Usefulness of modular neck adapter in partial hip revision

被引:0
|
作者
Pautasso, Andrea [1 ]
Bardellini, Giuseppe [2 ]
Stissi, Placido [1 ]
D'Angelo, Fabio [1 ,3 ]
机构
[1] ASST Sette Laghi, Circolo Hosp & Macchi Fdn Varese, Univ Ctr, Orthopaed & Traumatol Dept, Viale Luigi Borri 57, I-21100 Varese, Italy
[2] Univ Studies Insubria, Orthopaed & Traumatol Dept, Varese, Italy
[3] Univ Studies Insubria, Dept Biotechnol & Life Sci DBSV, Varese, Italy
来源
ANNALS OF JOINT | 2023年 / 8卷
关键词
Hip revision arthroplasty; head-neck adapter; total hip arthroplasty (THA); Merete BioBall; FEMORAL OFFSET; CERAMIC HEADS; LEG LENGTH; ARTHROPLASTY; RESTORATION; MULTICENTER; FRACTURE;
D O I
10.21037/aoj-23-22
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and Objective: Modular neck adapters allow different length and offset changes to reach a stable total hip arthroplasty (THA) and permit a quick partial hip revision procedure without removing the existing components. The literature is poor on this matter and about the long-term related outcomes. This narrative review summarizes the most recent literature about these devices as an option of surgical treatment in partial total hip arthroplasty revision (THAr) focusing on indications, clinical and radiological outcomes, and related complications. Methods: The narrative review of the current available literature was conducted in December 2022 through electronic database. The terms used were: "Head neck taper" OR "Merete BioBall" AND "revision Total Hip Arthroplasty (MeSH Terms)". The timeframe was limited between 01/01/2000 and 01/12/2022. The studies regarding the clinical use of the Merete BioBall (R) system in hip revision surgery were included, while all the papers concerning modular stem prosthesis were excluded. Key Content and Findings: The surgical procedure is safe, quick and allows the surgeon to correct a well-fixed stem version, length and offset, besides retensioning soft tissues. Clinical and radiological outcomes are good with low complications rates. Conclusions: The modular neck adapter system seems to be a good surgical procedure for recurrent dislocation of THA, especially in case of a second THAr surgery. However, the main indication of adapter use remains the isolated acetabular cup revision. The related complications are rare: the worst is the redislocation due to an insufficient stem version and length correction. Re-dislocation rates reported in literature vary from 5.2% to 15%. Corrosion or fretting of the modular system are not reported in literature.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Evaluation of the Painful Dual Taper Modular Neck Stem Total Hip Arthroplasty: Do They All Require Revision?
    Kwon, Young-Min
    JOURNAL OF ARTHROPLASTY, 2016, 31 (07): : 1385 - 1389
  • [32] Contribution of the BioballTM head–neck adapter to the restoration of femoral offset in hip revision arthroplasty with retention of a well-fixed cup and stem
    Clara Chimeno
    Jenaro Ángel Fernández-Valencia
    Alfonso Alías
    Adrià Serra
    Yury Postnikov
    Andrés Combalia
    Ernesto Muñoz-Mahamud
    International Orthopaedics, 2023, 47 : 2245 - 2251
  • [33] Significantly Reduced Leg Length Discrepancy and Increased Femoral Offset by Application of a Head-Neck Adapter in Revision Total Hip Arthroplasty
    Woelfle, Julia V.
    Fraitzl, Christian R.
    Reichel, Heiko
    Wernerus, Dirk
    JOURNAL OF ARTHROPLASTY, 2014, 29 (06): : 1301 - 1307
  • [34] Survivorship of a Metal-on-Metal Total Hip Implant With Modular Titanium Adapter
    Semaan, Derek J.
    Rutledge-Jukes, Heath
    Berend, Keith R.
    Lombardi, Adolph V., Jr.
    Adams, Joanne B.
    Crawford, David A.
    JOURNAL OF ARTHROPLASTY, 2022, 37 (07): : S560 - S565
  • [35] Correlation of Corrosion and Biomechanics in the Retrieval of a Single Modular Neck Total Hip Arthroplasty Design Modular Neck Total Hip Arthroplasty System
    Lanting, Brent A.
    Teeter, Matthew G.
    Vasarhelyi, Edward M.
    Ivanov, Todor G.
    Howard, James L.
    Naudie, Douglas D. R.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (01): : 135 - 140
  • [36] Revision arthroplasty of the hip - Modularity of neck and metaphyseal components
    Koester, G.
    Walde, T. A.
    ORTHOPADE, 2009, 38 (03): : 238 - +
  • [37] Contribution of the Bioball™ head-neck adapter to the restoration of femoral offset in hip revision arthroplasty with retention of a well-fixed cup and stem
    Chimeno, Clara
    Fernandez-Valencia, Jenaro angel
    Alias, Alfonso
    Serra, Adria
    Postnikov, Yury
    Combalia, Andres
    Munoz-Mahamud, Ernesto
    INTERNATIONAL ORTHOPAEDICS, 2023, 47 (09) : 2245 - 2251
  • [38] Revision hip arthroplasty using a cementless modular tapered stem
    Pattyn, Christophe
    Mulliez, Alexander
    Verdonk, Rene
    Audenaert, Emmanuel
    INTERNATIONAL ORTHOPAEDICS, 2012, 36 (01) : 35 - 41
  • [39] Modular neck total hip arthroplasty - a perfect storm
    Humadi, Ali
    Jamieson, Richard
    Chao, Tat
    Tacey, Mark
    Khoury, Elie
    HIP INTERNATIONAL, 2016, 26 (02) : 128 - 131
  • [40] Factors influencing taper failure of modular revision hip stems
    Krull, A.
    Morlock, M. M.
    Bishop, N. E.
    MEDICAL ENGINEERING & PHYSICS, 2018, 54 : 65 - 73