Determination of 238U/235U, 236U/238U and uranium concentration in urine using SF-ICP-MS and MC-ICP-MS:: An interlaboratory comparison

被引:28
|
作者
Parrish, RR [1 ]
Thirlwall, MF
Pickford, C
Horstwood, M
Gerdes, A
Anderson, J
Coggon, D
机构
[1] Univ Leicester, Dept Geol, Keyworth NG12 5GG, Notts, England
[2] British Geol Survey, NERC, Isotope Geosci Lab, Keyworth NG12 5GG, Notts, England
[3] Univ London Royal Holloway & Bedford New Coll, Dept Geol, Egham TW20 0EX, Surrey, England
[4] Sci Ltd, Didcot OX11 0TB, Oxon, England
[5] Univ Southampton, Southampton Gen Hosp, MRC, Epidemiol Resource Ctr, Southampton SO16 6YD, Hants, England
来源
HEALTH PHYSICS | 2006年 / 90卷 / 02期
基金
英国医学研究理事会;
关键词
accident analysis; uranium; weapons; radiation; ionizing;
D O I
10.1097/01.HP.0000174809.43871.54
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Accidental exposure to depleted or enriched uranium may occur in a variety of circumstances. There is a need to quantify such exposure, with the possibility that the testing may post-date exposure by months or years. Therefore, it is important to develop a very sensitive test to measure precisely the isotopic composition of uranium in urine at low levels of concentration. The results of an interlaboratory comparison using sector field (SIF)-inductively coupled plasma-mass spectrometry (ICP-MS) and multiple collector (MC-ICP-MS for the measurement of uranium concentration and U-235/U-238 and U-236/U-231 isotopic ratios of human urine samples are presented. Three urine samples were verified to contain uranium at 1-5 ng L-1 and shown to have natural uranium isotopic composition. Portions of these urine batches were doped with depleted uranium (DU) containing small quantities of U-236, and the solutions were split into 100 mL and 400 mL aliquots that were subsequently measured blind by three laboratories. All methods investigated were able to measure accura tely U-238/U-235 With precisions of similar to 0.5% to similar to 4%, but only selected MC-ICP-MS methods were capable of consistently analyzi ng U-236/U-238 to reasonable precision at the similar to 20 fg L-1 level of U-236 abundance. Isotope dilution using a U-233 tracer demonstrates the ability to measure concentrations to better than +/- 4% with the MC-ICP-MS method, though sample heterogeneity in urine samples was shown to be problematic in some cases. MC-ICP-MS outperformed SF-ICP-MS methods, as was expected. The MC-ICP-MS methodology described is capable of measuring to similar to 1 % precision the U-238/U-235 of any sample of human urine over the entire range of uranium abundance down to < 1 Ing L-1, and detecting very small amounts of DU contained therein.
引用
收藏
页码:127 / 138
页数:12
相关论文
共 50 条
  • [31] 238U/235U Systematics in Terrestrial Uranium-Bearing Minerals
    Hiess, Joe
    Condon, Daniel J.
    McLean, Noah
    Noble, Stephen R.
    SCIENCE, 2012, 335 (6076) : 1610 - 1614
  • [32] 236U inventories, 236U/238U, and 236U/239Pu:: The stratospheric fallout signature
    Ketterer, Michael E.
    Groves, Aran D.
    Strick, Brian J.
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2007, 71 (15) : A480 - A480
  • [33] Rapid uranium preconcentration and separation method from fresh water samples for total U and 235U/238U isotope ratio measurements by ICP-MS
    Tagami, K.
    Uchida, S.
    ANALYTICA CHIMICA ACTA, 2007, 592 (01) : 101 - 105
  • [34] Determination of the isotopic ratio 236U/238U in Austrian water samples
    Srncik, M.
    Steier, P.
    Wallner, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2010, 268 (7-8): : 1146 - 1149
  • [35] 铀颗粒物中235U/238U的LA-MC-ICP-MS分析技术研究
    汪伟
    李志明
    徐江
    周国庆
    王凡
    沈小攀
    翟利华
    柯昌凤
    质谱学报, 2016, 37 (03) : 213 - 221
  • [36] 235U、238U薄靶的制备
    魏永钦
    曹德林
    王秋玉
    核技术, 1983, (03) : 56 - 56
  • [37] Variations of the 238U/235U isotope composition in rivers
    Noordmann, Janine
    Weyer, Stefan
    Sharma, Mukul
    Georg, R. Bastian
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2009, 73 (13) : A951 - A951
  • [38] 238U/235U分馏及其地质应用
    徐林刚
    矿床地质, 2014, 33 (03) : 497 - 510
  • [39] HETEROGENEITY IN THE 238U/235U RATIOS OF ANGRITES.
    Tissot, F. L. H.
    Dauphas, N.
    Grove, T. L.
    METEORITICS & PLANETARY SCIENCE, 2016, 51 : A624 - A624
  • [40] 234U/238U and 235U/238U ratios in domestic water from the environs of Obuasi mine in Ghana
    A. R. Awudu
    E. O. Darko
    Journal of Radioanalytical and Nuclear Chemistry, 2011, 287 : 129 - 134