Frequency Stability Measurement of Cryogenic Sapphire Oscillators With a Multichannel Tracking DDS and the Two-Sample Covariance

被引:9
|
作者
Calosso, Claudio E. [1 ]
Vernotte, Francois [2 ,3 ]
Giordano, Vincent [2 ,3 ]
Fluhr, Christophe [4 ]
Dubois, Benoit [4 ]
Rubiola, Enrico [1 ,2 ,3 ]
机构
[1] Ist Nazl Ric Metrol, Phys Metrol Div, I-10135 Turin, Italy
[2] UBFC, FEMTO ST Inst, Dept Time & Frequency, F-25030 Besancon, France
[3] CNRS, ENSMM, F-25030 Besancon, France
[4] FEMTO Engn, F-25000 Besancon, France
关键词
Allan variance; digital instrumentation; frequency measurement; microwave oscillator; three-cornered hat (TCH); VARIANCE; DESIGN; TESTS;
D O I
10.1109/TUFFC.2018.2870593
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper shows the first measurement of three 100-MHz signals exhibiting fluctuations from 2 x10(-16) to parts in 10(-15) for an integration time tau between 1 s and 1 day. Such stable signals are provided by three cryogenic sapphire oscillators (CSOs) operating at about 10 GHz, also delivering the 100-MHz output via a dedicated synthesizer. The measurement is made possible by a six-channel tracking direct digital synthesizer (TDDS) and the two-sample covariance tool, used to estimate the Allan variance. The use of two TDDS channels per CSO enables high rejection of the instrument background noise. The covariance outperforms the three-cornered hat (TCH) method in that the background converges to zero "out of the box," with no need of the hypothesis that the instrument channels are equally noisy, nor of more sophisticated techniques to estimate the background noise of each channel. Thanks to correlation and averaging, the instrument background (AVAR) rolls off with a slope 1/root m, the number of measurements, down to 10(-18) at tau = 10(4) s. For consistency check, we compare the results to the traditional TCH method beating the 10-GHz outputs down to the megahertz region. Given the flexibility of the TDDS, our methods find an immediate application to the measurement of the 250-MHz output of the femtosecond combs.
引用
收藏
页码:616 / 623
页数:8
相关论文
共 10 条
  • [1] Direct Stability Measurement of Cryogenic Sapphire Oscillators with Tracking DDSs and Two-Sample Covariance
    Calosso, Claudio E.
    Fluhr, Christophe
    Dubois, Benoit
    Vernotte, F.
    Giordano, V.
    Rubiola, E.
    [J]. 2018 IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM (IFCS), 2018, : 53 - 54
  • [2] Direct Measurement of Cryogenic Sapphire Oscillators with Tracking DDSs and 2-Sample Covariance
    Calosso, Claudio E.
    Vernotte, Francois
    Giordano, Vincent
    Fluhr, Christophe
    Dubois, Benoit
    Rubiola, Enrico
    [J]. 2018 EUROPEAN FREQUENCY AND TIME FORUM (EFTF), 2018, : 41 - 44
  • [3] Short term frequency stability tests of two cryogenic sapphire oscillators
    Watabe, Ken-ichi
    Hartnett, John G.
    Locke, Clayton R.
    Santarelli, Giorgio
    Yanagimachi, Shinya
    Shimazaki, Takeshi
    Ikegami, Takeshi
    Ohshima, Shin-ichi
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2006, 45 (12): : 9234 - 9237
  • [4] Short term frequency stability tests of two cryogenic sapphire oscillators
    Watabe, Ken-Ichi
    Hartnett, John G.
    Locke, Clayton R.
    Santarelli, Giorgio
    Yanagimachi, Shinya
    Shimazaki, Takeshi
    Ikegami, Takeshi
    Ohshima, Shin-Ichi
    [J]. Japanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers, 2006, 45 (12): : 9235 - 9237
  • [5] Stability measurements of frequency synthesis with cryogenic sapphire oscillators
    Hartnett, John G.
    Creedon, Daniel
    Chambon, D.
    Santarelli, G.
    [J]. 2009 JOINT MEETING OF THE EUROPEAN FREQUENCY AND TIME FORUM AND THE IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM, VOLS 1 AND 2, 2009, : 372 - +
  • [6] Generation of 100 GHz with Parts in 1016 Frequency Stability Using Cryogenic Sapphire Oscillators
    Bara, Romain
    Le Floch, Jean-Michel
    Tobar, Michael E.
    Stanwix, Paul L.
    Parker, Stephen
    Hartnett, John G.
    Ivanov, Eugene N.
    [J]. 2011 JOINT CONFERENCE OF THE IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM/EUROPEAN FREQUENCY AND TIME FORUM PROCEEDINGS, 2011, : 193 - 194
  • [7] Frequency stability and phase noise of a pair of X-band cryogenic sapphire oscillators
    Marra, Giuseppe
    Henderson, Dale
    Oxborrow, Mark
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2007, 18 (05) : 1224 - 1228
  • [8] On measurement accuracy of the two-sample Allen variance in frequency instability estimation
    Borisov, B.D.
    Mishnev, A.S.
    [J]. Avtometriya, 2000, (03): : 122 - 125
  • [9] Characterization of the Individual Short-Term Frequency Stability of Cryogenic Sapphire Oscillators at the 10-16 Level
    Fluhr, Christophe
    Grop, Serge
    Dubois, Benoit
    Kersale, Yann
    Rubiola, Enrico
    Giordano, Vincent
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2016, 63 (06) : 915 - 921
  • [10] Rotating Dual Cryogenic Sapphire Oscillators with 10-16 Fractional Frequency Stability for Tests of Lorentz Invariance
    Parker, Stephen
    Stanwix, Paul
    Tobar, Michael
    Hartnett, John
    Ivanov, Eugene
    Nagel, Moritz
    Peters, Achim
    [J]. 2011 JOINT CONFERENCE OF THE IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM/EUROPEAN FREQUENCY AND TIME FORUM PROCEEDINGS, 2011, : 198 - +