Accuracy of iodine removal using dual-energy CT with or without a tin filter: an experimental phantom study

被引:6
|
作者
Kawai, Tatsuya [1 ]
Takeuchi, Mitsuru [1 ]
Hara, Masaki [1 ]
Ohashi, Kazuya [2 ]
Suzuki, Hirochika [3 ]
Yamada, Kiyotaka [3 ]
Sugimura, Yuya [3 ]
Shibamoto, Yuta [1 ]
机构
[1] Nagoya City Univ, Grad Sch Med Sci, Dept Radiol, Nagoya, Aichi, Japan
[2] Nagoya City Univ Hosp, Div Cent Radiol, Nagoya, Aichi, Japan
[3] Tsushima City Hosp, Div Radiol, Tsushima City, Japan
关键词
Dual-energy; CT; contrast agents; tin filter; technology assessments; LUNG PERFUSION; DIFFERENTIATION; FEASIBILITY; TOMOGRAPHY; STONES;
D O I
10.1177/0284185113485068
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The effects of a tin filter on virtual non-enhanced (VNE) images created by dual-energy CT have not been well evaluated. Purpose: To compare the accuracy of VNE images between those with and without a tin filter. Material and Methods: Two different types of columnar phantoms made of agarose gel were evaluated. Phantom A contained various concentrations of iodine (4.5-1590 HU at 120 kVp). Phantom B consisted of a central component (0, 10, 25, and 40 mgl/cm(3)) and a surrounding component (0, 50, 100, and 200 mgl/cm(3)) with variable iodine concentration. They were scanned by dual-source CT in conventional single-energy mode and dual-energy mode with and without a tin filter. CT values on each gel at the corresponding points were measured and the accuracy of iodine removal was evaluated. Results: On VNE images, the CT number of the gel of Phantom A fell within the range between -15 and +15 HU under 626 and 881 HU at single-energy 120 kVp with and without a tin filter, respectively. With attenuation over these thresholds, iodine concentration of gels was underestimated with the tin filter but overestimated without it. For Phantom B, the mean CT numbers on VNE images in the central gel component surrounded by the gel with iodine concentrations of 0, 50, 100, and 200 mgl/cm(3) were in the range of -19-+6 HU and 21-100 HU with and without the tin filter, respectively. Conclusion: Both with and without a tin filter, iodine removal was accurate under a threshold of iodine concentration. Although a surrounding structure with higher attenuation decreased the accuracy, a tin filter improved the margin of error.
引用
收藏
页码:954 / 960
页数:7
相关论文
共 50 条
  • [41] Accurate Multi-Material Decomposition in Dual-Energy CT: A Phantom Study
    Xue, Yi
    Jiang, Yangkang
    Yang, Chunlin
    Lyu, Qihui
    Wang, Jing
    Luo, Chen
    Zhang, Luhan
    Desrosiers, Catherine
    Feng, Kun
    Sun, Xiaonan
    Hu, Xiuhua
    Sheng, Ke
    Niu, Tianye
    IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, 2019, 5 (04) : 515 - 529
  • [42] Lung iodine mapping images of acute postpartum dyspnea without pulmonary thromboembolism using dual-energy CT
    Okada, Munemasa
    Nomura, Takafumi
    Kunihiro, Yoshie
    Takegami, Kazuki
    Uehara, Takuya
    Tokurei, Shogo
    ACTA RADIOLOGICA, 2023, 64 (08) : 2401 - 2408
  • [43] Dual-energy CT for the diagnosis of gout: an accuracy and diagnostic yield study
    Bongartz, Tim
    Glazebrook, Katrina N.
    Kavros, Steven J.
    Murthy, Naveen S.
    Merry, Stephen P.
    Franz, Walter B., III
    Michet, Clement J.
    Veetil, Barath M. Akkara
    Davis, John M., III
    Mason, Thomas G., II
    Warrington, Kenneth J.
    Ytterberg, Steven R.
    Matteson, Eric L.
    Crowson, Cynthia S.
    Leng, Shuai
    McCollough, Cynthia H.
    ANNALS OF THE RHEUMATIC DISEASES, 2015, 74 (06) : 1072 - 1077
  • [44] Iodine Quantification Using Dual-Energy Multidetector Computed Tomography Imaging Phantom Study Assessing the Impact of Iterative Reconstruction Schemes and Patient Habitus on Accuracy
    Feuerlein, Sebastian
    Heye, Tobias J.
    Bashir, Mustafa R.
    Boll, Daniel T.
    INVESTIGATIVE RADIOLOGY, 2012, 47 (11) : 656 - 661
  • [45] Influence of contrast medium on tophus detection using dual-energy CT: phantom study and clinical illustration
    Kotlyarov, Maximilian
    Mews, Jurgen
    Ulas, Sevtap Tugce
    Ziegeler, Katharina
    Hamm, Bernd
    Diekhoff, Torsten
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2023, 7 (01)
  • [46] Gastrointestinal Applications of Iodine Quantification Using Dual-Energy CT: A Systematic Review
    Xu, Jack Junchi
    Taudorf, Mikkel
    Ulriksen, Peter Sommer
    Achiam, Michael Patrick
    Resch, Timothy Andrew
    Nielsen, Michael Bachmann
    Lonn, Lars Birger
    Hansen, Kristoffer Lindskov
    DIAGNOSTICS, 2020, 10 (10)
  • [47] Characterisation of urinary stones in the presence of iodinated contrast medium using dual-energy CT: a phantom study
    Jia Wang
    Mingliang Qu
    Xinhui Duan
    Naoki Takahashi
    Akira Kawashima
    Shuai Leng
    Cynthia H. McCollough
    European Radiology, 2012, 22 : 2589 - 2596
  • [48] Influence of contrast medium on tophus detection using dual-energy CT: phantom study and clinical illustration
    Maximilian Kotlyarov
    Jürgen Mews
    Sevtap Tugce Ulas
    Katharina Ziegeler
    Bernd Hamm
    Torsten Diekhoff
    European Radiology Experimental, 7
  • [49] Characterisation of urinary stones in the presence of iodinated contrast medium using dual-energy CT: a phantom study
    Wang, Jia
    Qu, Mingliang
    Duan, Xinhui
    Takahashi, Naoki
    Kawashima, Akira
    Leng, Shuai
    McCollough, Cynthia H.
    EUROPEAN RADIOLOGY, 2012, 22 (12) : 2589 - 2596
  • [50] Performance of Dual-Energy CT with Tin Filter Technology for the Discrimination of Renal Cysts and Enhancing Masses
    Leschka, Sebastian
    Stolzmann, Paul
    Baumueller, Stephan
    Scheffel, Hans
    Desbiolles, Lotus
    Schmid, Bernhard
    Marincek, Borut
    Alkadhi, Hatem