Glaucoma diagnosis optic disc analysis comparing Cirrus spectral domain optical coherence tomography and Heidelberg retina tomograph II

被引:9
|
作者
Shin, Hye Young [1 ]
Park, Hae-Young Lopilly [1 ]
Jung, Kyoung In [1 ]
Park, Chan Kee [1 ]
机构
[1] Catholic Univ Korea, Seoul St Marys Hosp, Coll Med, Dept Ophthalmol & Visual Sci, Seoul 137701, South Korea
关键词
Cirrus spectral domain optical coherence tomography; Glaucoma; Optic nerve head analysis; Confocal scanning laser ophthalmoscopy imaging; NERVE-FIBER LAYER; THICKNESS MEASUREMENTS; TIME-DOMAIN; ABILITY; OCT;
D O I
10.1007/s10384-012-0205-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To evaluate the diagnostic ability of optic nerve head parameters, measured by Cirrus spectral domain optical coherence tomography (SD OCT) and Heidelberg retina tomograph II (HRT II) to detect concomitant glaucoma. Cirrus OCT and HRT examinations of 62 eyes of 62 patients (32 with no perimetric glaucoma and 30 with concomitant perimetric glaucoma) evaluated between August 2010 and December 2010 were retrospectively analyzed and compared with regard to the optic disc morphometric parameter, disc area, rim area, cup-to-disc ratio (CDR), and cup volume. Receiver operating characteristic curves were constructed for the parameters and areas under the curves (AUCs) were compared. All parameters except disc area were significantly different between Cirrus OCT and HRT. Average cup-to-disc ratio (CDR), vertical CDR, and cup volume were greater when measured by OCT compared with HRT II (P = 0.002, P < 0.001, and P < 0.001, respectively). Rim area was smaller by OCT than by HRT II (P < 0.001). Of the parameters evaluated, rim area (0.938), average CDR (0.865), and vertical CDR (0.897) had higher AUCs with OCT than with HRT II. Glaucoma diagnostic capability using the AUC was greater for OCT than for the HRT. Optic nerve head parameters measured by Cirrus OCT seem to be useful in differentiating glaucomatous optic nerve heads. The two types of instrument compared here should not be used interchangeably to obtain measurements of the optic disc for diagnosis of glaucoma.
引用
收藏
页码:41 / 46
页数:6
相关论文
共 50 条
  • [1] Glaucoma diagnosis optic disc analysis comparing Cirrus spectral domain optical coherence tomography and Heidelberg retina tomograph II
    Hye Young Shin
    Hae-Young Lopilly Park
    Kyoung In Jung
    Chan Kee Park
    Japanese Journal of Ophthalmology, 2013, 57 : 41 - 46
  • [2] Comparison of Spectral-Domain Optical Coherence Tomography and Heidelberg Retina Tomograph III Optic Nerve Head Parameters in Glaucoma
    Seymenoglu, Goktug
    Baser, Esin
    Ozturk, Bilge
    OPHTHALMOLOGICA, 2013, 229 (02) : 101 - 105
  • [3] Optic Disc Imaging with Spectral-Domain Optical Coherence Tomography Variability and Agreement Study with Heidelberg Retinal Tomograph
    Yang, Bingzhi
    Ye, Cong
    Yu, Marco
    Liu, Shu
    Lam, Dennis Shun Chiu
    Leung, Christopher Kai-shun
    OPHTHALMOLOGY, 2012, 119 (09) : 1852 - 1857
  • [4] ROC analysis of Heidelberg Retina Tomograph optic disc shape measures in glaucoma
    Iester, M
    Mikelberg, FS
    Swindale, NV
    Drance, SM
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 1997, 32 (06): : 382 - 388
  • [5] Comparison of optic nerve head parameters using Heidelberg Retina Tomograph 3 and spectral-domain optical coherence tomography
    Sato, Shino
    Hirooka, Kazuyuki
    Baba, Tetsuya
    Shiraga, Fumio
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (07): : 721 - 726
  • [6] Automated analysis of heidelberg retina tomograph optic disc images by glaucoma probability score
    Coops, Annemiek
    Henson, David Barry
    Kwartz, Anna J.
    Artes, Paul Habib
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (12) : 5348 - 5355
  • [7] Comparison of Optic Disc Morphology and Its Correlation With Visual Field Test Results Between Spectral-Domain Optical Coherence Tomography, Heidelberg Retina Tomograph II and Stereophotographs in Glaucomatous Eyes
    Saito, H.
    Tomidokoro, A.
    Araie, M.
    Iwase, A.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [8] Optic disc classification by the Heidelberg Retina Tomograph and by physicians with varying experience of glaucoma
    Andersson, S.
    Heijl, A.
    Bengtsson, B.
    EYE, 2011, 25 (11) : 1401 - 1407
  • [9] Optic disc classification by the Heidelberg Retina Tomograph and by physicians with varying experience of glaucoma
    S Andersson
    A Heijl
    B Bengtsson
    Eye, 2011, 25 : 1401 - 1407
  • [10] Influence of Disc Size on the Diagnostic Accuracy of Cirrus Spectral-Domain Optical Coherence Tomography in Glaucoma
    Sato, Shino
    Ukegawa, Kaori
    Nitta, Eri
    Hirooka, Kazuyuki
    JOURNAL OF OPHTHALMOLOGY, 2018, 2018