Exploiting sequential injection analysis technique to automate on-line sample treatment and quantitative determination of morphine in human urine

被引:38
|
作者
Idris, Abubakr M. [1 ]
Alnajjar, Ahmed O. [1 ]
机构
[1] King Faisal Univ, Dept Chem, Al Hufuf 31982, Saudi Arabia
关键词
Morphine; Sequential injection analysis; Forensic analysis; Clinical analysis;
D O I
10.1016/j.talanta.2008.04.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple uni-stream sequential injection analysis (SIA) manifold was developed to automate a method for the assay of morphine in human sample. The proposed SIA method includes on-line sample treatment, coupling reaction and spectrophotometric measurement. A rapid algorithm controlled the adopted procedure was critically programmed. For sample treatment, solid-phase extraction (SPE) was carried out into a homemade microcolumn, installed in the SIA manifold. Sufficient sample clean-Lip, extraction and preconcentration Were obtained by SPE. A coupling reaction of morphine with diazonium salt of aniline hydrochloride was adapted to SIA. The product of the reaction, an azo-morphine derivative, was spectrophotometrically detected at 390 nm. Parameters that influenced the efficiency of the proposed method, including Solution Volumes, diazonium Concentration, flow Fate and residence time, Were optimized. The proposed method was linear in a range of 0.10-2.5 mu g ml(-1). The limits of detection and quantification were 0.023 and 0.076 mu g ml(-1), respectively. The detectability of the method was enhanced by the preconcentration a rid the use of an extended path length (50 mm) of a flow cell. The method was validated by an H PLC method. Comparable results with respect to accuracy(recovery 96.3-97.1), repeatability (R.S.D. < 2.4%) and intermediate precision (R.S.D. < 3.1) were gained. The full-automation and miniaturization of the utilized technique offer rapidity, safety in handling urine sample and reagents as well as reduction of reagent and sample volumes. The method is Suitable for the application in forensic cases as an initial test and clinical analysis to prevent overdose-induced toxicity. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:522 / 526
页数:5
相关论文
共 50 条
  • [1] Using on-line solid phase extraction for determination of amiloride in human urine by sequential injection technique
    Huclova, Jitka
    Satinsky, Dalibor
    Pavlicek, Ondrej
    Vedralova, Lucie
    Karlicek, Rolf
    ANALYTICA CHIMICA ACTA, 2006, 573 : 376 - 382
  • [2] Simultaneous determination of iodide and creatinine in human urine by flow analysis with an on-line sample treatment column
    Sitanurak, Jirayu
    Inpota, Prawpan
    Mantim, Thitirat
    Ratanawimarnwong, Nuanlaor
    Wilairat, Prapin
    Nacapricha, Duangjai
    ANALYST, 2015, 140 (01) : 295 - 302
  • [3] Flow injection-capillary electrophoresis coupling to automate on-line sample treatment for the determination of inorganic ions in waters
    Arce, L
    Ríos, A
    Valcárcel, M
    JOURNAL OF CHROMATOGRAPHY A, 1997, 791 (1-2) : 279 - 287
  • [4] Recent advances in on-line solvent extraction exploiting flow injection/sequential injection analysis
    Miró, M
    Estela, JM
    Cerdà, V
    CURRENT ANALYTICAL CHEMISTRY, 2005, 1 (03) : 329 - 343
  • [5] Expanding the separation capability of sequential injection chromatography: Determination of melamine in milk exploiting micellar medium and on-line sample preparation
    Batista, Alex D.
    Nascimento, Carina F.
    Melchert, Wanessa R.
    Rocha, Fabio R. P.
    MICROCHEMICAL JOURNAL, 2014, 117 : 106 - 110
  • [6] Automatic On-line Solid-phase Extraction-Electrothermal Atomic Absorption Spectrometry Exploiting Sequential Injection Analysis for Trace Vanadium, Cadmium and Lead Determination in Human Urine Samples
    Giakisikli, Georgia
    Quezada, Alejandro Ayala
    Tanaka, Junpei
    Anthemidis, Aristidis N.
    Murakami, Hiroya
    Teshima, Norio
    Sakai, Tadao
    ANALYTICAL SCIENCES, 2015, 31 (05) : 383 - 389
  • [7] Automatic On-line Solid-phase Extraction-Electrothermal Atomic Absorption Spectrometry Exploiting Sequential Injection Analysis for Trace Vanadium, Cadmium and Lead Determination in Human Urine Samples
    Georgia Giakisikli
    Alejandro Ayala Quezada
    Junpei Tanaka
    Aristidis N. Anthemidis
    Hiroya Murakami
    Norio Teshima
    Tadao Sakai
    Analytical Sciences, 2015, 31 : 383 - 389
  • [8] On-line dynamic fractionation and automatic determination of inorganic phosphorus in environmental solid substrates exploiting sequential injection microcolumn extraction and flow injection analysis
    Buanuam, Janya
    Miro, Manuel
    Hansen, Elo Harald
    Shiowatana, Juwadee
    ANALYTICA CHIMICA ACTA, 2006, 570 (02) : 224 - 231
  • [9] On-line capillary based quantitative analysis of a heterocyclic amine in human urine
    Viberg, Peter
    Wahlund, Karl-Gustav
    Skog, Kerstin
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1133 (1-2) : 347 - 352
  • [10] Sequential-injection analysis (SIA): A useful tool for on-line sample-handling and pre-treatment
    Economou, A
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2005, 24 (05) : 416 - 425