1E3 was typically 10 years for right ventricular assist, 8 years for left ventricular assist, and 5 years for cardiopulmonary bypass.
引用
收藏
页码:523 / 528
页数:6
相关论文
共 50 条
  • [21] Platelet activation in the Gyro C1E3 centrifugal pump: Comparison with the Terumo Capiox and the Nikkiso HPM-15
    Kawahito, K
    Adachi, H
    Ino, T
    [J]. ARTIFICIAL ORGANS, 2000, 24 (11) : 889 - 892
  • [22] Particles released from the Gyro C1E3 during simulated extracorporeal circulation
    Takano, T
    Nakata, K
    Schulte-Eistrup, S
    Kawahito, S
    Glueck, J
    Clarke, IC
    Williams, PA
    Maeda, T
    Nonaka, K
    Linneweber, J
    Yoshikawa, M
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    [J]. ARTIFICIAL ORGANS, 2000, 24 (06) : 446 - 449
  • [23] LONG-TERM CLINICAL USE OF EXTRACORPOREAL CARDIOPULMONARY ASSIST DEVICE WITH THE GYRO C1E3 CENTRIFUGAL BLOOD PUMP
    Tabayashi, N.
    Kameda, Y.
    Yoshikawa, Y.
    Abe, T.
    Naito, H.
    Hayata, Y.
    Hirose, T.
    Yokota, M.
    Iwashita, Y.
    Taniguchi, S.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2008, 31 (07): : 620 - 620
  • [24] Eccentric inlet port of the pivot bearing supported Gyro centrifugal pump
    Takami, Y
    Andrade, A
    Nakazawa, T
    Makinouchi, K
    Glueck, J
    Benkowski, R
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1997, 21 (04) : 312 - 317
  • [25] Improvement of anti-wear function of pivot bearing for centrifugal blood pump (part 1) - Wear mechanisms in spinning motion
    Nonaka, K
    Maruyama, O
    Yamane, T
    Miyoshi, H
    Mizuhara, K
    [J]. JOURNAL OF JAPANESE SOCIETY OF TRIBOLOGISTS, 2003, 48 (05) : 394 - 401
  • [26] New centrifugal blood pump with dual impeller and double pivot bearing system: Wear evaluation in bearing system, performance tests, and preliminary hemolysis tests
    Bock, Eduardo
    Ribeiro, Adriana
    Silva, Maxwell
    Antunes, Pedro
    Fonseca, Jeison
    Legendre, Daniel
    Leme, Juliana
    Arruda, Celso
    Biscegli, Jose
    Nicolosi, Denys
    Andrade, Aron
    [J]. ARTIFICIAL ORGANS, 2008, 32 (04) : 329 - 333
  • [27] DEVELOPMENT AND EVALUATION OF ANTITHROMBOGENIC CENTRIFUGAL PUMP - THE BAYLOR-C-GYRO-PUMP-ECCENTRIC-INLET-PORT-MODEL
    OHARA, Y
    MAKINOUCHI, K
    GLUECK, J
    SUTHERLAND, B
    SHIMONO, T
    NAITO, K
    TASAI, K
    ORIME, Y
    TAKATANI, S
    NOSE, Y
    [J]. ARTIFICIAL ORGANS, 1994, 18 (09) : 673 - 679
  • [28] STUDIES ON MONOCLONAL-ANTIBODY PRODUCTION BY A HYBRIDOMA CELL-LINE (C1E3) IMMOBILIZED IN A FIXED-BED, POROSPHERE CULTURE SYSTEM
    RACHER, AJ
    LOOBY, D
    GRIFFITHS, JB
    [J]. JOURNAL OF BIOTECHNOLOGY, 1990, 15 (1-2) : 129 - 145
  • [29] Wear evaluation and microstructure analysis of cryogenically treated AISI 440C bearing steel
    Idayan, A.
    Elanchezhian, C.
    Ramnath, B. Vijaya
    Palanikumar, K.
    [J]. TRANSACTIONS OF THE CANADIAN SOCIETY FOR MECHANICAL ENGINEERING, 2022, 46 (01) : 175 - 187
  • [30] Wear evaluation and microstructure analysis of cryogenically treated AISI 440C bearing steel
    Idayan, A.
    Elanchezhian, C.
    Ramnath, B. Vijaya
    Palanikumar, K.
    [J]. Idayan, A. (idayanphd@gmail.com), 1600, Canadian Science Publishing (46): : 175 - 187