Preparation of polymer gel dosimeters for low gamma irradiation dose

被引:2
|
作者
El-Kelany, Moushera A. [2 ]
Khozemy, Ehab E. [1 ]
Goda, Heba [3 ]
Awad, Boshra [3 ]
机构
[1] Egyptian Atom Energy Author, Natl Ctr Radiat Res & Technol, Polymer Chem Dept, Cairo, Egypt
[2] Egyptian Atom Energy Author, Natl Ctr Radiat Res & Technol, Radiat Protect & Dosimetry Dept, Cairo, Egypt
[3] Ain Shams Univ, Fac Women Arts Sci & Educ, Chem Dept, Cairo, Egypt
关键词
CMC; gelatin; hydrogels; irradiation dosimeter; PVA; CONGO RED-DYE; CROSS-LINKING; RADIATION; GELATIN; DEGRADATION; CARBOXYMETHYLCELLULOSE; HYDROGELS;
D O I
10.1515/ract-2022-0106
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this study, two separate hydrogel dosimeters were prepared for low-dose measurement. The first system is based on Congo Red (CR) dye and a blend of natural polymers (gelatin/CMC) while the second system is based on 2,6-dichlorophenol indophenol (DCP) blue dye and a hybrid polymer blend (PVA/gelatin/CMC). The sensitivity of Congo red (CR)-(gelatin/CMC) gel dosimeter and 2, 6 - di choro phenol indophenols(DCP)-(PVA/gelatin/CMC) gel dosimetry systems were examined by studying the comparative results between the two different systems through the change in optical properties upon exposure to different gamma radiation doses, which showed the sensitivity of the second system (DCP)-(PVA/gelatin/CMC) to radiation is more than the first system(CR)-(gelatin/CMC) (CR) and (DCP)dyes have absorbance crest at 492 and 622 nm respectively. With the increase of the radiation dose, the decomposition of the dyes increases, and thus the color of the prepared hydrogel changes from orange to colorless in the case of (CR) dye at the dose (0-3.5 kGy) and from blue to colorless in the case (DCP) dye at the absorbed doses ranged from (0-1000 Gy). Dose-response curves, sensitivity, FTIR analysis, and the relative stability of the prepared hydrogel before and after irradiation were studied.
引用
收藏
页码:195 / 202
页数:8
相关论文
共 50 条
  • [31] The dose response of normoxic polymer gel dosimeters measured using X-ray CT
    Hill, B
    Venning, A
    Baldock, C
    BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (931): : 623 - 630
  • [32] Effect of inorganic salts and matrix crosslinking on the dose response of polymer gel dosimeters based on acrylamide
    Chacon, David
    Strumia, Miriam
    Valente, Mauro
    Mattea, Facundo
    RADIATION MEASUREMENTS, 2018, 117 : 7 - 18
  • [33] Severe dose inaccuracies caused by an oxygen-antioxidant imbalance in normoxic polymer gel dosimeters
    Sedaghat, Mahbod
    Bujold, Rachel
    Lepage, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (03): : 601 - 625
  • [34] Low dose gamma irradiation effect on aluminosilicate compounds
    Vanea, E.
    Simon, V.
    JOURNAL OF OPTOELECTRONICS AND ADVANCED MATERIALS, 2010, 12 (02): : 365 - 369
  • [35] Enhancement in dose sensitivity of polymer gel dosimeters composed of radiation-crosslinked gel matrix and less toxic monomers
    Hiroki, A.
    Yamashita, S.
    Taguchi, M.
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [36] Novel liquid dosimeters for low-energy electron beam irradiation in low and medium dose ranges
    Besecke, Joana K.
    Ullrich, Daniel
    von Hauff, Elizabeth
    Teichmann, Tobias
    Dietze, Marleen
    Koenig, Ulla
    Schopf, Simone
    RADIATION PHYSICS AND CHEMISTRY, 2024, 222
  • [37] LOW-DOSE GAMMA-IRRADIATION OF SOME FLUOROPOLYMERS - EFFECT OF POLYMER CHEMICAL-STRUCTURE
    ROSENBERG, Y
    SIEGMANN, A
    NARKIS, M
    SHKOLNIK, S
    JOURNAL OF APPLIED POLYMER SCIENCE, 1992, 45 (05) : 783 - 795
  • [38] LOW DOSE MEASUREMENTS WITH THERMOLUMINESCENT DOSIMETERS
    SHAMBON, A
    MCLAUGHL.JE
    HEALTH PHYSICS, 1968, 15 (02): : 177 - +
  • [39] Effect of low irradiation temperature on the gamma-ray response of dyed and undyed PMMA dosimeters
    Biramontri, S
    Haneda, N
    Tachibana, H
    Kojima, T
    RADIATION PHYSICS AND CHEMISTRY, 1996, 48 (01): : 105 - 109
  • [40] Polymer Gel Dosimeters for use in Targeted Radionuclide Therapy
    Gear, J. I.
    Flux, G. D.
    Partridge, M.
    Ott, R. J.
    Charles-Edwards, E.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S207 - S207