Total hip arthroplasty using a three-dimensional porous titanium acetabular cup: an examination of micromotion using subject-specific finite element analysis

被引:6
|
作者
Miyagawa, Takaki [1 ]
Matsumoto, Kazu [1 ]
Komura, Shingo [1 ]
Akiyama, Haruhiko [1 ]
机构
[1] Gifu Univ, Dept Orthopaed Surg, Sch Med, Yanagido 1-1, Gifu 5011194, Japan
关键词
Finite element analysis; Total hip arthroplasty; Three-dimensional porous structure; Radiolucent line;
D O I
10.1186/s12891-021-04174-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundWe investigated the mid-term clinical and radiological results of total hip arthroplasty (THA) using a three-dimensional (3D) porous titanium cup and analyzed the micromotion at the interface of the cup using subject-specific finite element (FE) analysis.MethodsWe evaluated 73 hips of 65 patients (6 men and 59 women; mean age at the time of surgery, 62.2years; range, 45-86years) who had undergone THA using a 3D porous titanium cup. Clinical evaluations were performed using the Japanese Orthopaedic Association (JOA) hip score system. We assessed the fixation of the acetabular component based on the presence of radiolucent lines and cup migration using anteroposterior radiographs. Subject-specific FE models were constructed from computed tomography data.ResultsThe JOA score improved from a preoperative mean of 52.2 (range, 23-82) to a mean of 87.8 (range, 71-100) at the final follow-up. None of the patients underwent revisions during the follow-up period. Radiolucent lines were observed in 26 cases (35.6%) and frequently appeared at DeLee and Charnley Zone 3. Following the FE analysis, the micromotion at DeLee and Charnley Zone 3 was significantly larger than that at Zone 2. Furthermore, micromotion was large in the groups in which radiolucent lines appeared at Zone 3.ConclusionsThe mid-term clinical outcome of THA using a 3D porous titanium cup was excellent. However, radiolucent lines frequently appeared at DeLee and Charnley Zone 3. FE analysis indicated that micromotion was large at the same site, strongly suggesting that it contributes to the emergence of radiolucent lines. The 3D porous titanium cups are useful in THA, and with improvements focused on micromotion, we anticipate better long-term outcomes.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Total hip arthroplasty using a three-dimensional porous titanium acetabular cup: an examination of micromotion using subject-specific finite element analysis
    Takaki Miyagawa
    Kazu Matsumoto
    Shingo Komura
    Haruhiko Akiyama
    BMC Musculoskeletal Disorders, 22
  • [2] Risk analysis of patients with an osteolytic acetabular defect after total hip arthroplasty using subject-specific finite-element modelling
    Munro, J. T.
    Millar, J. S.
    Fernandez, J. W.
    Walker, C. G.
    Howie, D. W.
    Shim, V. B.
    BONE & JOINT JOURNAL, 2018, 100B (11): : 1455 - 1462
  • [3] Total hip arthroplasty by using a cementless ultrashort stem: A subject-specific finite element analysis for a young patient clinical case
    Epasto, Gabriella
    Foti, Albina
    Guglielmino, Eugenio
    Rosa, Michele A.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (07) : 757 - 766
  • [4] Clinical and radiological outcomes of total hip arthroplasty in octogenarian patients using a three-dimensional porous titanium cup: a retrospective analysis in Japanese patients
    Taizo Kaneko
    Shinya Nakamura
    Kentaro Hayakawa
    Fumiaki Tokimura
    Tsuyoshi Miyazaki
    European Journal of Orthopaedic Surgery & Traumatology, 2023, 33 : 2361 - 2367
  • [5] Clinical and radiological outcomes of total hip arthroplasty in octogenarian patients using a three-dimensional porous titanium cup: a retrospective analysis in Japanese patients
    Kaneko, Taizo
    Nakamura, Shinya
    Hayakawa, Kentaro
    Tokimura, Fumiaki
    Miyazaki, Tsuyoshi
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2023, 33 (06): : 2361 - 2367
  • [6] Acute total hip arthroplasty in acetabular fractures using modern porous metal cup
    Malhotra, Rajesh
    Gautam, Deepak
    JOURNAL OF ORTHOPAEDIC SURGERY, 2019, 27 (02)
  • [7] Patient-specific three-dimensional evaluation of interface micromotion in two different short stem designs in cementless total hip arthroplasty: a finite element analysis
    Arata Kanaizumi
    Daisuke Suzuki
    Satoshi Nagoya
    Atsushi Teramoto
    Toshihiko Yamashita
    Journal of Orthopaedic Surgery and Research, 17
  • [8] Patient-specific three-dimensional evaluation of interface micromotion in two different short stem designs in cementless total hip arthroplasty: a finite element analysis
    Kanaizumi, Arata
    Suzuki, Daisuke
    Nagoya, Satoshi
    Teramoto, Atsushi
    Yamashita, Toshihiko
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [9] Is the acetabular cup orientation after total hip arthroplasty on a two-dimensional or three-dimensional model accurate?
    Kim, Yang-Hyun
    Cho, Kyung-Hwan
    Park, Yong-Gyu
    INTERNATIONAL ORTHOPAEDICS, 2015, 39 (04) : 819 - 820
  • [10] Is the acetabular cup orientation after total hip arthroplasty on a two-dimensional or three-dimensional model accurate?
    Yang-Hyun Kim
    Kyung-Hwan Cho
    Yong-Gyu Park
    International Orthopaedics, 2015, 39 : 819 - 820