Malfunction of the da Vinci robotic system in urology

被引:15
|
作者
Chen, Cheng-Che [1 ]
Ou, Yen-Chuan [1 ,2 ,3 ]
Yang, Cheng-Kuang [1 ]
Chiu, Kun-Yuan [1 ]
Wang, Shian-Shiang [1 ]
Su, Chung-Kuang [1 ]
Ho, Hao-Chung [1 ]
Cheng, Chen-Li [1 ]
Chen, Chuan-Shu [1 ]
Lee, Jian-Ri [1 ]
Chen, Wen-Min [1 ]
机构
[1] Taichung Vet Gen Hosp, Div Urol, Dept Surg, Taichung 40705, Taiwan
[2] Chung Sang Med Univ, Inst Med, Taichung, Taiwan
[3] Natl Chi Nan Univ, Grad Inst Biomed & Biomed Technol, Dept Appl Chem, Nantou, Taiwan
关键词
da Vinci; malfunction; robot; urology; RADICAL PROSTATECTOMIES;
D O I
10.1111/j.1442-2042.2012.03010.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To analyze the incidence of malfunction of the da Vinci robotic system in a single center and to provide potential solutions. Methods: A total of 400 patients underwent da Vinci robotic urological surgery at Taichung Veterans General Hospital in Taichung, Taiwan, from December 2005 to April 2011. Episodes of malfunction of the robotic system were analyzed by period of operation, type of procedure, type of malfunction and management of the event. Results: Overall, 14 cases of malfunction occurred (3.5% of the entire series). Among them, five (1.25%) occurred before the surgery and nine (2.25%) intraoperatively. Operative procedures included radical prostatectomy, bilateral pelvic lymph node dissection, dismembered pyeloplasty, partial nephrectomy, nephroureterectomy, and radical and partial cystectomies. Areas of malfunctions included the robotic arm system and joint (11/14), optical system (1/14), power system and connector (1/14), endoscopic instrument (1/14), and software (1/14). In 10 cases, the failure was recoverable, whereas in four cases there was a critical failure, requiring a conversion to standard laparoscopy in three of them, and the rescheduling of the surgery in one case. Conclusions: The da Vinci robotic system is extremely reliable for use in urology. Malfunction is rare and the risk of critical failure is very low. Managing mechanical failure before or during the surgery is the key to maintaining the safety of patients undergoing robotic surgical procedures.
引用
收藏
页码:736 / 740
页数:5
相关论文
共 50 条
  • [21] Camera-Robot Calibration for the Da Vinci Robotic Surgery System
    Ozguner, Orhan
    Shkurti, Thomas
    Huang, Siqi
    Hao, Ran
    Jackson, Russell C.
    Newman, Wyatt S.
    Cavusoglu, M. Cenk
    IEEE TRANSACTIONS ON AUTOMATION SCIENCE AND ENGINEERING, 2020, 17 (04) : 2154 - 2161
  • [22] Resection of ectopic mediastinal parathyroid glands with the da Vinci® robotic system
    Ismail, M.
    Maza, S.
    Swierzy, M.
    Tsilimparis, N.
    Rogalla, P.
    Sandrock, D.
    Rueckert, R. I.
    Mueller, J. M.
    Rueckert, J. C.
    BRITISH JOURNAL OF SURGERY, 2010, 97 (03) : 337 - 343
  • [23] Robotic mitral valve repair: Experience with the da Vinci system - Discussion
    Smith, CR
    Nifong
    ANNALS OF THORACIC SURGERY, 2003, 75 (02): : 443 - 443
  • [24] Feasibility study of intraocular robotic surgery with the da Vinci Surgical System
    Bourla, Dan H.
    Hubschman, Jean Pierre
    Culjat, Martin
    Tsirbas, Angelo
    Gupta, Anurag
    Schwartz, Steven D.
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2008, 28 (01): : 154 - 158
  • [25] Da Vinci Robotic Surgery in a Pediatric Hospital
    Mattioli, Girolamo
    Prato, Alessio Pini
    Razore, Barbara
    Leonelli, Lorenzo
    Pio, Luca
    Avanzini, Stefano
    Boscarelli, Alessandro
    Barabino, Paola
    Disma, Nicola Massimo
    Zanaboni, Clelia
    Garzi, Alfredo
    Martigli, Sofia Paola
    Buffi, Nicolo Maria
    Rosati, Ubaldo
    Petralia, Paolo
    JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2017, 27 (05): : 539 - 545
  • [26] Robotic da Vinci prostatectomy: The Greek experience
    Andripoulos, Nick
    Voulgarides, George
    Kosmaoglou, Elent
    Pardalidis, Nick
    JOURNAL OF ENDOUROLOGY, 2007, 21 : A103 - A103
  • [27] Robotic Partial Nephrectomy with the Da Vinci Xi
    Kallingal, George J. S.
    Swain, Sanjaya
    Darwiche, Fadi
    Punnen, Sanoj
    Manoharan, Murugesan
    Gonzalgo, Mark L.
    Parekh, Dipen J.
    ADVANCES IN UROLOGY, 2016, 2016
  • [28] Robotic assisted mitral valve repair: early experience with the da Vinci S robotic system
    Mega, S.
    Bono, M. C.
    Castiglione, I.
    Ranocchi, F.
    Casacalenda, A.
    Contento, C.
    Menichetti, A.
    Patti, G.
    Di Sciascio, G.
    Musumeci, F.
    EUROPEAN HEART JOURNAL, 2013, 34 : 979 - 979
  • [29] Is da Vinci Xi Better than da Vinci Si in Robotic Rectal Cancer Surgery? Comparison of the 2 Generations of da Vinci Systems
    Ozben, Volkan
    Cengiz, Turgut B.
    Atasoy, Deniz
    Bayraktar, Onur
    Aghayeva, Afag
    Erguner, Ilknur
    Baca, Bilgi
    Hamzaoglu, Ismail
    Karahasanoglu, Tayfun
    SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 2016, 26 (05): : 417 - 423
  • [30] COMPARATIVE STUDY OF THE DA VINCI XI VERSUS THE DA VINCI SI SURGICAL SYSTEM FOR BARIATRIC BYPASS SURGERY Robotic bariatric surgery
    Niclauss, N.
    Hagen, M.
    Jung, M.
    Morel, P.
    OBESITY SURGERY, 2017, 27 : 221 - 221