Quality assurance evaluation of spot scanning beam proton therapy with an anthropomorphic prostate phantom

被引:12
|
作者
Iqbal, K. [1 ,2 ,3 ]
Gillin, M. [1 ]
Summers, P. A. [1 ,4 ]
Dhanesar, S. [1 ]
Gifford, K. A. [1 ]
Buzdar, S. A. [2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Islamia Univ Bahawalpur, Dept Phys, Bahawalpur, Pakistan
[3] Shaukat Khanum Canc Hosp & Res Ctr, Dept Radiat Oncol, Lahore, Pakistan
[4] Univ Texas MD Anderson Canc Ctr, Radiol Phys Ctr, Houston, TX 77030 USA
来源
BRITISH JOURNAL OF RADIOLOGY | 2013年 / 86卷 / 1031期
关键词
RADIOTHERAPY; SYSTEM;
D O I
10.1259/bjr.20130390
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this study was to evaluate spot scanning proton therapy with an anthropomorphic prostate phantom at the Proton Therapy Center of The University of Texas MD Anderson Cancer Center at Houston, TX (PTCH). Methods: An anthropomorphic prostate phantom from the Radiological Physics Center (RPC), The University of Texas MD Anderson Cancer Center, Houston, TX, was used, which contained thermoluminescent dosemeters and GAFCHROMIC (R) EBT2 film (ISP Technologies, Wayne, NJ). The phantom was irradiated by the Hitachi synchrotron (Hitachi America, Ltd, Tarrytown, NY), and the results were compared between the treatment planning system (TPS) and RPC measurements. Results: RPC results show that the right/left, inferior/superior and posterior/anterior aspects of the coronal/sagittal and EBT2 film measurements were within 67%/64mm of the TPS. The RPC thermoluminescent dosemeter measurements of the prostate and femoral heads were within 3% of the TPS. Conclusion: The RPC prostate phantom is a useful mechanism to evaluate spot scanning beam proton therapy within certain confidence levels. Advances in knowledge: The RPC anthropomorphic prostate phantom could be used to establish quality assurance of spot scanning proton beam for patients with prostate cancer.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Development of a Quality Assurance Phantom for Proton Radiographic Images Using the MEVION Spot Scanning Proton Therapy System
    Pelas, C.
    Richards, Z.
    Alsbou, N.
    Ahmad, S.
    Ali, I.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E527 - E527
  • [2] A fast and reliable method for daily quality assurance in spot scanning proton therapy with a compact and inexpensive phantom
    Bizzocchi, Nicola
    Fracchiolla, Francesco
    Schwarz, Marco
    Algranati, Carlo
    [J]. MEDICAL DOSIMETRY, 2017, 42 (03) : 238 - 246
  • [3] Real-Time Visualization of Discrete Spot Scanning Proton Therapy Beam for Quality Assurance
    Matsuzaki, Y.
    Jenkins, C.
    Yang, Y.
    Yoshimura, T.
    Fujii, Y.
    Umegaki, K.
    Xing, L.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3758 - 3759
  • [4] Evaluation of An Anthropomorphic Pelvis Phantom for Proton Therapy
    Grant, R.
    Ibbott, G.
    Sahoo, N.
    Tucker, S.
    Zhu, X.
    Followill, D.
    [J]. MEDICAL PHYSICS, 2009, 36 (06) : 2607 - +
  • [5] Clinical Validation of Proton Pencil Beam Scanning on an Anthropomorphic Lung Phantom
    Wang, P.
    Tang, S.
    Cummings, D.
    Lee, A. K.
    Chang, C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E736 - E736
  • [6] Detectors for quality assurance of pencil beam scanning gantries for proton therapy
    Gagnon-Moisan, F.
    van der Meer, R.
    Chowdhuri, Z.
    Eichin, M.
    Koenig, S.
    Schmid, F. -J.
    Weber, D. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S44 - S45
  • [7] Applications of a novel detector for pencil beam scanning proton therapy beam quality assurance
    Yang, Pei-Ying
    Hsieh, Yang-Wei
    Kang, Chen-Lin
    Tseng, Chin-Dar
    Lin, Chih-Hsueh
    Shieh, Chin-Shiuh
    Chao, Pei-Ju
    Lee, Tsair-Fwu
    [J]. INTERNATIONAL JOURNAL OF MODERN PHYSICS B, 2021, 35 (14N16):
  • [8] A Comparison between Pencil Beam and Monte Carlo Algorithms Against Film Measurements in an Anthropomorphic Phantom for Proton Spot Scanning
    Saini, J.
    Maes, D.
    Bowen, S. R.
    Egan, A. J.
    Bloch, C.
    Wong, T. P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E717 - E718
  • [9] Design of an anthropomorphic intensity modulated radiation therapy quality assurance phantom
    Radford, DA
    Followill, DS
    Hanson, WF
    [J]. PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 197 - 200
  • [10] Clinical examination of proton pencil beam scanning on a moving anthropomorphic lung phantom
    Wang, Peng
    Tang, Shikui
    Taylor, Paige A.
    Cummings, David E.
    Janson, Martin
    Traneus, Erik
    Sturgeon, Jared D.
    Lee, Andrew K.
    Chang, Chang
    [J]. MEDICAL DOSIMETRY, 2019, 44 (02) : 122 - 129