Achieving high spatial and temporal resolution with perfusion MRI in the head and neck region using golden-angle radial sampling

被引:9
|
作者
Tomppert, Andrea [1 ]
Wuest, Wolfgang [1 ,2 ]
Wiesmueller, Marco [1 ]
Heiss, Rafael [1 ]
Kopp, Markus [1 ]
Nagel, Armin M. [1 ,3 ]
Tomita, Hayato [4 ]
Meixner, Christian [1 ,3 ]
Uder, Michael [1 ,2 ]
May, Matthias Stefan [1 ,2 ]
机构
[1] Univ Hosp Erlangen, Dept Radiol, Maximilianspl 3, D-91054 Erlangen, Germany
[2] Univ Hosp Erlangen, Imaging Sci Inst, Erlangen, Germany
[3] Friedrich Alexander Univ Erlangen Nurnberg FAU, Inst Med Phys, Erlangen, Germany
[4] St Marianna Univ, Dept Radiol, Sch Med, Miyamae Ku, Kawasaki, Kanagawa, Japan
关键词
Adult; Magnetic resonance imaging; Neck; Perfusion; SQUAMOUS-CELL CARCINOMA; CONTRAST-ENHANCED MRI; PAROTID TUMORS; DCE-MRI; RECONSTRUCTION; BENIGN;
D O I
10.1007/s00330-020-07263-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives Conventional perfusion-weighted MRI sequences often provide poor spatial or temporal resolution. We aimed to overcome this problem in head and neck protocols using a golden-angle radial sparse parallel (GRASP) sequence. Methods We prospectively included 58 patients for examination on a 3.0-T MRI using a study protocol. GRASP (A) was applied to a volumetric interpolated breath-hold examination (VIBE) with 135 reconstructed pictures and high temporal (2.5 s) and spatial resolution (0.94 x 0.94 x 3.00 mm). Additional sequences of matching temporal resolution (B: 2.5 s, 1.88 x 1.88 x 3.00 mm), with a compromise between temporal and spatial resolution (C: 7.0 s, 1.30 x 1.30 x 3.00 mm) and with matching spatial resolution (D: 145 s, 0.94 x 0.94 x 3.00 mm), were subsequently without GRASP. Instant inline-image reconstructions (E) provided one additional series of averaged contrast information throughout the entire acquisition duration of A. Overall diagnostic image quality, edge sharpness and contrast of soft tissues, vessels and lesions were subjectively rated using 5-point Likert scales. Objective image quality was measured as contrast-to-noise ratio in D and E. Results Overall, the anatomic and pathologic image quality was substantially better with the GRASP sequence for the temporally (A/B/C, allp < 0.001) and spatially resolved comparisons (D/E, allp < 0.002 except lesion edge sharpness withp = 0.291). Image artefacts were also less likely to occur with GRASP. Differences in motion, aliasing and truncation were mainly significant, but pulsation and fat suppression were comparable. In addition, the contrast-to-noise ratio of E was significantly better than that of D (p(D-E) < 0.001). Conclusions High temporal and spatial resolution can be obtained synchronously using a GRASP-VIBE technique for perfusion evaluation in head and neck MRI.
引用
收藏
页码:2263 / 2271
页数:9
相关论文
共 50 条
  • [1] Achieving high spatial and temporal resolution with perfusion MRI in the head and neck region using golden-angle radial sampling
    Andrea Tomppert
    Wolfgang Wuest
    Marco Wiesmueller
    Rafael Heiss
    Markus Kopp
    Armin M. Nagel
    Hayato Tomita
    Christian Meixner
    Michael Uder
    Matthias Stefan May
    European Radiology, 2021, 31 : 2263 - 2271
  • [2] 4D Golden-Angle Radial MRI at Subsecond Temporal Resolution
    Feng, Li
    NMR IN BIOMEDICINE, 2023, 36 (02)
  • [3] Golden-Angle Radial Sparse Parallel MRI: Combination of Compressed Sensing, Parallel Imaging, and Golden-Angle Radial Sampling for Fast and Flexible Dynamic Volumetric MRI
    Feng, Li
    Grimm, Robert
    Block, Kai Tobias
    Chandarana, Hersh
    Kim, Sungheon
    Xu, Jian
    Axel, Leon
    Sodickson, Daniel K.
    Otazo, Ricardo
    MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (03) : 707 - 717
  • [4] Golden-Angle Radial Sparse Parallel (GRASP) MRI differentiates head & neck paragangliomas from schwannomas
    Demerath, T.
    Blackham, K.
    Anastasopoulos, C.
    Block, K. T.
    Stieltjes, B.
    Schubert, T.
    MAGNETIC RESONANCE IMAGING, 2020, 70 : 73 - 80
  • [5] Impact of temporal resolution and motion correction for dynamic contrast-enhanced MRI of the liver using an accelerated golden-angle radial sequence
    Mansour, Rihab
    Antonacci, Alana
    Bilodeau, Laurent
    Romaguera, Liset
    Cerny, Milena
    Huet, Catherine
    Gilbert, Guillaume
    Tang, An
    Kadoury, Samuel
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (08):
  • [6] Dynamic Contrast-Enhanced MRI of the Prostate With High Spatiotemporal Resolution Using Compressed Sensing, Parallel Imaging, and Continuous Golden-Angle Radial Sampling: Preliminary Experience
    Rosenkrantz, Andrew B.
    Geppert, Christian
    Grimm, Robert
    Block, Tobias K.
    Glielmi, Christian
    Feng, Li
    Otazo, Ricardo
    Ream, Justin M.
    Romolo, Melanie Moccaldi
    Taneja, Samir S.
    Sodickson, Daniel K.
    Chandarana, Hersh
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2015, 41 (05) : 1365 - 1373
  • [7] High-speed whole-brain oximetry by golden-angle radial MRI
    Cao, Wen
    Chang, Yulin V.
    Englund, Erin K.
    Song, Hee Kwon
    Barhoum, Suliman
    Rodgers, Zachary B.
    Langham, Michael C.
    Wehrli, Felix W.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (01) : 217 - 223
  • [8] High-resolution CINE imaging of active guided knee motion using continuously acquired golden-angle radial MRI and rotary sensor information
    Aleksiev, Martin
    Kraemer, Martin
    Brisson, Nicholas M.
    Maggioni, Marta B.
    Duda, Georg N.
    Reichenbach, Juergen R.
    MAGNETIC RESONANCE IMAGING, 2022, 92 : 161 - 168
  • [9] Quantification of the Effects of Golden-Angle Radial K-Space Under Sampling on DCE-MRI Accuracy and Precision
    Knight, S.
    Clemence, M.
    Meaney, J.
    Fagan, A.
    MEDICAL PHYSICS, 2018, 45 (06) : E585 - E585
  • [10] Respiratory motion compensation using data binning in dynamic contrast enhanced golden-angle radial MRI
    Shahzadi, Iram
    Siddiqui, Muhammad Faisal
    Aslam, Ibtisam
    Omer, Hammad
    MAGNETIC RESONANCE IMAGING, 2020, 70 : 115 - 125