Accuracy of iodine removal using dual-energy CT with or without a tin filter: an experimental phantom study

被引:6
|
作者
Kawai, Tatsuya [1 ]
Takeuchi, Mitsuru [1 ]
Hara, Masaki [1 ]
Ohashi, Kazuya [2 ]
Suzuki, Hirochika [3 ]
Yamada, Kiyotaka [3 ]
Sugimura, Yuya [3 ]
Shibamoto, Yuta [1 ]
机构
[1] Nagoya City Univ, Grad Sch Med Sci, Dept Radiol, Nagoya, Aichi, Japan
[2] Nagoya City Univ Hosp, Div Cent Radiol, Nagoya, Aichi, Japan
[3] Tsushima City Hosp, Div Radiol, Tsushima City, Japan
关键词
Dual-energy; CT; contrast agents; tin filter; technology assessments; LUNG PERFUSION; DIFFERENTIATION; FEASIBILITY; TOMOGRAPHY; STONES;
D O I
10.1177/0284185113485068
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The effects of a tin filter on virtual non-enhanced (VNE) images created by dual-energy CT have not been well evaluated. Purpose: To compare the accuracy of VNE images between those with and without a tin filter. Material and Methods: Two different types of columnar phantoms made of agarose gel were evaluated. Phantom A contained various concentrations of iodine (4.5-1590 HU at 120 kVp). Phantom B consisted of a central component (0, 10, 25, and 40 mgl/cm(3)) and a surrounding component (0, 50, 100, and 200 mgl/cm(3)) with variable iodine concentration. They were scanned by dual-source CT in conventional single-energy mode and dual-energy mode with and without a tin filter. CT values on each gel at the corresponding points were measured and the accuracy of iodine removal was evaluated. Results: On VNE images, the CT number of the gel of Phantom A fell within the range between -15 and +15 HU under 626 and 881 HU at single-energy 120 kVp with and without a tin filter, respectively. With attenuation over these thresholds, iodine concentration of gels was underestimated with the tin filter but overestimated without it. For Phantom B, the mean CT numbers on VNE images in the central gel component surrounded by the gel with iodine concentrations of 0, 50, 100, and 200 mgl/cm(3) were in the range of -19-+6 HU and 21-100 HU with and without the tin filter, respectively. Conclusion: Both with and without a tin filter, iodine removal was accurate under a threshold of iodine concentration. Although a surrounding structure with higher attenuation decreased the accuracy, a tin filter improved the margin of error.
引用
收藏
页码:954 / 960
页数:7
相关论文
共 50 条
  • [31] Differentiation of heroin and cocaine using dual-energy CT—an experimental study
    Jochen Grimm
    Ramona Wudy
    Edvard Ziegeler
    Stefan Wirth
    Michael Uhl
    Maximilian F. Reiser
    Michael Scherr
    International Journal of Legal Medicine, 2014, 128 : 475 - 482
  • [32] Dual-energy CT iodine overlay technique for characterization of renal masses as cyst or solid: a phantom feasibility study
    Brown, C. L.
    Hartman, R. P.
    Dzyubak, O. P.
    Takahashi, N.
    Kawashima, A.
    McCollough, C. H.
    Bruesewitz, M. R.
    Primak, A. M.
    Fletcher, J. G.
    EUROPEAN RADIOLOGY, 2009, 19 (05) : 1289 - 1295
  • [33] Dual-energy CT iodine overlay technique for characterization of renal masses as cyst or solid: a phantom feasibility study
    C. L. Brown
    R. P. Hartman
    O. P. Dzyubak
    N. Takahashi
    A. Kawashima
    C. H. McCollough
    M. R. Bruesewitz
    A. M. Primak
    J. G. Fletcher
    European Radiology, 2009, 19
  • [34] Quantification of Metal Artifact Reduction of Dual-Energy CT: A Phantom Study
    Li, K.
    Leung, W.
    MEDICAL PHYSICS, 2017, 44 (06) : 3036 - 3036
  • [35] Accuracy and precision of volumetric bone mineral density assessment using dual-source dual-energy versus quantitative CT: a phantom study
    Koch, Vitali
    Hokamp, Nils Grosse
    Albrecht, Moritz H.
    Gruenewald, Leon D.
    Yel, Ibrahim
    Borggrefe, Jan
    Wesarg, Stefan
    Eichler, Katrin
    Burck, Iris
    Gruber-Rouh, Tatjana
    Lenga, Lukas
    Vogl, Thomas J.
    Martin, Simon S.
    Wichmann, Julian L.
    Hammerstingl, Renate M.
    Alizadeh, Leona S.
    Mader, Christoph
    Huizinga, Nicole A.
    D'Angelo, Tommaso
    Ascenti, Giorgio
    Mazziotti, Silvio
    Booz, Christian
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2021, 5 (01)
  • [36] Accuracy and precision of volumetric bone mineral density assessment using dual-source dual-energy versus quantitative CT: a phantom study
    Vitali Koch
    Nils Große Hokamp
    Moritz H. Albrecht
    Leon D. Gruenewald
    Ibrahim Yel
    Jan Borggrefe
    Stefan Wesarg
    Katrin Eichler
    Iris Burck
    Tatjana Gruber-Rouh
    Lukas Lenga
    Thomas J. Vogl
    Simon S. Martin
    Julian L. Wichmann
    Renate M. Hammerstingl
    Leona S. Alizadeh
    Christoph Mader
    Nicole A. Huizinga
    Tommaso D’Angelo
    Giorgio Ascenti
    Silvio Mazziotti
    Christian Booz
    European Radiology Experimental, 5
  • [37] Accuracy of Dual-Energy CT Virtual Non-Contrast/unenhanced and Material Density Images: A Phantom Study
    Li, B.
    Pomerleau, M.
    Gupta, A.
    Soto, J.
    Anderson, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E511 - E512
  • [38] Spectral Imaging of Iodine and Gadolinium Nanoparticles Using Dual-Energy CT
    Badea, C. T.
    Holbrook, M.
    Clark, D. P.
    Ghaghada, K.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [39] Differentiation of heroin and cocaine using dual-energy CT-an experimental study
    Grimm, Jochen
    Wudy, Ramona
    Ziegeler, Edvard
    Wirth, Stefan
    Uhl, Michael
    Reiser, Maximilian F.
    Scherr, Michael
    INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 2014, 128 (03) : 475 - 482
  • [40] Calcium scoring with dual-energy CT in men and women: an anthropomorphic phantom study
    Li, Qin
    Liu, Songtao
    Myers, Kyle
    Gavrielides, Marios A.
    Zeng, Rongping
    Sahiner, Berkman
    Petrick, Nicholas
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783