An initial investigation of a wireless patient radiation dosimeter for use in interventional radiology

被引:26
|
作者
Inaba, Yohei [1 ,2 ]
Chida, Koichi [1 ,2 ]
Murabayashi, Yuuki [1 ]
Endo, Mime [1 ]
Otomo, Kazuki [1 ]
Zuguchi, Masayuki [1 ]
机构
[1] Tohoku Univ, Grad Sch Med, Dept Radiol Technol, Aoba Ku, 2-1 Seiryo, Sendai, Miyagi 9808575, Japan
[2] Tohoku Univ, Int Res Inst Disaster Sci, Div Disaster Med, Aoba Ku, Sendai, Miyagi 9808579, Japan
基金
日本学术振兴会;
关键词
Interventional radiology (IR); Radiation safety; Dosimetry; Radiation protection; Radiation dose; Fluoroscopy; FOLLOW-UP; SKIN; PHYSICIANS; TIME; PROTECTION;
D O I
10.1007/s12194-020-00575-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiation exposure during interventional radiology (IR) procedures is a critical issue. We have developed a wireless real-time dosimeter for IR patients that use nontoxic phosphor (four sensors). We evaluated the basic performance parameters (such as dose linearity, batch uniformity, reproducibility, and wireless-communication conditions) of the developed system using an IR X-ray system. Further, we investigated the influence of noise from other medical equipment on our wireless real-time dosimeter in the IR X-ray room. Overall, our wireless system exhibited excellent performance in terms of uniformity, reproducibility, and linearity; moreover, the wireless communication performance was better. The developed system enabled real-time visualization of patient radiation dose, without noise contamination from other medical equipment. In addition, the wireless system can be easily installed in a location where the PC screen (display) can be readily viewed by the IR physician. Hence, we developed a wireless system that can display the patient radiation dose data in real time; the system performed satisfactorily upon application in radiation dosimetry. Therefore, our wireless system will facilitate the real-time monitoring/management of patient radiation dose during IR.
引用
收藏
页码:321 / 326
页数:6
相关论文
共 50 条
  • [21] Quality Initiatives Establishing an Interventional Radiology Patient Radiation Safety Program
    Steele, Joseph R.
    Jones, A. Kyle
    Ninan, Elizabeth P.
    RADIOGRAPHICS, 2012, 32 (01) : 277 - 287
  • [22] An audit of patient radiation doses in interventional radiology at a South African hospital
    Slave, Oneile
    Mahomed, Nasreen
    SA JOURNAL OF RADIOLOGY, 2023, 27 (01):
  • [23] Patient Radiation Dose Reduction Considerations in a Contemporary Interventional Radiology Suite
    Panick, Catherine
    Wunderle, Kevin
    Sands, Mark
    Martin, Charles
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 41 (12) : 1925 - 1934
  • [24] Management of Patient and Staff Radiation Dose in Interventional Radiology: Current Concepts
    Bartal, Gabriel
    Vano, Eliseo
    Paulo, Graciano
    Miller, Donald L.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2014, 37 (02) : 289 - 298
  • [25] Patient Radiation Dose Reduction Considerations in a Contemporary Interventional Radiology Suite
    Catherine Panick
    Kevin Wunderle
    Mark Sands
    Charles Martin
    CardioVascular and Interventional Radiology, 2018, 41 : 1925 - 1934
  • [26] Sepsis in the Interventional Radiology Patient
    Miller, Michael J.
    Smith, Tony P.
    Ryan, J. Mark
    TECHNIQUES IN VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 9 (02) : 64 - 68
  • [27] Interventional radiology for the obstetric patient
    Lambrecht, Sigrid
    Van de Velde, Marc
    CURRENT OPINION IN ANESTHESIOLOGY, 2020, 33 (04) : 566 - 570
  • [28] Ergonomic investigation of interventional radiology
    Shinohara, Kazuhiko
    6TH INTERNATIONAL CONFERENCE ON APPLIED HUMAN FACTORS AND ERGONOMICS (AHFE 2015) AND THE AFFILIATED CONFERENCES, AHFE 2015, 2015, 3 : 308 - 311
  • [29] Patient care in interventional radiology
    Maskell, G.
    CLINICAL RADIOLOGY, 2023, 78 (04) : 286 - 287
  • [30] Sepsis in the interventional radiology patient
    Smith, TP
    Ryan, JM
    Niklason, LE
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (04) : 317 - 325