Detection of γ-rays with a 3.5 1 liquid xenon ionization chamber triggered by the primary scintillation light

被引:10
|
作者
Aprile, E [1 ]
Bolotnikov, A [1 ]
Chen, DL [1 ]
Muhkerjee, R [1 ]
Xu, F [1 ]
机构
[1] Columbia Univ, Columbia Astrophys Lab, New York, NY 10027 USA
基金
美国国家航空航天局;
关键词
D O I
10.1016/S0168-9002(01)01204-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A gridded ionization chamber with a drift length of 4.5 cm and a total volume of 3.5 1, was operated with high-purity liquid xenon and extensively tested with gamma-rays from Cs-137, Na-22 and Co-60 radioactive sources. An electron lifetime in excess of I ins was inferred from two independent measurements. The electric Field dependence of the collected charge and energy resolution was studied in the range 0.1-4 kV/cm, for different gamma-ray energies, With an electric field of 4 kV/cm, the spectral performance of the detector is consistent with an energy resolution of 5.9% at 1 MeV, scaling with energy as E-0.5 The chamber was also used to detect the primary scintillation light produced by gamma-ray interactions in liquid xenon. The light signal was successfully used to trigger the acquisition of the charge signal with a FADC readout. A trigger efficiency of similar to85% was measured at 662 keV. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:636 / 650
页数:15
相关论文
共 50 条
  • [31] Development of PPDs to detect scintillation light from liquid xenon
    Iwamoto, Toshiyuki
    Bai, Xue
    Chiba, Teppei
    Fujii, Yuki
    Haruyama, Tomiyoshi
    Kaneko, Daisuke
    Maki, Akihiro
    Mihara, Satoshi
    Mori, Toshinori
    Natori, Hiroaki
    Nishiguchi, Hajime
    Ootani, Wataru
    Sawada, Ryu
    Uchiyama, Yusuke
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 1589 - 1594
  • [32] Time behaviour of the scintillation light in mixtures of liquid argon and xenon
    Universita di Padova, Padova, Italy
    Nucl Instrum Methods Phys Res Sect A, 3 (475-478):
  • [33] Performance of a chamber for studying the liquid xenon response to γ-rays and nuclear recoils
    Neves, F
    Chepel, V
    Solovov, V
    Pereira, A
    Lopes, MI
    da Cunha, JP
    Mendes, P
    Lindote, A
    Silva, CP
    Marques, RF
    Policarpo, AJPL
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (06) : 2793 - 2800
  • [34] Detection of scintillation light in coincidence with ionizing tracks in a liquid argon time projection chamber
    Cennini, P.
    Revol, J.P.
    Rubbia, C.
    Sergiampietri, F.
    Bueno, A.
    Campanelli, M.
    Goudsmit, P.
    Rubbia, A.
    Periale, L.
    Suzuki, S.
    Chen, C.
    Chen, Y.
    He, K.
    Huang, X.
    Li, Z.
    Lu, F.
    Ma, J.
    Xu, G.
    Xu, Z.
    Zhang, C.
    Zhang, Q.
    Zheng, S.
    Cavanna, F.
    Mazza, D.
    Mortari, G. Piano
    Petrera, S.
    Rossi, C.
    Mannocchi, G.
    Picchi, P.
    Arneodo, F.
    De Mitri, I.
    Palamara, O.
    Cavalli, D.
    Ferrari, A.
    Sala, P.
    Di Tigliole, A. Borio
    Cesana, A.
    Terrani, M.
    Zavattari, C.
    Baibussinov, S.
    Bettini, A.
    Carpanese, C.
    Centro, S.
    Favaretto, D.
    Pascoli, D.
    Pepato, A.
    Pietropaolo, F.
    Ventura, S.
    Benetti, P.
    Calligarich, E.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999, 432 (2-3): : 240 - 248
  • [35] TIME RESOLUTION OF IONIZATION-CHAMBER FILLED WITH LIQUID KRYPTON AND XENON
    LEBEDEV, PK
    MURAVIEV, SV
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 327 (01): : 138 - 140
  • [36] Pulse shape analysis in the liquid xenon multiwire ionization chamber for PET
    Crespo, P
    Chepel, V
    Lopes, MI
    Janeiro, L
    Marques, RF
    Policarpo, AJPL
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 561 - 567
  • [37] DETECTION OF SOFT X-RAYS WITH A XENON PROPORTIONAL SCINTILLATION-COUNTER
    POLICARPO, AJ
    ALVES, MAF
    SALETE, M
    LEITE, SCP
    DOSSANTO.MC
    NUCLEAR INSTRUMENTS & METHODS, 1974, 118 (01): : 221 - 226
  • [38] Simultaneous reconstruction of scintillation light and ionization charge produced by 511 keV photons in liquid xenon: Potential application to PET
    Amaudruz, P.
    Bryman, D.
    Kurchaninov, L.
    Lu, P.
    Marshall, C.
    Martin, J. P.
    Muennich, A.
    Retiere, F.
    Sher, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 607 (03): : 668 - 676
  • [39] A PHOTO-IONIZATION DETECTOR FOR THE DETECTION OF XENON LIGHT
    ANDERSON, DF
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) : 842 - 848
  • [40] A Multi-Pixel Photon Counter detector prototype for direct detection of scintillation light in liquid xenon
    Di Giovanni, Adriano
    CONFERENCE ON NEUTRINO AND NUCLEAR PHYSICS (CNNP2017), 2018, 1056