Design and performance of a multi-pinhole collimation device for small animal imaging with clinical SPECT and SPECT-CT scanners

被引:52
|
作者
DiFilippo, Frank P. [1 ]
机构
[1] Cleveland Clin, Dept Nucl Med, Cleveland, OH 44195 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2008年 / 53卷 / 15期
关键词
D O I
10.1088/0031-9155/53/15/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A multi-pinhole collimation device is developed that uses the gamma camera detectors of a clinical SPECT or SPECT-CT scanner to produce high-resolution SPECT images. The device consists of a rotating cylindrical collimator having 22 tungsten pinholes with 0.9 mm diameter apertures and an animal bed inside the collimator that moves linearly to provide helical or ordered-subsets axial sampling. CT images also may be acquired on a SPECT-CT scanner for purposes of image co-registration and SPECT attenuation correction. The device is placed on the patient table of the scanner without attaching to the detectors or scanner gantry. The system geometry is calibrated in-place from point source data and is then used during image reconstruction. The SPECT imaging performance of the device is evaluated with test phantom scans. Spatial resolution from reconstructed point source images is measured to be 0.6 mm full width at half maximum or better. Micro-Derenzo phantom images demonstrate the ability to resolve 0.7 mm diameter rod patterns. The axial slabs of a Micro-Defrise phantom are visualized well. Collimator efficiency exceeds 0.05% at the center of the field of view, and images of a uniform phantom show acceptable uniformity and minimal artifact. The overall simplicity and relatively good imaging performance of the device make it an interesting low-cost alternative to dedicated small animal scanners.
引用
收藏
页码:4185 / 4201
页数:17
相关论文
共 50 条
  • [21] Design of a Multi-Pinhole Collimator in a Dual-Headed, Stationary, Small-Animal SPECT
    Shokouhi, S.
    Fritz, M. A.
    McDonald, B. S.
    Wilson, D. W.
    Metzler, S. D.
    Peterson, T. E.
    [J]. 2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2399 - 2402
  • [22] Small animal imaging with attenuation correction using clinical SPECT/CT scanners
    Boutchko, Rostyslav
    Balakrishnan, Karthikayan
    Reutter, Bryan W.
    Sauve, Anne
    Gullberg, Grant T.
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4294 - 4295
  • [23] Performance evaluation of a clinical multi-pinhole brain SPECT system: comparison with a conventional SPECT system
    Thilsing-Hansen, Kasper
    Simonsen, Jane
    Hvidsten, Svend
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [24] Quantification of the Multiplexing Effects in Multi-Pinhole Small Animal SPECT: A Simulation Study
    Mok, Greta S. P.
    Wang, Yuchuan
    Tsui, Benjamin M. W.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2636 - 2643
  • [25] 188Re image performance assessment using small animal multi-pinhole SPECT/PET/CT system
    Esquinas, Pedro L.
    Rodriguez-Rodriguez, Cristina
    De La Vega, Jose Carlos
    Bokharaei, Mehrdad
    Saatchi, Katayoun
    Shirmohammad, Maryam
    Hafeli, Urs O.
    Sossi, Vesna
    Celler, Anna
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 33 : 26 - 37
  • [26] Dedicated cardiac SPECT imaging with multi-pinhole collimators on a clinical scanner
    Liu, Hui
    Ma, Tianyu
    Wu, Jing
    Chen, Si
    Wang, Shi
    Wu, Zhaoxia
    Xia, Yan
    Liu, Yaqiang
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [27] Multi-Pinhole SPECT Imaging With Silicon Strip Detectors
    Peterson, Todd E.
    Shokouhi, Sepideh
    Furenlid, Lars R.
    Wilson, Donald W.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 646 - 652
  • [28] Triple and dual PET quantification of a Small-Animal Multi-Pinhole PET/SPECT/CT System
    Echegoyen, P.
    Cuenca Bandin, T.
    Ecay, M.
    Prieto, E.
    Collantes, M.
    Pareja, F.
    Simon Martinez, J.
    Penuelas, I.
    Marti-Climent, J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S144 - S144
  • [29] Design of an Optimized Multi-pinhole Collimator for Brain SPECT
    Chen, L.
    Mok, G. S.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S212 - S212
  • [30] Multi-Isotope Capabilities of a Small-Animal Multi-Pinhole SPECT System
    Lukas, Mathias
    Kluge, Anne
    Beindorff, Nicola
    Brenner, Winfried
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2020, 61 (01) : 152 - 161