Malfunction of the da Vinci robotic system in urology

被引:15
|
作者
Chen, Cheng-Che [1 ]
Ou, Yen-Chuan [1 ,2 ,3 ]
Yang, Cheng-Kuang [1 ]
Chiu, Kun-Yuan [1 ]
Wang, Shian-Shiang [1 ]
Su, Chung-Kuang [1 ]
Ho, Hao-Chung [1 ]
Cheng, Chen-Li [1 ]
Chen, Chuan-Shu [1 ]
Lee, Jian-Ri [1 ]
Chen, Wen-Min [1 ]
机构
[1] Taichung Vet Gen Hosp, Div Urol, Dept Surg, Taichung 40705, Taiwan
[2] Chung Sang Med Univ, Inst Med, Taichung, Taiwan
[3] Natl Chi Nan Univ, Grad Inst Biomed & Biomed Technol, Dept Appl Chem, Nantou, Taiwan
关键词
da Vinci; malfunction; robot; urology; RADICAL PROSTATECTOMIES;
D O I
10.1111/j.1442-2042.2012.03010.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To analyze the incidence of malfunction of the da Vinci robotic system in a single center and to provide potential solutions. Methods: A total of 400 patients underwent da Vinci robotic urological surgery at Taichung Veterans General Hospital in Taichung, Taiwan, from December 2005 to April 2011. Episodes of malfunction of the robotic system were analyzed by period of operation, type of procedure, type of malfunction and management of the event. Results: Overall, 14 cases of malfunction occurred (3.5% of the entire series). Among them, five (1.25%) occurred before the surgery and nine (2.25%) intraoperatively. Operative procedures included radical prostatectomy, bilateral pelvic lymph node dissection, dismembered pyeloplasty, partial nephrectomy, nephroureterectomy, and radical and partial cystectomies. Areas of malfunctions included the robotic arm system and joint (11/14), optical system (1/14), power system and connector (1/14), endoscopic instrument (1/14), and software (1/14). In 10 cases, the failure was recoverable, whereas in four cases there was a critical failure, requiring a conversion to standard laparoscopy in three of them, and the rescheduling of the surgery in one case. Conclusions: The da Vinci robotic system is extremely reliable for use in urology. Malfunction is rare and the risk of critical failure is very low. Managing mechanical failure before or during the surgery is the key to maintaining the safety of patients undergoing robotic surgical procedures.
引用
收藏
页码:736 / 740
页数:5
相关论文
共 50 条
  • [1] MALFUNCTION OF DA VINCI ROBOTIC SURGERY IN UROLOGY
    Chen, Cheng-Che
    Ou, Yen-Chuan
    Yang, Cheng-Kuang
    Chiu, Kun-Yuan
    Wang, Shian-Shiang
    Su, Chung-Kuang
    Ho, Hao-Chung
    Cheng, Chen-Li
    Chen, Chuan-Shu
    Lee, Jian-Ri
    Chen, Wen-Min
    JOURNAL OF ENDOUROLOGY, 2011, 25 : A60 - A60
  • [2] Leonardo da Vinci, urology and the da Vinci robot
    Thomas, J.
    Rao, A. R.
    Motiwala, H.
    Karim, O.
    BJU INTERNATIONAL, 2009, 103 : 50 - 50
  • [3] Malfunction of the da Vinci Robotic System During Robot-Assisted Laparoscopic Prostatectomy: An International Survey
    Kaushik, Dharam
    High, Robin
    Clark, Curtis J.
    LaGrange, Chad A.
    JOURNAL OF ENDOUROLOGY, 2010, 24 (04) : 571 - 575
  • [4] The da Vinci robotic system in head and neck surgery
    Fakhoury, R.
    Prasad, V. M. N.
    Sayin, I.
    Van der Vorst, S.
    Remacle, M.
    Lawson, G.
    B-ENT, 2015, 11 : 1 - 6
  • [5] Application of the da Vinci robotic system in thoracic surgery
    Ismail, M.
    Swierzy, M.
    Ulrich, M.
    Rueckert, J. C.
    CHIRURG, 2013, 84 (08): : 643 - 650
  • [6] Effect of using the da Vinci Skills Simulator on robotic skill acquisition in urology residents
    Lu, Jirong
    Tiong, Ho Yee
    INTERNATIONAL JOURNAL OF UROLOGY, 2016, 23 : 19 - 20
  • [7] Malfunction of da Vinci Robotic System-Disassembled Surgeon's Console Hand Piece: Case Report and Review of the Literature
    Ham, Won Sik
    Park, Sung Yul
    Yu, Ho Song
    Choi, Young Deuk
    Hong, Sung Joon
    Rha, Koon Ho
    UROLOGY, 2009, 73 (01) : 209.e7 - 209.e8
  • [8] Incidence of device malfunction using the da Vinci robot during robotic assisted laparoscopic urologic surgery
    Lee, Chester
    Uberoi, Jayant
    Babayan, Richard K.
    Wang, David S.
    JOURNAL OF ENDOUROLOGY, 2007, 21 : A142 - A143
  • [9] Thoracoscopic thymectomy with the da Vinci robotic system for myasthenia gravis
    Rueckert, Jens C.
    Ismail, Mahmoud
    Swierzy, Marc
    Sobel, Holger
    Rogalla, Patrik
    Meisel, Andreas
    Wernecke, Klaus D.
    Rueckert, Ralph I.
    Mueller, Joachim M.
    MYASTHENIA GRAVIS AND RELATED DISORDERS: 11TH INTERNATIONAL CONFERENCE, 2008, 1132 : 329 - 335
  • [10] Impact of Refractive Errors on Da Vinci SI Robotic System
    Tuna, Mustafa Bilal
    Kilavuzoglu, Ayse Ebru
    Mourmouris, Panogiotis
    Argun, Omer Burak
    Doganca, Tunkut
    Obek, Can
    Ozisik, Ozan
    Kural, Ali Riza
    JSLS-JOURNAL OF THE SOCIETY OF LAPAROENDOSCOPIC SURGEONS, 2020, 24 (03) : 1 - 6