A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study

被引:15
|
作者
Rubio, Jorge [1 ]
Zarzosa, Jose Ignacio [1 ]
Pahares, Antonio [1 ]
机构
[1] Catholic Univ Valencia, Dept Endodont & Restorat Dent, Valencia, Spain
关键词
Dental High-Speed Equipment; Fatigue; Torque; WAVEONE GOLD; ROTARY INSTRUMENTS; CURRENT CHALLENGES; CANAL PREPARATION; RECIPROC BLUE; RESISTANCE; FILES; SYSTEMS; ONESHAPE; PROTAPER;
D O I
10.15644/asc53/1/3
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: This study was designed to test the null hypothesis that there were no significant differences between size 25 files F360, F6 SkyTaper, Hyfiex EDM, iRace, Neoniti, One Shape Protaper Next, Reciproc, Revo-S and Wave One Gold in terms of resistance to cyclic fatigue and length of broken fragments. Material and methods: 300 new size 25 files of the systems studied were selected (n=30). The instruments were mechanized with a X-Smart Plus endo motor at the speed and torque recommended by the manufacturer, holding the instruments steady with a clamping mechanism, with passive adjustment and without pressure in a stainless-steel block. The time was calculated in seconds until fracture. The number of fatigue cycles was calculated as (Resistance (s) x Speed)/60. The separated fragment lengths were measured with a digital Vernier calliper. A statistical analysis was performed with the SPSS 18 programme at a 95% confidence level, using the Levene's Test to compare variances, the Welch's Test to compare means, and the Games-Howell's Test to reveal differences between groups. Results: The Levene's Test showed no equal variances (P<0.05). The Welch's Test (P<0.05) showed significant differences in cyclic fatigue and separated fragment lengths. The Games-Howell test (P<0.05) exhibited significant differences in multiple comparisons (P<0.05). Conclusions: The systems with CM-Wire (Hyflex EDM and Neoniti) were superior in resistance to the other systems for cyclic fatigue. For separated fragment lengths, F360 (conventional NiTi) and Reciproc (M-Wire) were better significantly better in terms of resistance.
引用
收藏
页码:28 / 36
页数:9
相关论文
共 50 条
  • [31] Effect of Temperature on the Cyclic Fatigue Resistance and Phase Transformation Behavior of Three Different NiTi Endodontic Instruments
    Yigt, Esra trem
    Cetinkaya, Trem
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (01)
  • [32] Influence of NiTi Wire Diameter on Cyclic and Torsional Fatigue Resistance of Different Heat-Treated Endodontic Instruments
    Pedulla, Eugenio
    Canova, Francesco Saverio
    La Rosa, Giusy Rita Maria
    Naaman, Alfred
    Diemer, Franck
    Generali, Luigi
    Nehme, Walid
    [J]. MATERIALS, 2022, 15 (19)
  • [33] Cyclic fatigue resistance of WaveOne Gold instruments with different amplitudes of axial movement: a dynamic study
    Alves, Viviane
    Carvalho, Rafael
    Silva, Rene
    Soares, Adriana de Jesus
    Frozoni, Marcos
    [J]. GIORNALE ITALIANO DI ENDODONZIA, 2022, 36 (01): : 10 - 16
  • [34] The Cyclic Fatigue Resistance of Different Lengths of CM Gold Wire and CM Blue Wire NiTi Alloy Endodontic Rotary Files: An In Vitro Study
    Faus-Matoses, Vicente
    Faus-Llacer, Vicente
    Ruiz-Sanchez, Celia
    Prats Gallego, Sofia
    Zubizarreta-Macho, Alvaro
    Solano-Mendoza, Beatriz
    Biedma, Benjamin Martin
    Faus-Matoses, Ignacio
    [J]. APPLIED SCIENCES-BASEL, 2023, 13 (07):
  • [35] Does the type of endodontic access influence in the cyclic fatigue resistance of reciprocating instruments?
    Emmanuel João Nogueira Leal Silva
    Rafael Santos Attademo
    Marcia Cristina Dutra da Silva
    Karem Paula Pinto
    Henrique dos Santos Antunes
    Victor Talarico Leal Vieira
    [J]. Clinical Oral Investigations, 2021, 25 : 3691 - 3698
  • [36] The Effect of Temperature on Cyclic Fatigue of Nickel-titanium Rotary Endodontic Instruments
    Dosanjh, Amandeep
    Paurazas, Susan
    Askar, Mazin
    [J]. JOURNAL OF ENDODONTICS, 2017, 43 (05) : 823 - 826
  • [37] Cyclic fatigue of endodontic nickel titanium rotary instruments: Static and dynamic tests
    Li, UM
    Lee, BS
    Shih, CT
    Lan, WH
    Lin, CP
    [J]. JOURNAL OF ENDODONTICS, 2002, 28 (06) : 448 - 451
  • [38] Correlation between cyclic fatigue and the bending properties of nickel titanium endodontic instruments
    Pia, Jindan
    Miyara, Kana
    Ebihara, Arata
    Nomura, Naoyuki
    Hanawa, Takao
    Suda, Hideaki
    [J]. DENTAL MATERIALS JOURNAL, 2014, 33 (04) : 539 - 544
  • [39] Does the type of endodontic access influence in the cyclic fatigue resistance of reciprocating instruments?
    Nogueira Leal Silva, Emmanuel Joao
    Attademo, Rafael Santos
    Dutra da Silva, Marcia Cristina
    Pinto, Karem Paula
    Antunes, Henrique dos Santos
    Leal Vieira, Victor Talarico
    [J]. CLINICAL ORAL INVESTIGATIONS, 2021, 25 (06) : 3691 - 3698
  • [40] Effect of Rotational Speed on the Resistance of NiTi Alloy Endodontic Rotary Files to Cyclic Fatigue-An In Vitro Study
    Faus-Matoses, Vicente
    Faus-Llacer, Vicente
    Ruiz-Sanchez, Celia
    Jaramillo-Vasconez, Sharon
    Faus-Matoses, Ignacio
    Martin-Biedma, Benjamin
    Zubizarreta-Macho, Alvaro
    [J]. JOURNAL OF CLINICAL MEDICINE, 2022, 11 (11)