Traceable S-parameter measurements up to 90 GHz in 1.35 mm coaxial

被引:3
|
作者
Stokes, Daniel [1 ]
Gellersen, Frauke [2 ]
Allal, Djamel [3 ]
Skinner, James [1 ]
Phung, Gia Ngoc [2 ]
Kuhlmann, Karsten [2 ]
机构
[1] Natl Phys Lab NPL, Teddington TW11, England
[2] Phys Tech Bundesanstalt PTB, D-38116 Braunschweig, Germany
[3] Bur Natl Metrol LNE, F-75014 Paris, France
关键词
S-parameters; mm-wave; coaxial; measurement; traceability; intercomparison;
D O I
10.1088/1361-6501/acc04c
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
As part of a joint program three National Measurement Institutes around Europe-the National Physical Laboratory in the UK, the Physikalisch-Technische Bundesanstalt in Germany and the Laboratoire National de metrologie et d'essais in France-have each established new capabilities for the traceable measurement of S-parameters up to 90 GHz in the new 1.35 mm (E band) coaxial connector. Each institution has established their own measurement systems to access the full frequency band for the connector, utilising a variety of commercially available equipment and measurement setups. Each have also developed their own calibration techniques and verification methods using different vector network analyser calibration routines. These include optimised multi-line Through Reflect Line, multiple Offset Shorts Through and Short Open Load Reciprocal (SOLR) schemes. Uncertainties for each of these setups and calibration schemes were determined utilising theoretical models of the systems and calibration standards, or through verification methods, with different effects being derived from the dimensional characteristics for each of the different standards or verification artifacts. The different dimensional systems used to make these measurements were also developed and established previously as part of the same program. The last part of this exercise was a three-way intercomparison between the laboratories. The comparison involved the measurement of a set of four traveling standards, including both 1-port and 2-port as well as plug (male) and jack (female) devices, each with different reflection and transmission characteristics to better test and demonstrate the equivalence of the capabilities.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Characterization of Coaxial Adapters for S-parameter Measurements
    Hoffmann, Johannes Paul
    Leuchtman, Pascal
    Kretz, Adjan
    Ruefenacht, Juerg
    Vahldieck, Ruediger
    [J]. 2008 EUROPEAN MICROWAVE CONFERENCE, VOLS 1-3, 2008, : 401 - +
  • [2] Development and Traceable Measurements of 0.800 mm Coaxial Airline: Broadband Standard to 145 GHz
    Roberts, Tom
    Lee, Yeou Song
    Martens, Jon
    Oldfield, William
    [J]. 2013 81ST ARFTG MICROWAVE MEASUREMENT CONFERENCE (ARFTG), 2013,
  • [3] Broadband Coaxial S-Parameter Measurements for Cryogenic Quantum Technologies
    Shin, Sang-Hee
    Stanley, Manoj
    Skinner, James
    de Graaf, Sebastian E.
    Ridler, Nick M.
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2024, 72 (04) : 2193 - 2201
  • [4] On-wafer CPW standards for S-parameter measurements of balanced circuits up to 40 GHz
    Thi Dao Pham
    Allal, Djamel
    Ziade, Francois
    Bergeault, Eric
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [5] A 110 GHz scalable FET model based on 50 GHz S-parameter measurements
    Cidronali, A
    Collodi, G
    Santarelli, A
    Vannini, G
    [J]. 2000 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM DIGEST, VOLS 1-3, 2000, : 1377 - 1380
  • [6] Design of Verification Standards for S-parameter Measurements from 110 GHz to 170 GHz
    Liu Jie
    Yang Chuntao
    Chen Ting
    Zhang Na
    Cheng Chunyue
    Ma Hongmei
    [J]. 2020 XXXIIIRD GENERAL ASSEMBLY AND SCIENTIFIC SYMPOSIUM OF THE INTERNATIONAL UNION OF RADIO SCIENCE, 2020,
  • [7] Establishing Traceability for On-Wafer S-Parameter Measurements of Membrane Technology Devices up to 110 GHz
    Arz, Uwe
    Zinal, Sherko
    Probst, Thorsten
    Hechtfischer, Gerd
    Schmueckle, Franz-Josef
    Heinrich, Wolfgang
    [J]. 2017 90TH ARFTG MICROWAVE MEASUREMENT SYMPOSIUM (ARFTG), 2017,
  • [8] On-Wafer Coplanar Waveguide Standards for S-Parameter Measurements of Balanced Circuits Up to 40 GHz
    Thi Dao Pham
    Allal, Djamel
    Ziade, Francois
    Bergeault, Eric
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2019, 68 (06) : 2160 - 2167
  • [9] DEVELOPMENT OF A 2.4 mm COAXIAL MICROCALORIMETER FOR POWER MEASUREMENTS UP TO 50 GHz
    Bourghes, M.
    Allal, D.
    Belieres, D.
    Kazemipour, A.
    Litwin, A.
    [J]. 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 324 - 325
  • [10] On peculiarities of S-parameter measurements
    Rolain, Yves
    Van Moer, Wendy
    Jargon, Jeffrey A.
    DeGroot, Donald C.
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (05) : 1967 - 1972