Comparison of Less Invasive Stabilization System Plate and Retrograde Intramedullary Nail in the Fixation of Femoral Supracondylar Fractures in the Elderly: A Biomechanical Study

被引:17
|
作者
Du, Yu-ren [1 ]
Ma, Jian-xiong [2 ]
Wang, Shuo [3 ,4 ]
Sun, Lei [2 ]
Wang, Ying [2 ]
Lu, Bin [2 ]
Bai, Hao-hao [2 ]
Hu, Yong-cheng [3 ]
Ma, Xin-long [1 ,2 ]
机构
[1] Tianjin Med Univ, Gen Hosp, Dept Orthopaed, Tianjin, Peoples R China
[2] Tianjin Hosp, Tianjin Inst Orthopaed, Tianjin 300050, Peoples R China
[3] Tianjin Hosp, Dept Orthopaed, Tianjin 300211, Peoples R China
[4] 464 Hosp Peoples Liberat Army, Dept Orthopaed, Tianjin, Peoples R China
关键词
Biomechanical testing; Comminuted femoral supracondylar fracture; Less invasive stabilization system plate; Retrograde intramedullary nail; DISTAL FEMUR FRACTURES; INTERNAL-FIXATION; LCP; EPIDEMIOLOGY; LISS;
D O I
10.1111/os.12449
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective To compare the biomechanical stabilities of less invasive stabilization system (LISS) plate and retrograde intramedullary nail (IMN) for the comminuted femoral supracondylar fracture fractures in the elderly. Methods Sixteen pairs of embalmed cadaver femurs were obtained to simulate a comminuted supracondylar femur fracture (AO/OTA33-A3) gap model. All left-side specimens were fixed with LISS plate, and retrograde IMN were applied to the right-side specimens. All specimens were tested in torsional, axial and cyclic load mode on an Instron testing machine. Results The mean torsional stiffness for LISS plate group was 34.1% greater than retrograde IMN group (2.90 vs. 1.91 Nm/degree, P = 0.002), but the mean axial stiffness was greater for the retrograde IMN (199.16 vs. 303.93 N/mm, P < 0.001). The total deformation of LISS plate caused by cyclic axial loading was greater than retrograde IMN (4.17 vs. 3.57 mm, P = 0.014). Significantly less mean irreversible deformation was detected in LISS plate than in retrograde IMN (1.64 vs. 1.69 mm, P = 0.699). Failure loads of the constructs were significantly different between the two groups (LISS plate: 2941 +/- 128 N; retrograde IMN: 4022 +/- 176 N, P < 0.001). Conclusion For comminuted femoral supracondylar fractures in the elderly, the tested instruments can both maintain sufficient biomechanical stabilities, but retrograde IMN is superior to LISS plate in deformation of fracture site.
引用
收藏
页码:311 / 317
页数:7
相关论文
共 50 条
  • [1] Biomechanical evaluation of the Less Invasive Stabilization System, angled blade plate, and retrograde intramedullary nail for the internal fixation of distal femur fractures
    Zlowodzki, M
    Williamson, S
    Cole, PA
    Zardiackas, LD
    Kregor, PJ
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) : 494 - 502
  • [2] Biomechanical analysis of retrograde intramedullary nail fixation in distal femoral fractures
    Chen, Shih-Hao
    Yu, Tzai-Chiu
    Chang, Chih-Han
    Lu, Yuan-Chiao
    KNEE, 2008, 15 (05): : 384 - 389
  • [3] Treatment of supracondylar femoral fractures with a retrograde intramedullary nail
    Gellman, RE
    Paiement, GD
    Green, HD
    Coughlin, RR
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (332) : 90 - 97
  • [4] The retrograde intramedullary supracondylar nail: an alternative in the treatment of distal femoral fractures in the elderly?
    Janzing, HMJ
    Stockman, B
    Van Damme, G
    Rommens, P
    Broos, PLO
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1998, 118 (1-2) : 92 - 95
  • [5] The retrograde intramedullary supracondylar nail: an alternative in the treatment of distal femoral fractures in the elderly?
    H. M. J. Janzing
    B. Stockman
    G. Van Damme
    P. Rommens
    P. L. O. Broos
    Archives of Orthopaedic and Trauma Surgery, 1998, 118 : 92 - 95
  • [6] Biomechanical performance of retrograde nail for supracondylar fractures stabilization
    Nattapon Chantarapanich
    Kriskrai Sitthiseripratip
    Banchong Mahaisavariya
    Pongwit Siribodhi
    Medical & Biological Engineering & Computing, 2016, 54 : 939 - 952
  • [7] Distal femoral fixation: A biomechanical comparison of retrograde nail, retrograde intramedullary nail, and prototype locking retrograde nail
    Heiney, Jake P.
    Battula, Suneel
    O'Connor, Jill A.
    Ebraheim, Nabil
    Schoenfeld, Andrew J.
    Vrabec, Gregory
    CLINICAL BIOMECHANICS, 2012, 27 (07) : 692 - 696
  • [8] Biomechanical performance of retrograde nail for supracondylar fractures stabilization
    Chantarapanich, Nattapon
    Sitthiseripratip, Kriskrai
    Mahaisavariya, Banchong
    Siribodhi, Pongwit
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2016, 54 (06) : 939 - 952
  • [9] Treatment of distal femoral fractures in the elderly. Results with the retrograde intramedullary supracondylar nail
    Janzing H.M.J.
    Vaes F.
    Van Damme G.
    Stockman B.
    Broos P.L.O.
    Unfallchirurgie, 1998, 24 (2): : 55 - 59
  • [10] The supracondylar intramedullary nail in elderly patients with distal femoral fractures
    Dunlop, DG
    Brenkel, IJ
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1999, 30 (07): : 475 - 484