Assessment of matrix effects associated with Fe isotope analysis using 266 nm femtosecond and 193 nm nanosecond laser ablation multi-collector inductively coupled plasma mass spectrometry

被引:48
|
作者
Zheng, Xin-Yuan [1 ,2 ]
Beard, Brian L. [1 ,2 ]
Johnson, Clark M. [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Geosci, 1215 W Dayton St, Madison, WI 53706 USA
[2] NASA, Astrobiol Inst, Washington, DC 20546 USA
关键词
MC-ICP-MS; LANGMUIR PROBE MEASUREMENTS; HIGH-PRECISION FE; IN-SITU; ELECTRON-DENSITY; SPACE-CHARGE; NONTRADITIONAL ISOTOPES; ELEMENTAL FRACTIONATION; ANALYTICAL-CHEMISTRY; IRON-METEORITES;
D O I
10.1039/c7ja00272f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study evaluated matrix effects during Fe isotope analysis using a 266 nm femtosecond (fs) laser and a 193 nm nanosecond (ns) laser coupled to a multi-collector ICP-MS. During 150 second spot analysis on pyrite, Fe isotope fractionation was not observed for fs-laser ablation (fs-LA), but was evident for ns-LA. The observed downhole Fe isotope fractionation during ns-LA is caused by multiple processes comprising ablation, transport, and ionization in the ICP. Contrary to the common perception of "matrix-free" analysis, matrix effects clearly exist during fs-LA analysis; small deviations of up to similar to 0.2 parts per thousand in the measured Fe-56/Fe-54 ratios from the true value of magnetite grains with >=similar to 8 wt% impurities were resolved using a nearly pure magnetite as the bracketing standard. Moreover, inaccurate and imprecise Fe-56/Fe-54 results were obtained when magnetite and pyrrhotite was measured against a non-matrix-matchedl standard (pyrite or Fe metal). The observed matrix effects during fs-LA cannot be explained by formation of a large heat-affected zone during ablation, but result from the influence of different chemical compositions of samples and standards on space-charge effects in the ICP-MS. Such matrix effects can be largely suppressed by water addition during analysis at a price of reduced sensitivity, so that precise and accurate Fe isotope analysis to a similar to 0.1 parts per thousand level can be routinely achieved under "wet" conditions without matrix-matching between sample and standard. These results may reconcile dramatically different precisions previously reported for Fe isotope analysis by fs-lasers, and also highlight fs-LA-MC-ICP-MS as an appealing option for in situ Fe isotope analysis on samples with complex matrices and high-symmetry minerals, both of which encounter significant analytical difficulties using secondary ion mass spectrometry (SIMS). For ns-LA, in addition to similar composition-related matrix effects experienced by fs-LA in the ICP, matrix effects also originate from ablation-related processes that produce sample particles with matrix-dependent size distributions and, often, larger aerodynamic sizes, resulting in highly inaccurate Fe-56/Fe-54 results during non-matrix-matched analysis under "dry" conditions. The collective matrix effects during Fe isotope analysis by ns-LA cannot be fully suppressed by water addition, therefore, matrix matching is required for accurate Fe isotope analysis by ns-LA.
引用
收藏
页码:68 / 83
页数:16
相关论文
共 50 条
  • [3] Comparison of 795 nm and 265 nm femtosecond and 193 nm nanosecond laser ablation inductively coupled plasma mass spectrometry for the quantitative multi-element analysis of glass materials
    Ohata, Masaki
    Tabersky, Daniel
    Glaus, Reto
    Koch, Joachim
    Hattendorf, Bodo
    Guenther, Detlef
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (08) : 1345 - 1353
  • [4] The effect of matrix interferences on in situ boron isotope analysis by laser ablation multi-collector inductively coupled plasma mass spectrometry
    Standish, Christopher D.
    Chalk, Thomas B.
    Babila, Tali L.
    Milton, J. Andy
    Palmer, Martin R.
    Foster, Gavin L.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2019, 33 (10) : 959 - 968
  • [5] Comparison of 265 nm Femtosecond and 213 nm Nanosecond Laser Ablation Inductively Coupled Plasma Mass Spectrometry for Pb Isotope Ratio Measurements
    Masaki Ohata
    Naoko Nonose
    Ladina Dorta
    Detlef Günther
    Analytical Sciences, 2015, 31 : 1309 - 1315
  • [6] Comparison of 265 nm Femtosecond and 213 nm Nanosecond Laser Ablation Inductively Coupled Plasma Mass Spectrometry for Pb Isotope Ratio Measurements
    Ohata, Masaki
    Nonose, Naoko
    Dorta, Ladina
    Guenther, Detlef
    ANALYTICAL SCIENCES, 2015, 31 (12) : 1309 - 1315
  • [7] Homogenised 266 nm femtosecond laser ablation for isotopic and elemental microanalyses using inductively coupled plasma mass spectrometry
    Kimura, Jun-Ichi
    Ohki, Koichi
    Chang, Qing
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (06) : 1203 - 1210
  • [8] Strontium isotope analysis in prehistoric cod otoliths by laser ablation multi-collector inductively coupled plasma mass spectrometry
    Glykou, Aikaterini
    Ritchie, Kenneth
    Hargrave, Matthew S.
    Visch, Wouter
    Liden, Kerstin
    JOURNAL OF ARCHAEOLOGICAL SCIENCE-REPORTS, 2021, 37
  • [9] Enhanced sensitivity in laser ablation multi-collector inductively coupled plasma mass spectrometry
    Cottle, John M.
    Burrows, Andrew J.
    Kylander-Clark, Andrew
    Freedman, Philip A.
    Cohen, Roy S.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (11) : 1700 - 1706
  • [10] In situ calcium isotope analysis of Sr-rich carbonates using laser ablation multi-collector inductively coupled plasma mass spectrometry
    Jiang, Jiao
    Chen, Wei
    Lu, Jue
    Liu, Yonghong
    Li, Ming
    Sun, Jian
    Zhao, Kuidong
    Jiang, Shaoyong
    Liu, Yongsheng
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2025, 40 (04)