MEAN GLANDULAR DOSE IN SIX DIGITAL MAMMOGRAPHY SERVICES IN SANTIAGO, CHILE: PRELIMINARY REFERENCE LEVELS

被引:3
|
作者
Leyton, Fernando [1 ,2 ]
Nogueira, Maria Do Socorro [3 ]
Dantas, Marcelino [4 ]
Paz Duran, Maria [5 ]
Ubeda, Carlos [6 ,7 ]
机构
[1] Diego Portales Univ, Fac Hlth & Odontol, Santiago, Chile
[2] Tarapaca Univ, Study Ctr Radiol Sci CECRAD, Arica, Chile
[3] Nucl Technol Dev Ctr CDTN CNEN, Belo Horizonte, MG, Brazil
[4] INB, Caldas, Brazil
[5] Clin Alemana, Santiago, Chile
[6] Tarapaca Univ, Fac Hlth Sci, Study Ctr Radiol Sci CECRAD, Med Technol Dept, Arica, Chile
[7] Tarapaca Univ, CHIDE, Arica, Chile
关键词
SCREEN-FILM MAMMOGRAPHY; QUALITY; PARAMETERS;
D O I
10.1093/rpd/ncv092
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this paper was to estimate mean glandular dose levels (D-G) in six digital mammography systems in Santiago, Chile, and to propose preliminary reference levels to execute mammography in Chile. The study was carried out assessing two direct digital systems and four computer-based radiography (CR) systems. Estimates of D-G were calculated for different thicknesses of polymethyl methacrylate according to the quality control protocol in digital mammography of the Spanish Society of Medical Physics and NHSBSP Equipment Report 0604 Version 3. D-G values ranged between 0.64 and 7.26 mGy for a range of 20-to 70-mm thickness, respectively. Thirty-six per cent of D-G was higher than the acceptable dose level and 100 % of D-G was higher than the desirable level. It is therefore necessary to optimise doses. The initial proposal to establish dose reference levels for D-G would range between 0.90 and 6.40 mGy for a thickness range of 20 to 70 mm.
引用
收藏
页码:115 / 120
页数:6
相关论文
共 50 条
  • [1] ASSESSMENT OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY
    FAULKNER, K
    LAW, J
    ROBSON, KJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (812): : 877 - 881
  • [2] Historical trends in image quality and mean glandular dose in digital mammography
    Mackenzie, Alistair
    Khan, Riyad
    Young, Kenneth C.
    [J]. 15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
  • [3] Dosimetric evaluation of the mean glandular dose for mammography in Korean women: A preliminary report
    Oh, KK
    Hur, J
    Kim, EK
    Choo, SS
    [J]. YONSEI MEDICAL JOURNAL, 2003, 44 (05) : 863 - 868
  • [4] MODEL FOR COMPUTING TUNGSTEN TUBE SPECTRA AND MEAN GLANDULAR DOSE FOR DIGITAL MAMMOGRAPHY
    SPIVEY, BA
    ANDRE, MP
    STOLLER, M
    TRAN, J
    [J]. RADIOLOGY, 1995, 197 : 291 - 291
  • [5] Evaluation of depth dose and glandular dose for digital mammography
    Tsai, H. Y.
    Chong, N. S.
    Ho, Y. J.
    Tyan, Y. S.
    [J]. RADIATION MEASUREMENTS, 2010, 45 (3-6) : 726 - 728
  • [6] Assessment of Mean Glandular Dose and Entrance Surface Dose in Mammography
    Abdallah, Yousif M.
    Abuhadi, Nouf H.
    Aldousari, Hanan A.
    [J]. JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2021, 9 (02): : 8 - 11
  • [7] Clinical image quality assessment and mean glandular dose for full field digital mammography
    Nassar, Joyce
    Rizk, Chadia
    Fares, Georges
    Tohme, Carla
    Braidy, Chady
    Farah, Jad
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (01)
  • [8] Decreased breast compression using digital mammography at screen/film mean glandular dose
    Kimme-Smith, C
    Yee, TY
    Bassett, LW
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 51 - 55
  • [9] Impact of age and breast thickness on mean glandular dose of standard digital mammography and digital breast tomosynthesis
    Norsuddin, Norhashimah Mohd
    Sin, Justine Go Mei
    Ravintaran, Rathieswari
    Arasaratnam, Shantini
    Karim, Muhammad Khalis Abdul
    [J]. APPLIED RADIATION AND ISOTOPES, 2023, 192
  • [10] AN INVESTIGATION INTO VARIATIONS IN THE ESTIMATION OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY
    FAULKNER, K
    CRANLEY, K
    [J]. RADIATION PROTECTION DOSIMETRY, 1995, 57 (1-4) : 405 - 407