摘要
Objectives: To compare the practical use, safety, and clinical outcomes associated with the TandemHeart (TH) versus Impella Recover 2.5 (IR2.5) devices when used for circulatory support during high-risk percutaneous coronary intervention (PCI). Background: Small studies and registries suggest safety and efficacy for the TH and IR2.5 percutaneous-left ventricular assist devices (P-LVADs). However, these P-LVADs differ markedly in their insertion, operation, and manner of circulatory augmentation. To date, no study has compared these devices. Methods: We identified 68 patients (49 males, 19 females; age 71.1 +/- 12.1 years) from our single-center database that underwent high-risk PCI with P-LVAD support from April 2005 to June 2010 (32 with TH, 36 with IR2.5). Relevant data were extracted for analysis. Results: Baseline demographics were similar, including low LVEF (overall mean 31.0 +/- 13.7%) and elevated STS mortality risk score (4.2 +/- 3.7%). Angiographic characteristics were also similar, with a mean of 2.4 +/- 1.0 lesions treated per patient, and 29% undergoing left main PCI. PCI success rates were 99% in both groups, with similar in-hospital outcomes and a combined 7% major vascular access site complication rate. A single episode of left atrial perforation occurred during TH use. No patient required emergent CABG and no in-hospital deaths occurred. The 30-day MACE rate (death, myocardial infarction, target lesion revascularization) was 5.8%. There were no differences between the IR2.5 and TH groups with respect to short- or long-term clinical outcomes. Conclusions: The IR2.5 and TH assist devices are safe, equally effective, and associated with acceptable short- and long-term clinical outcomes in patients undergoing high-risk PCI. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:E28 / E37
页数:10
相关论文
共 50 条
  • [31] The Duration of Impella 2.5 Circulatory Support and Length of Hospital Stay of Patients Undergoing High-risk Percutaneous Coronary Interventions
    Anusionwu, Obiora
    Fischman, Daniel
    Cheriyath, Pramil
    CARDIOLOGY RESEARCH, 2012, 3 (04) : 154 - 157
  • [32] The effectiveness and safety of the Impella ventricular assist device for high-risk percutaneous coronary interventions: A systematic review
    Ichou, Jamal Ait
    Larivee, Natasha
    Eisenberg, Mark J.
    Suissa, Karine
    Filion, Kristian B.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2018, 91 (07) : 1250 - 1260
  • [33] Left Ventricular Rupture After Impella® Placement During High-Risk Percutaneous Coronary Intervention
    de Armas, Ismael A. Salas
    Kumar, Sachin
    Almustafa, Ahmed
    Akkanti, Bindu
    Akay, M. Hakan
    Patel, Manish K.
    Patel, Jayeshkumar
    Rajagopal, Keshava
    Marcano, Juan
    Goswami, Rajiv
    Gregoric, Igor D.
    Kar, Biswajit
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2020, 21 (11) : S100 - S102
  • [34] Trans-Aortic Percutaneous Heart Valve Deployment of Impella CP™ to Assist in High-Risk Percutaneous Coronary Intervention
    Chan, William
    Seidelin, Peter H.
    HEART LUNG AND CIRCULATION, 2014, 23 (05): : E142 - E144
  • [35] Use of the Impella 2.5 for prophylactic circulatory support during elective high-risk percutaneous coronary intervention
    Alasnag, Mirvat A.
    Gardi, Delair O.
    Elder, Mahir
    Kannam, Hari
    Ali, Farhan
    Petrina, Mircea
    Kheterpal, Vipin
    Hout, Mariah S.
    Schreiber, Theodore L.
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2011, 12 (05) : 299 - 303
  • [36] Effects of Mechanical Left Ventricular Unloading by Impella on Left Ventricular Dynamics in High-Risk and Primary Percutaneous Coronary Intervention Patients
    Remmelink, Maurice
    Sjauw, Krischan D.
    Henriques, Jose P. S.
    de Winter, Robbert J.
    Vis, Marije M.
    Koch, Karel T.
    Paulus, Walter J.
    de Mol, Bas A. J. M.
    Tijssen, Jan G. P.
    Piek, Jan J.
    Baan, Jan, Jr.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 75 (02) : 187 - 194
  • [37] Percutaneous Left Axillary Artery Approach for Impella 2.5 Liter Circulatory Support for Patients with Severe Aortoiliac Arterial Disease Undergoing High-Risk Percutaneous Coronary Intervention
    Lotun, Kapildeo
    Shetty, Ranjith
    Patel, Manishkumar
    Arain, Salman A.
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2012, 25 (02) : 210 - 213
  • [38] Impella ventricular assist device: A "valvular bypass" to support high-risk percutaneous coronary intervention or complicated transcatheter aortic valve replacement
    Tarantini, Giuseppe
    Nai Fovino, Luca
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 95 (03) : 363 - 364
  • [39] Left ventricular assistant device impella 2.5 usage for patients undergoing high-risk PCI
    Palacios, Igor F.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 79 (07) : 1135 - 1137
  • [40] The evolving role of percutaneous ventricular assist devices in high-risk cardiac patients
    Marcuschamer, Ilan A.
    Abelow, Aryeh
    Kornowski, Ran
    CORONARY ARTERY DISEASE, 2018, 29 (04) : 344 - 353