Effect of Radiofrequency Use on Hip Arthroscopy Irrigation Fluid Temperature

被引:24
|
作者
McCormick, Frank [1 ]
Alpaugh, Kyle [2 ]
Nwachukwu, Benedict U. [2 ]
Xu, Siyan [3 ]
Martin, Scott D. [2 ]
机构
[1] Midwest Orthoped Rush Sports Med, Rush Sports Med Fellowship Program, Chicago, IL USA
[2] Brigham & Womens Hosp, Dept Orthoped Surg, Boston, MA 02115 USA
[3] Boston Univ, Sch Med, Dept Biostat, Boston, MA 02118 USA
关键词
THERMAL CHONDROPLASTY; SHOULDER ARTHROSCOPY; ENERGY; BIPOLAR; CHONDROLYSIS; SURGERY;
D O I
10.1016/j.arthro.2012.10.001
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: The purpose of this study was to determine operating parameters for joint fluid lavage using radiofrequency (RF) in maintaining intra-articular temperatures < 50 degrees C in the hip joint and to then quantify the influence of flow rate on maintaining safe intra-articular temperatures. Methods: Fiberoptic intra-articular thermometers at radial distance intervals of 1, 2, 5, and 10 mm, spanning cross-sectional areas of 3.14, 12.56, 78.5, and 314.1 mm(2), respectively, from the RF probe were used in 3 human hip cadaveric specimens at room temperature, with 9 trials per variable, using a 3-portal technique with a capsule release. Using a new Dyonics RF System continuously at the superior capsulolabral junction for 90 seconds, continuous temperatures were recorded at 50 mm Hg inflow and variable outflow intervals: no flow and 5, 15, and 30 seconds. Lavages were 1 second in duration on suction. Statistical comparison was through multivariate regression analysis and a logistic model. Results: Temperatures reached >= 50 degrees C at 5-mm radial distance from the probe in all but the 5-second pulse lavage group. Elevated temperatures were reached within 1 to 2 seconds locally (1- to 2-mm radial distance) and at a radial distance of 5 mm in the 15-, 30-, and 0-second lavage groups. Logistic regression revealed a reduction in the odds that temperatures >= 50 degrees C will occur as flow frequency increases every 30 (odds ratio = 0.68, P = .086); 15 (odds ratio = -1.22, P = .0067); and 5 (odds ratio = -4.26, P < .0001) seconds. Conclusions: Increasing-interval pulsed irrigation is effective in maintaining intra-articular temperature profiles below 50 degrees C during use of continuous RF ablation. Five-second-interval pulsed lavage is the longest flow interval identified during which fluid 5 mm radially from the RF device never reached temperatures > 50 degrees C. Clinical Relevance: Clinical guidelines for using the RF ablation include: meticulous technique, intermittent use, good inflow and outflow, and pulsed lavage at frequent intervals.
引用
收藏
页码:336 / 342
页数:7
相关论文
共 50 条
  • [31] The effect of irrigation fluid temperature on core body temperature in arthroscopic shoulder surgery
    Board, Timothy N.
    Srinivasan, Makram S.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (05) : 531 - 533
  • [32] The effect of irrigation fluid temperature on core body temperature in arthroscopic shoulder surgery
    Timothy N. Board
    Makram S. Srinivasan
    Archives of Orthopaedic and Trauma Surgery, 2008, 128 : 531 - 533
  • [33] Effect of Bipolar Radiofrequency Energy on Canine Stifle Joint Fluid Temperature
    Aoki, Koji
    Waldner, Cheryl L.
    Sathya, Suresh
    Shmon, Cindy
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2018, 31 (03) : 188 - 193
  • [34] Warm irrigation fluid does not raise the subacromial temperature to harmful levels while using radiofrequency device
    Gereli, Arel
    Kocaoglu, Baris
    Guven, Osman
    Turkmen, Metin
    INTERNATIONAL JOURNAL OF SHOULDER SURGERY, 2015, 9 (03): : 99 - +
  • [35] Effect of Saline Irrigation Flow Rate on Temperature Profile during Cooled Radiofrequency Ablation
    Wai Shun Wong
    Brian A. VanderBrink BS*
    Richard E. Riley MS
    Mark Pomeranz MS
    Mark S. Link
    Munther K. Homoud
    N.A. Mark Estes III
    Paul J. Wang
    Journal of Interventional Cardiac Electrophysiology, 2000, 4 (1) : 321 - 326
  • [36] Effect of saline irrigation flow rate on temperature profile during cooled radiofrequency ablation
    Wong, WS
    VanderBrink, BA
    Riley, RE
    Pomeranz, M
    Link, MS
    Homoud, MK
    Estes, NAM
    Wang, PJ
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2000, 4 (01) : 321 - 326
  • [37] Extravasated Fluid in Hip Arthroscopy and Pain: Is Quadratus Lumborum Block the Answer?
    Ben-David, Bruce
    La Colla, Luca
    ANESTHESIA AND ANALGESIA, 2017, 125 (01): : 364 - 364
  • [38] Intra-abdominal fluid after hip arthroscopy : A case report
    Jaekers, Joris
    Van Melkebeek, Jan
    Van Cleemput, Marc
    Van Riet, Roger
    ACTA ORTHOPAEDICA BELGICA, 2012, 78 (06): : 800 - 803
  • [39] Avoiding fluid loss from an unused portal during hip arthroscopy
    van Winterswijk, P.
    Wilson, M.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2016, 98 (08) : 590 - 590
  • [40] Oxygen embolism after hydrogen peroxide irrigation during hip arthroscopy: a case report
    Zhengwu Peng
    Hui Li
    Ziqin Cao
    Wenchao Zhang
    Hongxin Li
    Ruping Dai
    Lei Liu
    Xinzhan Mao
    Daniel M. George
    Tianlong Huang
    BMC Musculoskeletal Disorders, 21