Effects of a lateral row anchor position on the suture holding strength of a double-row knotless fixation in rotator cuff repair

被引:5
|
作者
Acan, Ahmet Emrah [1 ]
Hapa, Onur [2 ]
Horoz, Levent [2 ]
Kara, Aylin [3 ]
Havitcioglu, Hasan [2 ]
机构
[1] Mugla Sitki Kocman Univ, Training & Res Hosp, Dept Orthoped & Traumatol, Mugla, Turkey
[2] Dokuz Eylul Univ, Med Fac, Dept Orthoped & Traumatol, Izmir, Turkey
[3] Dokuz Eylul Univ, Dept Biomech, Izmir, Turkey
关键词
Biomechanics; double-row; rotator cuff; FAILURE; MODEL; BONE;
D O I
10.5606/ehc.2018.57519
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: This study aims to evaluate the effects of anchor positions on the suture holding strength of a double-row knotless fixation in rotator cuff repair. Materials and methods: Four different double-row fixation techniques were assessed. In group 1, a 15-mm-wide mattress suture was fixed using a knotless lateral row anchor, horizontal to the shaft. In group 2, the medial sutures were fixed with a 5-mm more lateral anchor that was placed at 45 degrees to the long axis of the humeral shaft. In group 3, different from group 2, medial sutures were fixed with a 30-mm mattress suture width. In group 4, the mattress sutures coming from the medial row anchors were fixed to the 10-mm more lateral row, vertical to the long axis of the humeral shaft. The specimens were cyclically loaded from 10 N to 30 N at 0.5 Hz for 50 cycles, and then loaded to failure. Results: Group 4 had higher cyclic elongation values than group 1 (p=0.021) and group 3 (p=0.006). Group 1 had lower maximum load value than group 3 (p=0.011). Most of the specimens failed with suture ruptures. Unlike the other groups, none of the specimens in group 4 failed via a suture pull through the lateral anchor. Conclusion: A horizontal lateral row anchor positioned closer to the medial anchor resulted in less cyclic elongation when compared to a more vertically positioned lateral row anchor. The vertical or oblique positioning of the lateral row anchor did not result in any increase in the failure load value; however, the vertical placement prevented a suture pull through the lateral row anchor.
引用
收藏
页码:46 / 51
页数:6
相关论文
共 50 条
  • [31] Arthroscopic hybrid double-row rotator cuff repair
    Kim, Kyung-Cheon
    Rhee, Kwang-Jin
    Shin, Hyun-Dae
    Kim, Young-Mo
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2007, 15 (06) : 794 - 799
  • [32] A biomechanical comparison of single and double-row fixation in arthroscopic rotator cuff repair
    Smith, Christopher D.
    Alexander, Susan
    Hill, Adam M.
    Huijsmans, Pol E.
    Bull, Anthony M. J.
    Amis, Andrew A.
    De Beer, Joe F.
    Wallace, Andrew L.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (11): : 2425 - 2431
  • [33] How to Maximize Suture Tension in Double-Row Suture-Bridge Rotator Cuff Repair?
    Tanpowpong, Thanathep
    Itthipanichpong, Thun
    Limskul, Danaithep
    ARTHROSCOPY TECHNIQUES, 2021, 10 (10): : E2207 - E2212
  • [34] The low-profile Roman bridge technique for knotless double-row repair of the rotator cuff
    Umile Giuseppe Longo
    Francesco Franceschi
    Filippo Spiezia
    Andrea Marinozzi
    Nicola Maffulli
    Vincenzo Denaro
    Archives of Orthopaedic and Trauma Surgery, 2011, 131 : 357 - 361
  • [35] The Mattress-Tension-Band Technique A Knotless Double-Row Arthroscopic Rotator Cuff Repair
    Boileau, Pascal
    Brassart, Nicolas
    Roussanne, Yannick
    Scobercea, Razvan
    TECHNIQUES IN SHOULDER AND ELBOW SURGERY, 2009, 10 (01): : 22 - 25
  • [36] The low-profile Roman bridge technique for knotless double-row repair of the rotator cuff
    Longo, Umile Giuseppe
    Franceschi, Francesco
    Spiezia, Filippo
    Marinozzi, Andrea
    Maffulli, Nicola
    Denaro, Vincenzo
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (03) : 357 - 361
  • [37] Biomechanical Properties of Double-Row Transosseous Rotator Cuff Repair Combined With the Cinch Stitch in the Lateral Row
    Ito, Kentaro
    Sakaguchi, Katsunobu
    Sekihata, Hirosi
    Sugita, Naoki
    Kadono, Yuho
    ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2021, 9 (05)
  • [38] Outcome of open massive rotator cuff repairs with double-row suture knotless anchors: case series
    Tara M. Connelly
    Aileen Shaw
    Paul O’Grady
    International Orthopaedics, 2015, 39 : 1109 - 1114
  • [39] A NEW KNOTLESS SUTURE STAPLE DOUBLE ROW CONSTRUCT FOR MEDIAL REPAIR OF ROTATOR CUFF TEARS
    Schlegel, Theodore F.
    Otarodifard, Karimdad Amir
    Hawkins, Richard J.
    Vangsness, Thomas, Jr.
    Bui, Christopher
    McGarry, Michelle H.
    Lee, Thay Q.
    PROCEEDINGS OF THE 4TH FRONTIERS IN BIOMEDICAL DEVICES CONFERENCE AND EXPOSITION - 2009, 2009, : 17 - 18
  • [40] Outcome of open massive rotator cuff repairs with double-row suture knotless anchors: case series
    Connelly, Tara M.
    Shaw, Aileen
    O'Grady, Paul
    INTERNATIONAL ORTHOPAEDICS, 2015, 39 (06) : 1109 - 1114