Investigation of optimum anti-scatter grid selection for digital radiography: Physical imaging properties and detectability of low-contrast signals

被引:13
|
作者
Tanaka N. [1 ]
Naka K. [1 ,4 ]
Saito A. [2 ,5 ]
Morishita J. [3 ]
Toyofuku F. [3 ]
Ohki M. [3 ]
Higashida Y. [3 ]
机构
[1] Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku
[2] Department of Health Sciences, School of Medicine, Kyushu University, 3-1-1 Maidashi, Higashi-ku
[3] Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku
[4] Toshiba Medical Systems Corporation, 1385 Shimoishigami
[5] National Hospital Organization, Okayama Medical Center, 1711-1 Tamasu, Kita-ku
关键词
Anti-scatter grid; Computed radiography; Contrast-detail (CD) phantom; Low-contrast detectability; Noise-equivalent quanta (NEQ); Physical imaging properties;
D O I
10.1007/s12194-012-0169-y
中图分类号
学科分类号
摘要
Our aim in this study was to evaluate the effect of the grid variations on the imaging performance for a computed radiographic system under identical exposure condition. Digital radiographies using a 20-cm Lucite phantom were performed without grid and with grid ratios of 5:1, 8:1, 10:1, 12:1, and 14:1. The scatter fraction, the incident dose to the image receptor, the Wiener spectrum (WS), and the noise-equivalent quanta (NEQ) were measured. Visibility of low-contrast signals was evaluated using a contrast-detail phantom. The scatter fractions decreased considerably with an increase in the grid ratio. On the other hand, the WSs were increased (the noise property deteriorated) as the grid ratio increased due to a decreased incident dose to the image receptor under the identical exposure condition. The NEQs were improved as the grid ratio increased. The high grid ratios provided higher low-contrast detectability compared to the low grid ratios. Our results indicated that the removal of scattered radiation was very effective in improvement of the NEQ in the digital system under the identical exposure condition. © 2012 Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:54 / 60
页数:6
相关论文
共 25 条
  • [1] EFFECT OF PIXEL SIZE ON DETECTABILITY OF LOW-CONTRAST SIGNALS IN DIGITAL RADIOGRAPHY
    GIGER, ML
    DOI, K
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05): : 966 - 975
  • [2] Guidelines for anti-scatter grid use in pediatric digital radiography
    Shannon Fritz
    A. Kyle Jones
    Pediatric Radiology, 2014, 44 : 313 - 321
  • [3] Guidelines for anti-scatter grid use in pediatric digital radiography
    Fritz, Shannon
    Jones, A. Kyle
    PEDIATRIC RADIOLOGY, 2014, 44 (03) : 313 - 321
  • [4] Digital selenium radiography: anti-scatter grid for chest radiography in a clinical study
    Bernhardt, TM
    Rapp-Bernhardt, U
    Hausmann, T
    Reichel, G
    Krause, UW
    Doehring, W
    BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (873): : 963 - 968
  • [5] Dual harmonic anti-scatter grid artefact removal for digital radiography
    Ostojic, Vladimir S.
    Starcevic, Dorde S.
    Petrovic, Vladimir S.
    2017 25TH TELECOMMUNICATION FORUM (TELFOR), 2017, : 326 - 329
  • [6] Investigating the Replacement of the Physical Anti-scatter Grid with Digital Image Processing
    Tromans, Christopher E.
    Diffey, Jennifer
    Brady, Michael
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 205 - +
  • [7] Radiation dose considerations in digital radiography with an anti-scatter grid: A study using adult and pediatric phantoms
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Kitao, Azusa
    Matsubara, Takashi
    Sugiura, Takumi
    Kobayashi, Tomohiro
    Kobayashi, Satoshi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09):
  • [8] INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .8. DETECTION OF SIMULATED LOW-CONTRAST OBJECTS IN DIGITAL SUBTRACTION ANGIOGRAPHIC IMAGES
    OHARA, K
    CHAN, HP
    DOI, K
    GIGER, ML
    FUJITA, H
    MEDICAL PHYSICS, 1986, 13 (03) : 304 - 311
  • [9] Slot scanning versus anti-scatter grid in digital mammography-comparison of low contrast performance using contrast-detail measurement
    Lai, CJ
    Shaw, CC
    Geiser, W
    Kappadath, SC
    Liu, XM
    Wang, TP
    Tu, SJ
    Altunbas, MC
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 734 - 742
  • [10] Dose efficiency and low-contrast detectability of an amorphous silicon x-ray detector for digital radiography
    Aufrichtig, R
    Xue, P
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09): : 2653 - 2669