For any radiotherapy verification technique, it is desirable that issues with the accelerator, multileaf collimator and patient position be detected. In previous works, an effective method for this level of verification was presented. This paper identifies second-order issues affecting the part of the process in which the incident energy fluence is verified. These problems will affect any rotational intensity-modulated radiotherapy delivery that divides each rotation or arc into projections; however the solutions offered in this paper are specific to the method previously developed. The issues affecting the energy fluence verification method include leaf bouncing, delivery implementation and leaf latency. All three matters were found to introduce small errors in the verified energy fluence values for a small fraction of leaf states. The overall effect on the deposited dose over the course of a rotational delivery involving thousands of beam pulses per rotation is negligible. Regardless, effective correction strategies are presented; these are utilized in order to characterize both the delivered energy fluence and deposited dose as accurately as possible.
引用
收藏
页码:2953 / 2965
页数:13
相关论文
共 50 条
  • [41] Modeling the Fluence Output Factor for TomoTherapy Dynamic Jaw Delivery
    Chen, Y.
    Lu, W.
    Sterpin, E.
    Chen, M.
    Ruchala, K.
    Olivera, G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S830 - S831
  • [42] TomoTherapy Based IMRT for Breast Cancer with Skull Metastases
    Cyriac, A.
    Lee, D.
    Mott, J.
    CLINICAL ONCOLOGY, 2011, 23 (03) : S34 - S34
  • [43] A Clinical Workflow for TomoTherapy Delivery Verification
    Lu, W.
    Parnell, D.
    Olivera, G.
    Galmarini, D.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [44] Delivery verification in sequential and helical tomotherapy
    Kapatoes, JM
    Olivera, GH
    Reckwerdt, PJ
    Fitchard, EE
    Schloesser, EA
    Mackie, TR
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07): : 1815 - 1841
  • [45] Evaluation of BANG® gel for the verification of tomotherapy
    Todorovic, M.
    Bartels, A.
    Fischer, M.
    Cremers, F.
    Maryanskiz, M.
    Schmidt, R.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S187 - S187
  • [46] Delivery Verification for Tomotherapy Treatment System
    Chen, Q.
    Chapman, T.
    Sobering, G.
    Lu, W.
    Ruchala, K.
    Olivera, G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S669 - S669
  • [47] Evaluation of EPID ghosting effects in IMRT fluence maps
    Figueira, A. C.
    Monteiro, A.
    Trindade, M.
    Pinto, G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S719 - S719
  • [48] Optimization of IMRT fluence-map smoothing can result in improved IMRT plans
    Salter, B. J.
    Anker, C. J.
    Tobler, M.
    Hitchcock, Y. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S681 - S681
  • [49] Fast approximate delivery of fluence maps for IMRT and VMAT
    Balvert, Marleen
    Craft, David
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (04): : 1225 - 1247
  • [50] IMRT VERIFICATION IN BALTIC STATES
    Gershkevitsh, E.
    Popovs, S.
    Meijers, A.
    Miller, A.
    Griskevicius, R.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S459 - S459