Background effects on Faraday collectors in gas-source mass spectrometry and implications for clumped isotope measurements

被引:117
|
作者
Bernasconi, Stefano M. [1 ]
Hu, Bin [1 ,2 ]
Wacker, Ulrike [3 ]
Fiebig, Jens [3 ,4 ]
Breitenbach, Sebastian F. M. [1 ]
Rutz, Tanja [3 ]
机构
[1] Swiss Fed Inst Technol, Inst Geol, CH-8092 Zurich, Switzerland
[2] Peking Univ, Dept Geog, Beijing 100871, Peoples R China
[3] Goethe Univ Frankfurt, Inst Geosci, Dept Paleontol, D-60438 Frankfurt, Germany
[4] Biodivers & Climate Res Ctr, D-60325 Frankfurt, Germany
关键词
CARBONATE MINERALS; C-13-O-18; BONDS; PALEOTHERMOMETER; THERMOMETRY; CO2;
D O I
10.1002/rcm.6490
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
RATIONALE The measurement of the abundances of minor isotopologues by mass spectrometry requires correction of subtle non-linearities in the mass spectrometer that cause deviations in the relationship between actual and measured isotope ratios. Here we show that negative backgrounds on the Faraday cups recording the minor ion beams are the cause of the observed non-linearities in the measurement of CO2 isotopologues, and propose a new correction procedure for clumped isotope measurements. METHODS We carefully investigated the cause of non-linearity effects in the measurement of the abundance of 13C18O16O, a minor isotopologue of CO2 with m/z 47, on two different mass spectrometers. By using gases of different composition with close to stochastic and with non-random distribution of isotopes we demonstrate that the apparent dependence of the excess abundance of the isotopologue of m/z 47 on the bulk isotopic composition of CO2 is due to a background interference that is linearly dependent on the partial pressure of the gas in the source of the mass spectrometer. CONCLUSIONS Background determination with gas flowing into the source of the mass spectrometer is necessary for accurate clumped isotope measurements of CO2. Background corrections can be performed accurately if the slit width of the m/z 44 Faraday cup significantly exceeds that of the one for m/z 47, using a correlation between m/z 44 signal intensity and the corresponding minimum in m/z 47 background. We propose two new correction schemes that reduce the time-consuming measurement of gases of different bulk isotopic compositions. These findings may also be relevant for the measurement of other rare isotopologues by mass spectrometry. Copyright (c) 2013 John Wiley & Sons, Ltd.
引用
收藏
页码:603 / 612
页数:10
相关论文
共 50 条
  • [31] NEW TECHNOLOGIES FOR SMALL SAMPLE STABLE ISOTOPE MEASUREMENT - STATIC VACUUM GAS SOURCE-MASS SPECTROMETRY, LASER PROBES, ION PROBES AND GAS-CHROMATOGRAPHY ISOTOPE RATIO MASS-SPECTROMETRY
    PILLINGER, CT
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1992, 118 : 477 - 501
  • [32] Nd isotope mixing during thermal ionisation mass spectrometry:: Implications for accurate 142Nd measurements
    Upadhyay, D.
    Scherer, E. E.
    Mezger, K.
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 2007, 71 (15) : A1047 - A1047
  • [33] ACCURACY AND PRECISION OF GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY FOR STABLE CARBON-ISOTOPE RATIO MEASUREMENTS
    WONG, WW
    HACHEY, DL
    ZHANG, S
    CLARKE, LL
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1995, 9 (11) : 1007 - 1011
  • [34] Low Temperature Catalytic Combustion Reactors for High Precision Carbon Isotope Measurements in Gas Chromatography Combustion Isotope Ratio Mass Spectrometry
    Tobias, Herbert J.
    Jones, Andrew
    Spanjers, Charlie
    Bowers, Larry
    Brenna, J. Thomas
    [J]. ANALYTICAL CHEMISTRY, 2019, 91 (04) : 2901 - 2907
  • [35] Simultaneous measurements of ethanol and ethyl-d(5) alcohol by stable isotope gas chromatography mass spectrometry
    Dean, R
    Thomasson, HR
    Dumaual, N
    Amann, D
    Li, TK
    [J]. CLINICAL CHEMISTRY, 1996, 42 (03) : 367 - 372
  • [36] Isotope ratio measurements using gas chromatography inductively coupled plasma mass spectrometry for the assessment of organolead sources
    Encinar, JR
    Granadillo, IL
    Alonso, JIG
    Sanz-Medel, A
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (05) : 475 - 480
  • [37] Improved analysis of micro- and nanomole-scale sulfur multi-isotope compositions by gas source isotope ratio mass spectrometry
    Au Yang, David
    Landais, Guillaume
    Assayag, Nelly
    Widory, David
    Cartigny, Pierre
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2016, 30 (07) : 897 - 907
  • [38] Determination of natural Cu-isotope variation by plasma-source mass spectrometry: implications for use as geochemical tracers
    Zhu, XK
    O'Nions, RK
    Guo, Y
    Belshaw, NS
    Rickard, D
    [J]. CHEMICAL GEOLOGY, 2000, 163 (1-4) : 139 - 149
  • [39] SOME INSTRUMENTAL EFFECTS IN THE DETERMINATION OF STABLE CARBON ISOTOPE RATIOS BY GAS-CHROMATOGRAPHY ISOTOPE RATIO MASS-SPECTROMETRY
    EAKIN, PA
    FALLICK, AE
    GERC, J
    [J]. CHEMICAL GEOLOGY, 1992, 101 (1-2) : 71 - 79
  • [40] Contribution of the Faraday cup coupled to 1012 Ω current amplifier to uranium 235/238 and 234/238 isotope ratio measurements by thermal ionization mass spectrometry
    Quemet, Alexandre
    Maloubier, Melody
    Ruas, Alexandre
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2016, 404 : 35 - 39