A geometric view of signal sensitivity metrics in multi-echo fMRI

被引:1
|
作者
Liu, Thomas T. [1 ,2 ,3 ,4 ]
Li, Bochao [5 ]
Fernandez, Brice [6 ]
Banerjee, Suchandrima [7 ]
机构
[1] Univ Calif San Diego, Ctr Funct MRI, 9500 Gilman Dr MC 0677, La Jolla, CA 92093 USA
[2] Univ Calif San Diego, Dept Radiol, 9500 Gilman Dr, La Jolla, CA 92093 USA
[3] Univ Calif San Diego, Dept Psychiat, 9500 Gilman Dr, La Jolla, CA 92093 USA
[4] Univ Calif San Diego, Dept Bioengn, 9500 Gilman Dr, La Jolla, CA 92093 USA
[5] Univ Southern Calif, Dept Biomed Engn, Los Angeles, CA 90007 USA
[6] GE Healthcare, Buc, France
[7] GE Healthcare, Menlo Pk, CA USA
关键词
fMRI; Multi-echo; CNR; tSNR; BOLD-CONTRAST SENSITIVITY; PHYSIOLOGICAL NOISE; OPTIMIZATION; ENHANCEMENT; EPI; MRI;
D O I
10.1016/j.neuroimage.2022.119409
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
In multi-echo fMRI (ME-fMRI), two metrics have been widely used to measure the performance of various acquisition and analysis approaches. These are temporal SNR (tSNR) and differential contrast-to-noise ratio (dCNR). A key step in ME-fMRI is the weighted combination of the data from multiple echoes, and prior work has examined the dependence of tSNR and dCNR on the choice of weights. However, most studies have focused on only one of these two metrics, and the relationship between the two metrics has not been examined. In this work, we present a geometric view that offers greater insight into the relation between the two metrics and their weight dependence. We identify three major regimes: (1) a tSNR robust regime in which tSNR is robust to the weight selection with most weight variants achieving close to optimal performance, whereas dCNR shows a pronounced dependence on the weights with most variants achieving suboptimal performance; (2) a dCNR robust regime in which dCNR is robust to the weight selection with most weight variants achieving close to optimal performance, while tSNR exhibits a strong dependence on the weights with most variants achieving significantly lower than optimal performance; and (3) a within-type robust regime in which both tSNR and dCNR achieve nearly optimal performance when the form of the weights are variants of their respective optimal weights and exhibit a moderate decrease in performance for other weight variants. Insight into the behavior observed in the different regimes is gained by considering spherical representations of the weight dependence of the components used to form each metric. For multi-echo acquisitions, dCNR is shown to be more directly related than tSNR to measures of CNR and signal-to-noise ratio (SNR) for task-based and resting-state fMRI scans, respectively.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Assessment of microstructural signal compartments across the corpus callosum using multi-echo gradient recalled echo at 7 T
    Thapaliya, Kiran
    Vegh, Viktor
    Bollmann, Steffen
    Barth, Markus
    NEUROIMAGE, 2018, 182 : 407 - 416
  • [42] Improved resting state functional connectivity sensitivity and reproducibility using a multiband multi-echo acquisition
    Cohen, Alexander D.
    Yang, Baolian
    Fernandez, Brice
    Banerjee, Suchandrima
    Wang, Yang
    NEUROIMAGE, 2021, 225
  • [43] Comparison of multi-echo spiral and echo planar imaging in functional MRI
    Klarhöfer, M
    Barth, M
    Moser, E
    MAGNETIC RESONANCE IMAGING, 2002, 20 (04) : 359 - 364
  • [44] Comparing the efficacy of data-driven denoising methods for a multi-echo fMRI acquisition at 7T
    Beckers, Abraham B.
    Drenthen, Gerhard S.
    Jansen, Jacobus F. A.
    Backes, Walter H.
    Poser, Benedikt A.
    Keszthelyi, Daniel
    NEUROIMAGE, 2023, 280
  • [45] Characterization of BOLD activation in multi-echo fMRI data using fuzzy cluster analysis and a comparison with quantitative modeling
    Barth, M
    Windischberger, C
    Klarhöfer, M
    Moser, E
    NMR IN BIOMEDICINE, 2001, 14 (7-8) : 484 - 489
  • [46] Robust resting state fMRI processing for studies on typical brain development based on multi-echo EPI acquisition
    Kundu, Prantik
    Benson, Brenda E.
    Baldwin, Katherine L.
    Rosen, Dana
    Luh, Wen-Ming
    Bandettini, Peter A.
    Pine, Daniel S.
    Ernst, Monique
    BRAIN IMAGING AND BEHAVIOR, 2015, 9 (01) : 56 - 73
  • [47] Robust resting state fMRI processing for studies on typical brain development based on multi-echo EPI acquisition
    Prantik Kundu
    Brenda E. Benson
    Katherine L. Baldwin
    Dana Rosen
    Wen-Ming Luh
    Peter A. Bandettini
    Daniel S. Pine
    Monique Ernst
    Brain Imaging and Behavior, 2015, 9 : 56 - 73
  • [48] Imaging of amygdala activation during delayed aversive conditioning using single-shot multi-echo fMRI
    Weiskopf, N
    Veit, R
    Mathiak, K
    Grodd, W
    Birbaumer, N
    NEUROIMAGE, 2001, 13 (06) : S486 - S486
  • [49] Social reward and nonsocial reward processing across the adult lifespan: An interim multi-echo fMRI and diffusion dataset
    Smith, David, V
    Sharp, Cooper J.
    Dachs, Abraham
    Wyngaarden, James
    Sazhin, Daniel
    Yang, Yi
    Kos, Melanie
    Tropea, Tia
    Kohli, Ishika
    Clithero, John A.
    Olson, Ingrid
    Giovannetti, Tania
    Fareri, Dominic
    Jarcho, Johanna M.
    DATA IN BRIEF, 2024, 56
  • [50] Signal compartments in ultra-high field multi-echo gradient echo MRI reflect underlying tissue microstructure in the brain
    Kadamangudi, Shrinath
    Reutens, David
    Sood, Surabhi
    Vegh, Viktor
    NEUROIMAGE, 2018, 178 : 403 - 413