Impact of Source Conditions on Collision Cross Section Determination by Trapped Ion Mobility Spectrometry

被引:0
|
作者
George, Anais C. [1 ]
Schmitz, Isabelle [1 ]
Colsch, Benoit [2 ]
Afonso, Carlos [1 ]
Fenaille, Francois [2 ]
Loutelier-Bourhis, Corinne [1 ]
机构
[1] Normandie Univ, Univ Rouen Normandie, INSA Rouen Normandie, CNRS,COBRAUMR 6014,INC3M FR 3038, F-76000 Rouen, France
[2] Univ Paris Saclay, CEA, INRAE, Dept Med & Technol Sante DMTS,MetaboHUB, F-91191 Gif sur Yvette, France
关键词
Lipids; Solvent-ion cluster; Collisioncross section; Ion mobility-mass spectrometry; Trappedion mobility spectrometry; STRUCTURAL-CHARACTERIZATION; GAS MODIFIERS; METABOLOMICS; VALUES;
D O I
10.1021/jasms.3c00361
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Collision cross section (CCS) values determined in ion mobility-mass spectrometry (IM-MS) are increasingly employed as additional descriptors in metabolomics studies. CCS values must therefore be reproducible and the causes of deviations must be carefully known and controlled. Here, we analyzed lipid standards by trapped ion mobility spectrometry-mass spectrometry (TIMS-MS) to evaluate the effects of solvent and flow rate in flow injection analysis (FIA), as well as electrospray source parameters including nebulizer gas pressure, drying gas flow rate, and temperature, on the ion mobility and CCS values. The stability of ion mobility experiments was studied over 10 h, which established the need for a delay-time of 20 min to stabilize source parameters (mostly pressure and temperature). Modifications of electrospray source parameters induced shifts of ion mobility peaks and even the occurrence of an additional peak in the ion mobility spectra. This behavior could be essentially explained by ion-solvent cluster formation. Changes in source parameters were also found to impact CCS value measurements, resulting in deviations up to 0.8%. However, internal calibration with the Tune Mix calibrant reduced the CCS deviations to 0.1%. Thus, optimization of source parameters is essential to achieve a good desolvation of lipid ions and avoid misinterpretation of peaks in ion mobility spectra due to solvent effects. This work highlights the importance of internal calibration to ensure interoperable CCS values, usable in metabolomics annotation. [GRAPHICS]
引用
收藏
页码:696 / 704
页数:9
相关论文
共 50 条
  • [31] Ion Mobility-Mass Spectrometry-Based Measurements of Collision Cross Section Values for Metabolites and Related Databases
    Chen X.
    Li T.-Z.
    Zhu Z.-J.
    Journal of Chinese Mass Spectrometry Society, 2022, 43 (05) : 596 - 610
  • [32] Large-Scale Prediction of Collision Cross-Section Values for Metabolites in Ion Mobility-Mass Spectrometry
    Zhou, Zhiwei
    Shen, Xiaotao
    Tu, Jia
    Zhu, Zheng-Jiang
    ANALYTICAL CHEMISTRY, 2016, 88 (22) : 11084 - 11091
  • [33] Evaluation of Collision Cross Section Calibrants for Structural Analysis of Lipids by Traveling Wave Ion Mobility-Mass Spectrometry
    Hines, Kelly M.
    May, Jody C.
    McLean, John A.
    Xu, Libin
    ANALYTICAL CHEMISTRY, 2016, 88 (14) : 7329 - 7336
  • [34] Untargeted Molecular Discovery in Primary Metabolism: Collision Cross Section as a Molecular Descriptor in Ion Mobility-Mass Spectrometry
    Nichols, Charles M.
    Dodds, James N.
    Rose, Bailey S.
    Picache, Jaqueline A.
    Morris, Caleb B.
    Codreanu, Simona G.
    May, Jody C.
    Sherrod, Stacy D.
    McLean, John A.
    ANALYTICAL CHEMISTRY, 2018, 90 (24) : 14484 - 14492
  • [35] Correlating Resolving Power, Resolution, and Collision Cross Section: Unifying Cross-Platform Assessment of Separation Efficiency in Ion Mobility Spectrometry
    Dodds, James N.
    May, Jody C.
    McLean, John A.
    ANALYTICAL CHEMISTRY, 2017, 89 (22) : 12176 - 12184
  • [36] Trapped ion mobility spectrometry: A short reviewMark
    Ridgeway, Mark E.
    Lubeck, Markus
    Jordens, Jan
    Mann, Mattias
    Park, Melvin A.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2018, 425 : 22 - 35
  • [38] Note: Integration of trapped ion mobility spectrometry with mass spectrometry
    Fernandez-Lima, F. A.
    Kaplan, D. A.
    Park, M. A.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (12):
  • [39] Collision Cross-Section Determination and Tandem Mass Spectrometric Analysis of Isomeric Carotenoids Using Electrospray Ion Mobility Time-of-Flight Mass Spectrometry
    Dong, Linlin
    Shion, Henry
    Davis, Roderick G.
    Terry-Penak, Brent
    Castro-Perez, Jose
    van Breemen, Richard B.
    ANALYTICAL CHEMISTRY, 2010, 82 (21) : 9014 - 9021
  • [40] Review of Ion Collision Cross Section Measurement Methods Based on Ion Trap Mass Spectrometry
    Xiang Y.
    Jiang T.
    Xu W.
    Journal of Chinese Mass Spectrometry Society, 2022, 43 (05) : 611 - 622