Estimation of absorbed organ doses and effective dose based on body mass index in digital radiography

被引:4
|
作者
Kim, HyunJi [1 ]
Park, MinSeok [1 ,2 ]
Park, SongYi [1 ,3 ]
Jeong, HoiWoun [4 ,5 ]
Kim, Jungmin [1 ]
Kim, Youhyun [1 ]
机构
[1] Korea Univ, Dept Radiol Sci, Coll Hlth Sci, Seoul 136703, South Korea
[2] Korea Inst Radiol & Med Sci, Seoul 139706, South Korea
[3] Asan Med Ctr, Dept Nucl Med, Seoul 138736, South Korea
[4] Baekseok Culture Univ, Dept Radiol Sci, Chungcheongnam Do 303705, South Korea
[5] Inje Univ, Seoul Paik Hosp, Seoul 100766, South Korea
基金
新加坡国家研究基金会;
关键词
RADIATION; OVERWEIGHT; OBESITY;
D O I
10.1093/rpd/ncs089
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
With the introduction of digital radiography, patients undergoing radiographic procedures are subject to being overexposed to radiation. Therefore, it is necessary to estimate the absorbed organ dose and the effective dose, which are significant for patient health, along with body type. During chest radiographic examinations conducted in 899 patients for screening, the absorbed dose of the 13 major organs, the average whole-body dose, and two effective doses weighted by factors published in ICRP 60 and ICRP 103 were calculated on the basis of patient information such as height, weight and examination condition, including kilovolt potential, focus-skin distance and entrance surface dose (ESD), using a PC-based Monte Carlo program simulation. It was found that dose per unit ESD had a tendency to decrease with body mass index (BMI). In particular, the absorbed dose for most organs was larger at high voltages (140 kVp) than at low voltages (120 kVp, 100 kVp). In addition, the effective dose which was based on ICRP 60 and ICRP 103 also represented the same tendency in respect of BMI and tube voltage.
引用
收藏
页码:92 / 99
页数:8
相关论文
共 50 条
  • [31] ESTIMATION OF ORGAN DOSES AND EFFECTIVE DOSES BASED ON IN-PHANTOM DOSIMETRY FOR PAEDIATRIC DIAGNOSTIC CARDIAC CATHETERISATION
    Kawasaki, Toshio
    Sakakubo, Masami
    Ito, Kanako
    Kitagawa, Ai
    [J]. RADIATION PROTECTION DOSIMETRY, 2019, 185 (02) : 215 - 221
  • [32] CALDose_X-a software tool for the assessment of organ and tissue absorbed doses, effective dose and cancer risks in diagnostic radiology
    Kramer, R.
    Khoury, H. J.
    Vieira, J. W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (22): : 6437 - 6459
  • [33] Effective dose and organ doses estimation taking tube current modulation into account with a commercial software package
    X. Lopez-Rendon
    H. Bosmans
    R. Oyen
    F. Zanca
    [J]. European Radiology, 2015, 25 : 1919 - 1925
  • [34] Effective dose and organ doses estimation taking tube current modulation into account with a commercial software package
    Lopez-Rendon, X.
    Bosmans, H.
    Oyen, R.
    Zanca, F.
    [J]. EUROPEAN RADIOLOGY, 2015, 25 (07) : 1919 - 1925
  • [35] Organ Absorbed Doses and Effective Doses to the Patient and the Medical Staff in Interventional Radiology Calculated from Voxel Phantom
    Park, Sang-hyun
    Lee, Choonsik
    Lee, Jai-ki
    [J]. JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2008, : 309 - 312
  • [36] ACR Dose Index Registry-Digital Radiography Pilot
    Don, Steven
    Sensakovic, William F.
    Gress, Dustin
    Zacharias-Andrews, Kay
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2021, 18 (08) : 1213 - 1215
  • [37] Digital skeletal radiography - Reduction of absorbed dose by adaptation of exposure factors and image processing
    Axelsson, B
    Petersen, U
    Wiltz, HJ
    [J]. ACTA RADIOLOGICA, 2001, 42 (06) : 592 - 598
  • [38] Body Mass and Body Mass Index estimation in medieval Switzerland
    Siegmund, Frank
    Papageorgopoulou, Christina
    [J]. ANTHROPOLOGISCHER ANZEIGER, 2011, 68 (04) : 524 - 524
  • [39] A Landmark-Based Approach for Robust Estimation of Exposure Index Values in Digital Radiography
    Irrera, Paolo
    Bloch, Isabelle
    Delplanque, Maurice
    [J]. MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2015, PT II, 2015, 9350 : 627 - 634
  • [40] Digital radiography: are the manufacturers' settings too high? Optimisation of the Kodak digital radiography system with aid of the computed radiography dose index
    Sinead E. Peters
    Patrick C. Brennan
    [J]. European Radiology, 2002, 12 : 2381 - 2387