Comparison of two dose calculation methods applied to extracranial stereotactic radiotherapy treatment planning

被引:15
|
作者
Hansen, AT [1 ]
Petersen, JB
Hoyer, M
Christensen, JJ
机构
[1] Aarhus Univ Hosp, Dept Med Phys, DK-8000 Aarhus, Denmark
[2] Aarhus Univ Hosp, Dept Oncol, DK-8000 Aarhus, Denmark
关键词
extracraniel stereotactic radiotherapy; treatment planning; dose calculation algorithms; collapsed cone;
D O I
10.1016/j.radonc.2005.04.018
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Lung tissue is a special challenge for a dose calculation algorithm, especially in the case of extracranial stereotactic radiotherapy (ESRT) due to small field sizes in combination with large variations in tissue density. The present study investigates the choice of dose calculation algorithm for 18 patients with a single lung tumor and 8 patients with a single liver tumor. The dose calculation is performed with both the pencil beam convolution algorithm and the collapsed cone convolution algorithm with the same number of monitor units in both cases. In addition, the dose calculation with the collapsed cone convolution algorithm is also performed with modified field sizes in order to match the Planning Target Volume (PTV) peripheral dose of the pencil beam based treatment. Results: For liver tumors, the mean Clinical Target Volume (CTV) dose calculated by the collapsed cone convolution algorithm and the pencil beam convolution algorithm is almost identical. For lung tumors, the mean CTV dose determined by the collapsed cone convolution algorithm differs up to 20%. Plans obtained by the two algorithms have field sizes which differ up to 8 mm for the same number of monitor units and minimum dose to the lung PTV. Conclusions: The choice of dose calculation algorithm can have a large influence on a treatment plan for ESRT of the lungs. (C) 2005 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:96 / 98
页数:3
相关论文
共 50 条
  • [1] Accuracy of treatment planning system and delivery of dose in extracranial stereotactic radiotherapy
    Dawod, T.
    Werner, M.
    Bremer, M.
    Karstens, J. H.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 : 105 - 106
  • [2] Extracranial stereotactic radiotherapy of lung and liver cancer - a clinical comparison of two treatment planning systems
    Bogner, J
    Dvorak, P
    Kroupa, B
    Dieckmann, K
    Georg, D
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S45 - S45
  • [3] Treatment planning system and dose delivery accuracy in extracranial stereotactic radiotherapy using Elekta body frame
    Dawod, Tamer
    Bremer, Michael
    Karstens, Johann H.
    Werner, Martin
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2010, 16 (01): : 23 - 34
  • [4] Stereotactic conformation radiotherapy with three dimensional planning for the treatment of extracranial lesions
    Tanaka, Y
    Kawamori, J
    Hirayama, M
    Saito, T
    Itoh, H
    Endoh, Y
    CAR '98 - COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1998, 1165 : 507 - 512
  • [5] Comparison of selected dose calculation algorithms in radiotherapy treatment planning for tissues with inhomogeneities
    Woon, Y. L.
    Heng, S. P.
    Wong, J. H. D.
    Ung, N. M.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [6] Stereotactic body radiotherapy treatment of extracranial metastases
    Salama, Joseph K.
    Kirkpatrick, John P.
    Yin, Fang-Fang
    NATURE REVIEWS CLINICAL ONCOLOGY, 2012, 9 (11) : 654 - 665
  • [7] Stereotactic body radiotherapy treatment of extracranial metastases
    Joseph K. Salama
    John P. Kirkpatrick
    Fang-Fang Yin
    Nature Reviews Clinical Oncology, 2012, 9 : 654 - 665
  • [8] Optimal treatment planning of extracranial stereotactic conformal radiotherapy for medically inoperable lung cancer
    Wang, L
    Feigenberg, S
    Paskalev, K
    Chen, L
    Qin, L
    Ma, CC
    Movsas, B
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S630 - S631
  • [9] CALCULATION OF ELECTRON DOSE DISTRIBUTION FOR RADIOTHERAPY TREATMENT PLANNING
    KAWACHI, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (04): : 571 - 577
  • [10] Comparison of various extracranial stereotactic Radiotherapy (ESRT) - Modalities
    Roidl, O.
    Moustakis, C.
    Eich, H. T.
    Haverkamp, U.
    Frick, S.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2014, 190 : 71 - 71