Design of linear anti-scatter grid geometry with optimum performance for screen-film and digital mammography systems

被引:12
|
作者
Khodajou-Chokami, H. [1 ]
Sohrabpour, M. [1 ]
机构
[1] Sharif Univ Technol, Dept Energy Engn, Tehran, Iran
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 15期
关键词
mammography; Monte Carlo; anti-scatter grid; simulation; ANTISCATTER GRIDS; SIMULATION; RADIOLOGY; REMOVAL;
D O I
10.1088/0031-9155/60/15/5753
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A detailed 3D Monte Carlo simulation of the grid geometrical parameters in screen-film mammography (SFM) and digital mammography (DM) systems has been performed. A combination of IEC 60627:2013 international standard conditions and other more clinically relevant parameters were used for this simulation. Accuracy of our results has been benchmarked with previously published data and good agreement has been obtained. Calculations in a wide range of linear anti-scatter grid geometries have been carried out. The evaluated parameters for the SFM system were the Bucky factor (BF) and contrast improvement factor (CIF) and for the DM system it was signal difference-to-noise ratio improvement factor (SIF). The CIF parameter was chosen to be nearly the same as the commercial grade, the BF and SIF were significantly improved compared to commercial grids in use today. Our optimized grid parameters for the SFM system were lead strip thickness d = 12 mu m, grid ratio r = 5 and strip density N = 65 lines/cm. And for the DM system these parameters were d = 5 mu m, r = 3 and N = 100 lines/cm. Both optimized grid sets have thinner d and higher N compared to the commercial grids.
引用
收藏
页码:5753 / 5765
页数:13
相关论文
共 20 条
  • [1] A comparison of the performance of modern screen-film and digital mammography systems
    Monnin, P
    Gutierrez, D
    Bulling, S
    Lepori, D
    Valley, JF
    Verdun, FR
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (11): : 2617 - 2631
  • [2] Effect of the anti-scatter grid and target/filters in full-field digital mammography
    Chakraborty, DP
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 878 - 885
  • [3] Experimental Evaluation of a New Anti-Scatter Grid for Digital Radiographic Systems
    Fetterly, K.
    Schueler, B.
    MEDICAL PHYSICS, 2008, 35 (06) : 2677 - +
  • [4] Clinical performance comparison of full-field digital mammography to screen-film mammography
    Berns, EA
    Hendrick, RE
    RADIOLOGY, 2001, 221 : 283 - 283
  • [5] Performance comparison of full-field digital mammography to screen-film mammography in clinical practice
    Berns, EA
    Hendrick, RE
    Cutter, GR
    MEDICAL PHYSICS, 2002, 29 (05) : 830 - 834
  • [6] Screen-film and digital mammography in the Finnish breast cancer screening programme: a randomised design
    不详
    EJC SUPPLEMENTS, 2006, 4 (02): : 57 - 57
  • [7] A comparison of the performance of new screen-film and digital mammography systems - art. no. 61423B
    Monnin, P.
    Gutierrez, D.
    Castella, C.
    Lepori, D.
    Verdun, F. R.
    Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : B1423 - B1423
  • [8] A comparison of physical image quality of two hologic digital mammography systems that utilise linear and 2D anti-scatter grids
    Moore, Craig S.
    Wood, Tim J.
    Saunderson, John R.
    Beavis, Andrew W.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (05)
  • [9] Slot scanning versus anti-scatter grid in digital mammography-comparison of low contrast performance using contrast-detail measurement
    Lai, CJ
    Shaw, CC
    Geiser, W
    Kappadath, SC
    Liu, XM
    Wang, TP
    Tu, SJ
    Altunbas, MC
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 734 - 742
  • [10] Investigation of optimum anti-scatter grid selection for digital radiography: Physical imaging properties and detectability of low-contrast signals
    Tanaka N.
    Naka K.
    Saito A.
    Morishita J.
    Toyofuku F.
    Ohki M.
    Higashida Y.
    Radiological Physics and Technology, 2013, 6 (1) : 54 - 60