Commissioning and clinical evaluation of the IDENTIFYTM surface imaging system for frameless stereotactic radiosurgery

被引:0
|
作者
Covington, Elizabeth L. [1 ,2 ]
Stanley, Dennis N. [1 ]
Sullivan, Rodney J. [1 ]
Riley, Kristen O. [3 ]
Fiveash, John B. [1 ]
Popple, Richard A. [1 ,4 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, Birmingham, AL USA
[2] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI USA
[3] Univ Alabama Birmingham, Dept Neurosurg, Birmingham, AL USA
[4] Univ Alabama Birmingham, Dept Radiat Oncol, 1719 6th Ave S, Birmingham, AL 35294 USA
来源
关键词
optical surface imaging; SGRT; SRS; stereotactic radiosurgery; surface guided radiation therapy;
D O I
10.1002/acm2.14058
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To commission and assess the clinical performance of a new commercial surface imaging (SI) system by analyzing intra-fraction motion from the initial cohort of patients treated with frameless stereotactic radiosurgery (fSRS). Methods: The IDENTIFYTM SI system was commissioned for clinical use on an Edge (VarianMedical Systems, Palo Alto, CA) linear accelerator. All patients who received intracranial radiotherapy with HyperArcTM (Varian Medical Systems, Palo Alto, CA) were immobilized with the EncompassTM (Qfix, Avondale, PA) thermoplastic mask and monitored for intra-fraction motion with SI. IDENTIFYTM log files were correlated with trajectory log files to correlate treatment parameters with SI-reported offsets. IDENTIFYTM reported offsets were correlated with gantry and couch angles to assess system performance for obstructed and clear camera field of view. Data were stratified by race to evaluate performance differences due to skin tone. Results: All commissioning data were found to meet recommended tolerances. IDENTIFYTM was used to monitor intra-fraction motion on 1164 fractions from 386 patients. The median magnitude of translational SI reported offsets at the end of treatment was 0.27 mm. SI reported offsets were shown to increase when camera pods are blocked by the gantry with larger increases seen at nonzero couch angles. With camera obstruction, the median magnitude of the SI reported offset was 0.50 and 0.80 mm forWhite and Black patients, respectively. Conclusions: IDENTIFYTM performance during fSRS is comparable to other commercially available SI systems where offsets are shown to increase at nonzero couch angles and during camera pod blockage.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Frameless, Real Time, Surface Imaging Guided Radiosurgery: Clinical Outcomes for Brain Metastases
    Pham, N. L.
    Reddy, P. V.
    Murphy, J. D.
    Sanghvi, P.
    Hattangadi, J. A.
    Kim, G. Y.
    Cervino, L. I.
    Pawlicki, T.
    Murphy, K. T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E105 - E105
  • [22] Commissioning Measurements for Stereotactic Radiosurgery (SRS) on Varian Halcyon System
    Grelewicz, Z. R.
    MEDICAL PHYSICS, 2024, 51 (10) : 7810 - 7811
  • [23] The Analysis of Positioning Accuracy for Frameless Noncoplanar Stereotactic Radiosurgery Treatment Using Optical Surface Monitoring System
    Wang, S.
    Qian, X.
    Ma, Y.
    MEDICAL PHYSICS, 2019, 46 (06) : E166 - E166
  • [24] CyberKnife frameless stereotactic radiosurgery for spinal lesions: Clinical experience in 125 cases
    Gerszten, PC
    Ozhasoglu, C
    Burton, SA
    Vogel, WJ
    Atkins, BA
    Kalnicki, S
    Welch, WC
    NEUROSURGERY, 2004, 55 (01) : 89 - 98
  • [25] Single-institution evaluation of intrafraction motion during frameless stereotactic radiosurgery
    Kowalchik, Kristin
    Kim, Siyong
    Serago, Christopher
    Smith, Ashley
    Ko, Stephen
    JOURNAL OF CLINICAL ONCOLOGY, 2012, 30 (34)
  • [26] Commissioning a New Frameless Intracranial Stereotactic Positioning System (ISPS): The Real-Time 3D Stereo-Vision Surface Imaging System
    Peng, L.
    Liu, C.
    Kahler, D.
    Samant, S.
    Li, J.
    Palta, J.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [27] Frameless stereotactic radiosurgery for brain metastases: a review of outcomes and prognostic scores evaluation
    Mok, S. T.
    Kam, Michael K. M.
    Tsang, W. K.
    Poon, Darren M. C.
    Loong, Herbert H.
    Yeung, W. M.
    Yeung, T. Y.
    Yu, Jimmy
    Wong, Carlos K. H.
    HONG KONG MEDICAL JOURNAL, 2017, 23 (06) : 599 - 608
  • [28] The feasibility of frameless stereotactic radiosurgery in the management of pediatric central nervous system tumors
    Nanda, Ronica
    Dhabbaan, Anees
    Janss, Anna
    Shu, Hui-Kuo
    Esiashvili, Natia
    JOURNAL OF NEURO-ONCOLOGY, 2014, 117 (02) : 329 - 335
  • [29] The Role of Frameless Stereotactic Radiosurgery in the Management of Pediatric Central Nervous System Tumors
    Hazariwala, R.
    Dhabbaan, A.
    Ogunleye, T.
    Janss, A.
    Shu, H.
    Esiashvilli, N.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S635 - S635
  • [30] Frameless immobilization system with roll correction for stereotactic radiosurgery of intracranial brain metastases
    Park, So-Yeon
    Choi, Noorie
    Jang, Na Young
    JOURNAL OF RADIATION RESEARCH, 2021, 62 (06) : 1015 - 1021