Malfunction of the da Vinci robotic system in urology

被引:15
|
作者
Chen, Cheng-Che [1 ]
Ou, Yen-Chuan [1 ,2 ,3 ]
Yang, Cheng-Kuang [1 ]
Chiu, Kun-Yuan [1 ]
Wang, Shian-Shiang [1 ]
Su, Chung-Kuang [1 ]
Ho, Hao-Chung [1 ]
Cheng, Chen-Li [1 ]
Chen, Chuan-Shu [1 ]
Lee, Jian-Ri [1 ]
Chen, Wen-Min [1 ]
机构
[1] Taichung Vet Gen Hosp, Div Urol, Dept Surg, Taichung 40705, Taiwan
[2] Chung Sang Med Univ, Inst Med, Taichung, Taiwan
[3] Natl Chi Nan Univ, Grad Inst Biomed & Biomed Technol, Dept Appl Chem, Nantou, Taiwan
关键词
da Vinci; malfunction; robot; urology; RADICAL PROSTATECTOMIES;
D O I
10.1111/j.1442-2042.2012.03010.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To analyze the incidence of malfunction of the da Vinci robotic system in a single center and to provide potential solutions. Methods: A total of 400 patients underwent da Vinci robotic urological surgery at Taichung Veterans General Hospital in Taichung, Taiwan, from December 2005 to April 2011. Episodes of malfunction of the robotic system were analyzed by period of operation, type of procedure, type of malfunction and management of the event. Results: Overall, 14 cases of malfunction occurred (3.5% of the entire series). Among them, five (1.25%) occurred before the surgery and nine (2.25%) intraoperatively. Operative procedures included radical prostatectomy, bilateral pelvic lymph node dissection, dismembered pyeloplasty, partial nephrectomy, nephroureterectomy, and radical and partial cystectomies. Areas of malfunctions included the robotic arm system and joint (11/14), optical system (1/14), power system and connector (1/14), endoscopic instrument (1/14), and software (1/14). In 10 cases, the failure was recoverable, whereas in four cases there was a critical failure, requiring a conversion to standard laparoscopy in three of them, and the rescheduling of the surgery in one case. Conclusions: The da Vinci robotic system is extremely reliable for use in urology. Malfunction is rare and the risk of critical failure is very low. Managing mechanical failure before or during the surgery is the key to maintaining the safety of patients undergoing robotic surgical procedures.
引用
收藏
页码:736 / 740
页数:5
相关论文
共 50 条
  • [31] Technique and perioperative outcomes of single-port robotic surgery using the da Vinci SP platform in urology
    Kim, Kwang Hyun
    Ahn, Hyun Kyu
    Kim, Myong
    Yoon, Hana
    ASIAN JOURNAL OF SURGERY, 2023, 46 (01) : 472 - 477
  • [32] The da Vinci robotic system for general surgical applications: a critical interim appraisal
    Bodner, J
    Augustin, F
    Wykypiel, H
    Fish, J
    Muehlmann, G
    Wetscher, G
    Schmid, T
    SWISS MEDICAL WEEKLY, 2005, 135 (45-46) : 674 - 678
  • [33] Establishment and Application of a Novel Difficulty Scoring System for da Vinci Robotic Pancreatoduodenectomy
    Sun, Hongfa
    Sun, Chuandong
    Zhang, Bingyuan
    Ma, Kai
    Wu, Zehua
    Visser, Brendan C.
    Han, Bing
    FRONTIERS IN SURGERY, 2022, 9
  • [34] The patterns and costs of the da Vinci robotic surgery system in a large academic institution
    Prewitt R.
    Bochkarev V.
    McBride C.L.
    Kinney S.
    Oleynikov D.
    Journal of Robotic Surgery, 2008, 2 (1) : 17 - 20
  • [35] Laparoscopic appendicovesicostomy (Mitrofanoff procedure) in a child using the da Vinci robotic system
    Pedraza, R
    Weiser, A
    Franco, I
    JOURNAL OF UROLOGY, 2004, 171 (04): : 1652 - 1653
  • [36] Advantages of using robotic Da Vinci® system for unilateral adrenalectomy:: early results
    Brunaud, L
    Bresler, L
    Ayav, A
    Tretou, S
    Cormier, L
    Klein, M
    Boissel, P
    ANNALES DE CHIRURGIE, 2003, 128 (08): : 530 - 535
  • [37] RoSS: Virtual reality robotic surgical simulator for the da Vinci surgical system
    Baheti, Ankur
    Seshadri, Sridhar
    Kumar, Amrish
    Srimathveeravalli, Govindarajan
    Kesavadas, Thenkurussi
    Guru, Khurshid
    SYMPOSIUM ON HAPTICS INTERFACES FOR VIRTUAL ENVIRONMENT AND TELEOPERATOR SYSTEMS 2008, PROCEEDINGS, 2008, : 479 - +
  • [38] Minimally invasive mitral valve repair using the da Vinci robotic system
    Tatooles, AJ
    Pappas, PS
    Gordon, PJ
    Slaughter, MS
    ANNALS OF THORACIC SURGERY, 2004, 77 (06): : 1978 - 1984
  • [39] Robotic-assisted simple prostatectomy using the Da Vinci sp system
    Kim, B. H.
    Byun, H. J.
    Shin, T. J.
    Jung, W. H.
    Ha, J. Y.
    EUROPEAN UROLOGY, 2023, 83