Comparison of the Accuracy of PET/CT and PET/MRI Spatial Registration of Multiple Metastatic Lesions

被引:36
|
作者
Rakheja, Rajan [1 ]
DeMello, Linda [1 ]
Chandarana, Hersh [2 ]
Glielmi, Christopher [3 ]
Geppert, Christian [3 ]
Faul, David [3 ]
Friedman, Kent P. [1 ]
机构
[1] NYU, Langone Med Ctr, Dept Radiol, Div Nucl Med, New York, NY 10016 USA
[2] NYU, Langone Med Ctr, Dept Radiol, Abdominal Imaging Sect, New York, NY 10016 USA
[3] Siemens Healthcare, MR R&D Collaborat, New York, NY USA
关键词
MRI; PET/CT; PET/MRI; registration; respiratory motion; MOTION ARTIFACTS; ATTENUATION CORRECTION; FDG PET; MRI; CT; REDUCTION; SCANNER;
D O I
10.2214/AJR.13.11305
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to compare the accuracy of the spatial registration of conventional PET/CT with that of hybrid PET/MRI of patients with FDG-avid metastatic lesions. SUBJECTS AND METHODS. Thirteen patients with known metastatic lesions underwent FDG PET/CT followed by PET/MRI with a hybrid whole-body system. The inclusion criterion for tumor analysis was spherical or oval FDG-avid tumor clearly identified with both CT and MRI. The spatial coordinates (x, y, z) of the visually estimated centers of the lesions were determined for PET/CT (PET and CT independently) and PET/MRI (PET, T1-weighted gradient-echo sequence with radial stack-of-stars trajectory, T2-weighted sequence), and the b0 images of an echo-planar imaging (EPI) diffusion-weighted imaging (DWI) acquisition. All MRI sequences were performed in the axial plane with free breathing. The spatial coordinates of the estimated centers of the lesions were determined for PET and CT and PET and MRI sequences. Distance between the isocenter of the lesion on PET images and on the images obtained with the anatomic modalities was measured, and misregistration (in millimeters) was calculated. The degree of misregistration was compared between PET/CT and PET/MRI with a paired Student t test. RESULTS. Nineteen lesions were evaluated. On PET/CT images, the average of the total misregistration in all planes of CT compared with PET was 4.13 +/- 4.24 mm. On PET/MR images, lesion misregistration between PET and T1-weighted gradient-echo images had a shift of 2.41 +/- 1.38 mm and between PET and b0 DW images was 5.97 +/- 2.83 mm. Similar results were calculated for 11 lesions on T2-weighted images. The shift on T2-weighted images compared with PET images was 2.24 +/- 1.12 mm. Paired Student t test calculations for PET/CT compared with PET/MRI T1-weighted gradient-echo images with a radial stack-of-stars trajectory, b0 DW images, and T2-weighted images showed significant differences (p < 0.05). Similar results were seen in the analysis of six lung lesions. CONCLUSION. PET/MRI T1-weighted gradient-echo images with a radial stack-of-stars trajectory and T2-weighted images had more accurate spatial registration than PET/CT images. This may be because that the whole-body PET/MRI system used can perform simultaneous acquisition, whereas the PET/CT system acquires data sequentially. However, the EPI-based b0 DWI datasets were significantly misregistered compared with the PET/CT datasets, especially in the thorax. Radiologists reading PET/MR images should be aware of the potential for misregistration on images obtained with EPI-based DWI sequences because of inherent spatial distortion associated with this type of MRI acquisition.
引用
收藏
页码:1120 / 1123
页数:4
相关论文
共 50 条
  • [31] Accuracy of rigid CT-FDG-PET image registration of the liver
    van Dalen, JA
    Vogel, W
    Huisman, HJ
    Oyen, WJG
    Jager, GJ
    Karssemeijer, N
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (23): : 5393 - 5405
  • [32] Detection of pelvic lesions in patients with endometrial cancer with fused PET/MRI. Compared to PET/CT or MRI
    Juri, H.
    Komori, T.
    Hamada, T.
    Tanaka, Y.
    Higashiyama, A.
    Nakai, G.
    Narumi, Y.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S284 - S284
  • [33] High accuracy of volumetric image registration of CT, NIR and PET images
    Li, G.
    Xie, H.
    Ning, H.
    Citrin, D.
    Capala, J.
    Arora, B.
    Coleman, C.
    Camphausen, K.
    Miller, R.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2037 - 2037
  • [34] Stereotactic imaging for radiotherapy:: accuracy of CT, MRI, PET and SPECT
    Karger, CP
    Hipp, P
    Henze, M
    Echner, G
    Höss, A
    Schad, L
    Hartmann, GH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (02): : 211 - 221
  • [35] Pulmonary amyloidosis mimicking multiple metastatic lesions on F-18 FDG PET/CT
    Seo, Ji Hyoung
    Lee, Sang Woo
    Ahn, Byeong-Cheol
    Lee, Jaetae
    [J]. LUNG CANCER, 2010, 67 (03) : 376 - 379
  • [36] Comparison of FDG PET/MRI and FDG PET/CT in Pediatric Oncology in Terms of Anatomic Correlation of FDG-positive Lesions
    Uslu-Besli, Lebriz
    Kapucu, Lutfiye Ozlem Atay
    Karadeniz, Ceyda
    Akdemir, Umit Ozgur
    Pinarli, Faruk Guclu
    Aydos, Uguray
    Okur, Arzu
    Kaya, Zuhre
    Samanci, Cesur
    Karabacak, Nese I.
    [J]. JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 2019, 41 (07) : 542 - 550
  • [37] A new method for improving PET and CT registration in the thorax on PET/CT
    Pan, T
    Mawlawi, O
    Nehmeh, S
    Erdi, Y
    Luo, D
    Liu, H
    Hunjan, S
    Balter, P
    Prado, K
    Mohan, R
    Macapinlac, H
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1943 - 1943
  • [38] 18F-FDG-PET and MRI in patients with malignancies of the liver and pancreas Accuracy of retrospective multimodality image registration by using the CT-component of PET/CT
    Donati, O. F.
    Reiner, C. S.
    Hany, T. F.
    Fornaro, J.
    von Schulthess, G. K.
    Marincek, B.
    Weishaupt, D.
    [J]. NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2010, 49 (03): : 106 - 114
  • [39] Imaging of Multiple Myeloma: Usefulness of MRI and PET/CT
    Koppula, Bhasker
    Kaptuch, Justin
    Hanrahan, Christopher J.
    [J]. SEMINARS IN ULTRASOUND CT AND MRI, 2013, 34 (06) : 566 - 577
  • [40] Comparison of PET/CT with Sequential PET/MRI Using an MR-Compatible Mobile PET System
    Nakamoto, Ryusuke
    Nakamoto, Yuji
    Ishimori, Takayoshi
    Fushimi, Yasutaka
    Kido, Aki
    Togashi, Kaori
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (05) : 846 - 851