Cyclic fatigue resistance of WaveOne Gold instruments with different amplitudes of axial movement: a dynamic study

被引:0
|
作者
Alves, Viviane [1 ]
Carvalho, Rafael [1 ]
Silva, Rene [2 ]
Soares, Adriana de Jesus [3 ]
Frozoni, Marcos [1 ]
机构
[1] Sao Leopoldo Mandic Sch Dent, Campinas, Brazil
[2] Univ Fed Vicosa, Vicosa, MG, Brazil
[3] Univ Estadual Campinas, Campinas, Brazil
来源
GIORNALE ITALIANO DI ENDODONZIA | 2022年 / 36卷 / 01期
关键词
cycling fatigue; fracture; reciprocating; WaveOne; NICKEL-TITANIUM INSTRUMENTS; TORSIONAL RESISTANCE; ROTARY INSTRUMENTS; RECIPROC; FILES; MOTION;
D O I
10.32067/GIE.2021.35.01.34
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim: Single-file reciprocating instruments were launched, aiming to diminish the risk of instrument separation. It operates in an in-and-out dynamic with an amplitude of approximately 3-4 mm. This study evaluated the impact of different amplitudes of axial movement in the dynamic cyclic fatigue resistance of the WaveOne Gold reciprocating file. Materials and Methods: Forty-five WaveOne Gold 25.07 were divided into three different groups (n=15) according to the amplitude of axial movement of choice. In the G-2.5 group, the instruments were used in an in-and-out amplitude of 2.5 mm; in the G-5 group, the instruments were used in an in-and-out amplitude of 5.0 mm, and in the G-7.5 group, the instruments were used in an in-and-out amplitude of 7.5 mm. All instruments were tested in a simulated canal notched in metallic block, with 9.04 mm in its cervical portion, 13.3 mm in its apical portion, and curvature of 2.5 mm and 69 degrees. The axial movement was applied with a back-and-forth speed of 2.5 mm/s during the dynamic cyclic fatigue test until the fracture was noticed. All of the tests were performed at a controlled temperature of 36 +/- 1 degrees C and under oil lubrication. Then, the time to fracture (TTF/s), the number of cycles to fracture (NCF), and the length of the separated fragments were registered. The level of significance was set at 5%. Results: The TTF/s was 11.40 +/- 9.83, 15.00 +/- 7.46 and 22.33 +/- 8.76 for G-2.5, G-5, and G-7.5, respectively. The NCF was 57.00 +/- 49.13, 75.00 +/- 37.32, and 111.67 +/- 43.82 for G-2.5, G-5, and G-7.5, respectively. For both TTF/s and NCF, G-7.5 was higher than G-2.5, and G-5.0 was similar to both groups (P<.05). The lengths of the fragments were 10.27 +/- 1.07, 10.37 +/- 0.66, and 10.58 +/- 0.77 for G-2.5, G-5, and G-7.5, respectively. There was no difference among the groups in regards to the length of the fragments (P>.05). The SEM images showed characteristics related to cyclic fatigue. Conclusions: It can be concluded that in-and-out movements affect the dynamic cyclic fatigue resistance of the WaveOne Gold reciprocating instrument. Appropriate pecking motions in the root canals are recommended to prevent the breakage of NiTi reciprocating instruments.
引用
收藏
页码:10 / 16
页数:7
相关论文
共 50 条
  • [1] Comparison of Cyclic Fatigue Resistance of WaveOne and WaveOne Gold Small, Primary, and Large Instruments
    Achguzel, Mehmet
    Capar, Ismail Davut
    [J]. JOURNAL OF ENDODONTICS, 2017, 43 (04) : 623 - 627
  • [2] Cyclic Fatigue Resistance of Reciproc, WaveOne, and WaveOne Gold Nickel-Titanium Instruments
    Ozyurek, Taha
    [J]. JOURNAL OF ENDODONTICS, 2016, 42 (10) : 1536 - 1539
  • [3] Cyclic Fatigue Resistance of Reciproc Blue, Reciproc, and WaveOne Gold Reciprocating Instruments
    Keskin, Cangul
    Ivan, Ugur
    Demiral, Murat
    Keles, Ali
    [J]. JOURNAL OF ENDODONTICS, 2017, 43 (08) : 1360 - 1363
  • [4] Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
    Oh, Soram
    Kum, Kee-Yeon
    Kim, Hyun-Jung
    Moon, Su-Young
    Kim, Hyeon-Cheol
    Chaniotis, Antonis
    Perinpanayagam, Hiran
    Pedulla, Eugenio
    Chang, Seok Woo
    [J]. JOURNAL OF DENTAL SCIENCES, 2020, 15 (04) : 472 - 478
  • [5] Cyclic fatigue resistance of OneShape and WaveOne instruments using different angles of curvature
    Elsaka, Shaymaa E.
    Elnaghy, Amr M.
    [J]. DENTAL MATERIALS JOURNAL, 2015, 34 (03) : 358 - 363
  • [6] Assessment of cyclic fatigue resistance of Protaper Next and WaveOne Gold in different kinematics
    Amira Galal Ismail
    Manar Galal
    Nehal Nabil Roshdy
    [J]. Bulletin of the National Research Centre, 44 (1)
  • [7] Bending Resistance and Dynamic and Static Cyclic Fatigue Life of Reciproc and WaveOne Large Instruments
    De-Deus, Gustavo
    Leal Vieira, Victor Talarico
    Nogueira da Silva, Emmanuel Joao
    Lopes, Helio
    Elias, Carlos Nelson
    Moreira, Edson Jorge
    [J]. JOURNAL OF ENDODONTICS, 2014, 40 (04) : 575 - 579
  • [8] Effect of sodium hypochlorite and saline on cyclic fatigue resistance of WaveOne Gold and Reciproc reciprocating instruments
    Elnaghy, A. M.
    Elsaka, S. E.
    [J]. INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (10) : 991 - 998
  • [9] Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments
    Keskin, Cangul
    Inan, Ugur
    Demiral, Murat
    Keles, Ali
    [J]. CLINICAL ORAL INVESTIGATIONS, 2018, 22 (09) : 3007 - 3012
  • [10] Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments
    Cangül Keskin
    Uğur İnan
    Murat Demiral
    Ali Keleş
    [J]. Clinical Oral Investigations, 2018, 22 : 3007 - 3012