MICROBIOLOGICAL CONTAMINATION OF A HEMODIALYSIS CENTER WATER DISTRIBUTION SYSTEM

被引:30
|
作者
Montanari, Lilian Bueno [1 ]
Sartori, Flavio Garcia [1 ]
de Oliveira Cardoso, Miguel Jorge [1 ]
Varo, Samuel Dutra [1 ]
Pires, Regina Helena [1 ]
Fujimura Leite, Clarice Queico [2 ]
Prince, Karina [2 ]
Gomes Martins, Carlos Henrique [1 ]
机构
[1] Univ Franca, Lab Res Appl Microbiol, Franca, SP, Brazil
[2] State Univ Sao Paulo, Dept Biol Sci, Araraquara, SP, Brazil
关键词
Dialysis; Water; Microbiological contamination; Water treatment; Dialysis fluid quality; Bacterial count; PYROGENIC REACTIONS; DIALYSIS WATER; MYCOBACTERIA; IDENTIFICATION; QUALITY; INFECTIONS; BACTERIA; FLUID;
D O I
10.1590/S0036-46652009000100007
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The microbiological monitoring of the water used for hemodialysis is extremely important, especially because of the debilitated immune system of patients suffering from chronic renal insufficiency. To investigate the occurrence and species diversity of bacteria in waters, water samples were collected monthly from a hemodialysis center in upstate Sao Paulo and tap water samples at the terminal sites of the distribution system was sampled repeatedly (22 times) at each of five points in the distribution system; a further 36 samples were taken from cannulae in 19 hemodialysis machines that were ready for the next patient, four samples from the reuse system and 13 from the water storage system. To identify bacteria, samples were filtered through 0.22 mu m-pore membranes; for mycobacteria, 0.45 mu m pores were used. Conventional microbiological and molecular methods were used in the analysis. Bacteria were isolated from the distribution system (128 isolates), kidney machine water (43) and reuse system (3). Among these isolates, 32 were Gram-positive rods, 120 Gram-negative rods, 20 Gram-positive cocci and 11 mycobacteria. We propose the continual monitoring of the water supplies in hemodialysis centers and the adoption of effective prophylactic measures that minimize the exposure of these immunodeficient patients to contaminated sources of water.
引用
收藏
页码:37 / 43
页数:7
相关论文
共 50 条
  • [21] Control of Legionella contamination in a hospital water distribution system by monochloramine
    Marchesi, Isabella
    Cencetti, Stefano
    Marchegiano, Patrizia
    Frezza, Giuseppina
    Borella, Paola
    Bargellini, Annalisa
    AMERICAN JOURNAL OF INFECTION CONTROL, 2012, 40 (03) : 279 - 281
  • [22] A contamination source identification model for water distribution system security
    Preis, A.
    Ostfeld, A.
    ENGINEERING OPTIMIZATION, 2007, 39 (08) : 941 - 951
  • [23] Changes of microbiological parameters of water in domestic distribution system in terms of water supply safety
    Zywiec, Jakub
    Tchorzewska-Cieslak, Barbara
    Papciak, Dorota
    Domon, Andzelika
    DESALINATION AND WATER TREATMENT, 2021, 226 : 37 - 51
  • [24] Microbiological quality assurance of purified water by ozonization of storage and distribution system
    Florjanic, M.
    Kristl, J.
    DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 2006, 32 (10) : 1113 - 1121
  • [25] INOVATIVE FILTRATION OF THE MICROBIOLOGICAL CONTAMINATION WATER USING NANOTEXTILE
    Dosek, M.
    Lev, J.
    Cerny, M.
    Kalhotka, L.
    MENDELNET 2012, 2012, : 1048 - 1054
  • [26] Sources of microbiological contamination in sachet water from Ghana
    Aslan, Asli
    Rochani, Haresh
    Oyibo, Oghenekpaobor
    Dotherow, J. Edward
    Anderson, Kendall W.
    Beslin, Christina
    Chapman, Asheley
    Thomas, Brittny
    Brown, Alicia
    Ampofo-Yeboah, Akwasi
    Afriyie-Gyawu, Evans
    JOURNAL OF WATER SANITATION AND HYGIENE FOR DEVELOPMENT, 2020, 10 (02) : 202 - 208
  • [27] Reflection about the hemodialysis water microbiological quality in Brazil
    Carvalho, Gabriela Correa
    Dua, Kamal
    Gupta, Gaurav
    Bugno, Adriana
    Andreoli Pinto, Terezinha de Jesus
    BRAZILIAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2022, 58
  • [28] RISK MANAGEMENT METHODS IN HOT WATER DISTRIBUTION SYSTEMS TO NEGATE HEALTH RISKS FROM MICROBIOLOGICAL CONTAMINATION
    Ocipova, Daniela
    Vranayova, Zuzana
    Sikula, Ondrej
    ENVIRONMENTAL ENGINEERING, VOLS 1-3, 2011, : 638 - 642
  • [29] Microbiological monitoring of ultrapure dialysis fluid in a hemodialysis center in Alexandria, Egypt
    Abbass, Aleya A.
    El-Koraie, Ahmed F.
    Hazzah, Walaa A.
    Omran, Eman A.
    Mahgoub, Mohammed A.
    ALEXANDRIA JOURNAL OF MEDICINE, 2018, 54 (04) : 523 - 527
  • [30] CONTAMINATION OF WATER DISTRIBUTION SYSTEMS
    Grayman, Walter M.
    INTERNATIONAL SEMINAR ON NUCLEAR WAR AND PLANETARY EMERGENCIES: 45TH SESSION, 2013, : 389 - 398