Accuracy and user performance evaluation of the Contour® Next Link 2.4 blood glucose monitoring system

被引:10
|
作者
Bailey, Timothy [1 ]
Wallace, Jane F. [2 ]
Greene, Carmine [2 ]
Pardo, Scott [2 ]
Brown, Daniel [2 ]
Pflug, Brian [2 ]
Klaff, Leslie J. [3 ]
机构
[1] AMCR Inst Inc, Escondido, CA 92025 USA
[2] Bayer HealthCare LLC, Diabet Care, Whippany, NJ 07981 USA
[3] Rainier Clin Res Ctr, Renton, WA 98057 USA
关键词
Self-monitoring of blood glucose; Type; 1; diabetes; 2; INSULIN; THERAPY;
D O I
10.1016/j.cca.2015.06.001
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The Contour (R) Next Link 2.4 blood glucose monitoring system (BGMS), which communicates wirelessly exclusively with a Medtronic insulin pump, was assessed in 2 separate studies. Methods: In the laboratory study (N = 100), subject capillary fingertip samples were tested in duplicate using 3 test strip lots; accuracy was assessed based on ISO 15197:2013 section 63 accuracy criteria (95% within +/- 15 mg/dl or +/- 15% of reference for glucose <100 and >= 100 mg/dl, respectively). In the clinical trial (N = 219), untrained subjects with diabetes tested capillary fingertip and palm blood samples and completed an ease-of-use questionnaire. BGMS and YSI glucose analyzer results were compared based on ISO 15197:2013 section 8 accuracy criteria. Results: In the laboratory study, 100% of results met ISO 15197:2013 section 6.3 accuracy criteria. In the clinical trial, 98.6% of capillary fingertip and 972% of palm results from subjects met ISO 15197:2013 section 8 accuracy criteria. By Parkes-Consensus Error Grid analysis, 100% of subject capillary fingertip results were within Zone A. Based on questionnaire results, the majority of subjects found the BGMS easy to use. Conclusions: The BGMS exceeds ISO 15197:2013 section 63 and section 8 accuracy criteria in the laboratory and in the hands of untrained users with diabetes. (C) 2015 The Authors. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:139 / 145
页数:7
相关论文
共 50 条
  • [31] Accuracy and Precision of the CONTOUR® PLUS Blood Glucose Monitoring System as Used by Health Care Professionals
    Ji, Linong
    Guo, Lixin
    Pardo, Scott
    Huang, April
    Yeo, Sandy
    DIABETES, 2014, 63 : A596 - A596
  • [32] ACCURACY ASSESSMENT OF THE CONTOUR TS BLOOD GLUCOSE MONITORING SYSTEM CONTAINING AN UPDATED METER ALGORITHM
    Harrison, B.
    Simmons, D. A.
    Warchal-Windham, M. E.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 : A81 - A81
  • [33] Clinical evaluation of the Bayer Ascensia® Contour® blood glucose monitoring system in neonates.
    Griffey, M. A.
    Wilhite, T. R.
    Heacook, J.
    Dietzen, D. J.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A216 - A217
  • [34] An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose
    Cao, Bingyan
    Wang, Rui
    Gong, Chunxiu
    Wu, Di
    Su, Chang
    Chen, Jiajia
    Yi, Yajun
    Liu, Min
    Liang, Xuejun
    Li, Wenjing
    JOURNAL OF DIABETES RESEARCH, 2019, 2019
  • [35] An evaluation of the accuracy of a flash glucose monitoring system in children with diabetes in comparison with venous blood glucose
    Cao, Bingyan
    Wang, Rui
    Gong, Chunxiu
    Wu, Di
    Su, Chang
    Chen, Jiajia
    Yi, Yajun
    Liu, Min
    Liang, Xuejun
    Li, Wenjing
    HORMONE RESEARCH IN PAEDIATRICS, 2019, 91 : 287 - 287
  • [36] System accuracy evaluation of blood glucose monitoring systems for point of care testing and self monitoring
    Freckmann, G.
    Baumstark, A.
    Link, M.
    Pleus, S.
    Stolberg, E.
    Haug, C.
    Sieber, J.
    DIABETOLOGIA, 2013, 56 : S432 - S433
  • [37] Miniaturized, User-Insertable Wireless Continuous Glucose Monitoring System with Next Generation Speed, Accuracy, and Longevity
    Mujeeb-U-Rahman, Muhammad
    Nazari, Meisam Honarvar
    Sencan, Mehmet
    DIABETES, 2020, 69
  • [38] System Accuracy Evaluation of the GlucoRx®Nexus TD-4280 Blood Glucose Monitoring System
    Khan, Muhammad
    Broadbent, Keith
    Morris, Mike
    Ewins, David
    Joseph, Franklin
    DIABETES, 2014, 63 : A222 - A223
  • [39] Accuracy validation and user performance analysis of two new self-monitoring blood glucose systems
    Huang, Tun-Yu
    Chen, Yu-Hsi
    Li, Yu-Ching
    Chiu, Yen-Hsin
    Kang, Hui-Jun
    Chen, Chih-Yu
    ENDOCRINOLOGY DIABETES AND METABOLISM CASE REPORTS, 2025, 2025 (01):
  • [40] Accuracy of self-monitoring Glycemic Contour® XT used with Contour® next reactive strips: results & evaluation
    Barberot, V.
    CORRESPONDANCES EN METABOLISMES HORMONES DIABETES ET NUTRITION, 2012, 16 (5-6): : 11 - 14