Accuracy of iodine removal using dual-energy CT with or without a tin filter: an experimental phantom study

被引:6
|
作者
Kawai, Tatsuya [1 ]
Takeuchi, Mitsuru [1 ]
Hara, Masaki [1 ]
Ohashi, Kazuya [2 ]
Suzuki, Hirochika [3 ]
Yamada, Kiyotaka [3 ]
Sugimura, Yuya [3 ]
Shibamoto, Yuta [1 ]
机构
[1] Nagoya City Univ, Grad Sch Med Sci, Dept Radiol, Nagoya, Aichi, Japan
[2] Nagoya City Univ Hosp, Div Cent Radiol, Nagoya, Aichi, Japan
[3] Tsushima City Hosp, Div Radiol, Tsushima City, Japan
关键词
Dual-energy; CT; contrast agents; tin filter; technology assessments; LUNG PERFUSION; DIFFERENTIATION; FEASIBILITY; TOMOGRAPHY; STONES;
D O I
10.1177/0284185113485068
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The effects of a tin filter on virtual non-enhanced (VNE) images created by dual-energy CT have not been well evaluated. Purpose: To compare the accuracy of VNE images between those with and without a tin filter. Material and Methods: Two different types of columnar phantoms made of agarose gel were evaluated. Phantom A contained various concentrations of iodine (4.5-1590 HU at 120 kVp). Phantom B consisted of a central component (0, 10, 25, and 40 mgl/cm(3)) and a surrounding component (0, 50, 100, and 200 mgl/cm(3)) with variable iodine concentration. They were scanned by dual-source CT in conventional single-energy mode and dual-energy mode with and without a tin filter. CT values on each gel at the corresponding points were measured and the accuracy of iodine removal was evaluated. Results: On VNE images, the CT number of the gel of Phantom A fell within the range between -15 and +15 HU under 626 and 881 HU at single-energy 120 kVp with and without a tin filter, respectively. With attenuation over these thresholds, iodine concentration of gels was underestimated with the tin filter but overestimated without it. For Phantom B, the mean CT numbers on VNE images in the central gel component surrounded by the gel with iodine concentrations of 0, 50, 100, and 200 mgl/cm(3) were in the range of -19-+6 HU and 21-100 HU with and without the tin filter, respectively. Conclusion: Both with and without a tin filter, iodine removal was accurate under a threshold of iodine concentration. Although a surrounding structure with higher attenuation decreased the accuracy, a tin filter improved the margin of error.
引用
收藏
页码:954 / 960
页数:7
相关论文
共 50 条
  • [1] Differentiation of Kidney Stones Using Dual-Energy CT With and Without a Tin Filter
    Fung, George S. K.
    Kawamoto, Satomi
    Matlaga, Brian R.
    Taguchi, Katsuyuki
    Zhou, Xiaodong
    Fishman, Elliot K.
    Tsui, Benjamin M. W.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (06) : 1380 - 1386
  • [2] Multivendor Comparison of Quantification Accuracy of Iodine Concentration and Attenuation Measurements by Dual-Energy CT: A Phantom Study
    Chen, Yong
    Zhong, Jingyu
    Wang, Lan
    Shi, Xiaomeng
    Chang, Rui
    Fan, Jing
    Jiang, Jiang
    Xia, Yihan
    Yan, Fuhua
    Yao, Weiwu
    Zhang, Huan
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (05) : 827 - 839
  • [3] Evaluation of Intra-Manufacturer Dual-Energy CT Iodine Quantification Accuracy: Phantom Studies
    Zhao, A.
    Borhani, A.
    Catania, R.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Re: Differentiation of Kidney Stones Using Dual-Energy CT With and Without a Tin Filter Editorial Comment
    Assimos, Dean
    JOURNAL OF UROLOGY, 2013, 189 (01): : 172 - 173
  • [5] Quantitative accuracy and dose efficiency of dual-contrast imaging using dual-energy CT: a phantom study
    Ren, Liqiang
    Rajendran, Kishore
    McCollough, Cynthia H.
    Yu, Lifeng
    MEDICAL PHYSICS, 2020, 47 (02) : 441 - 456
  • [6] Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model
    James D. Koonce
    Rozemarijn Vliegenthart
    U. Joseph Schoepf
    Bernhard Schmidt
    Amy E. Wahlquist
    Paul J. Nietert
    Gorka Bastarrika
    Thomas G. Flohr
    Felix G. Meinel
    European Radiology, 2014, 24 : 512 - 518
  • [7] Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study
    Jacobsen, Megan C.
    Schellingerhout, Dawid
    Wood, Cayla A.
    Tamm, Eric P.
    Godoy, Myrna C.
    Sun, Jia
    Cody, Dianna D.
    RADIOLOGY, 2018, 287 (01) : 224 - 234
  • [8] Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model
    Koonce, James D.
    Vliegenthart, Rozemarijn
    Schoepf, U. Joseph
    Schmidt, Bernhard
    Wahlquist, Amy E.
    Nietert, Paul J.
    Bastarrika, Gorka
    Flohr, Thomas G.
    Meinel, Felix G.
    EUROPEAN RADIOLOGY, 2014, 24 (02) : 512 - 518
  • [9] Iodine Quantification With Dual-Energy CT: Phantom Study and Preliminary Experience With Renal Masses
    Chandarana, Hersh
    Megibow, Alec J.
    Cohen, Benjamin A.
    Srinivasan, Ramya
    Kim, Danny
    Leidecker, Christianne
    Macari, Michael
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) : W693 - W700
  • [10] Dual-Energy CT Iodine Accuracy Across Vendors and Platforms
    Jacobsen, M.
    Wood, C.
    Cody, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3835 - 3836