Design limitations of using magnetic inductive coupling to verify nasogastric tube placement

被引:3
|
作者
Silva, Nuno P. [1 ]
Elahi, Adnan [1 ]
O'Halloran, Martin [1 ]
O'Keeffe, Derek T. [2 ]
O'Loughlin, Declan [3 ,4 ]
机构
[1] Univ Galway, Translat Med Device Lab, Galway, Ireland
[2] Univ Galway, Dept Med, Div Endocrinol, Galway, Ireland
[3] Trinity Coll Dublin, Elect & Elect Engn, Dublin, Ireland
[4] Trinity Coll Dublin, Trinity Ctr Biomed Engn, Dublin, Ireland
基金
欧洲研究理事会;
关键词
electromagnetic coupling; electromagnetic induction; inductive power transmission; medical devices and systems; mutual coupling;
D O I
10.23919/EuCAP57121.2023.10133045
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Nasogastric tube insertion is a standard procedure but with potentially fatal consequences if misplaced. Many bedside methods have been proposed for identifying misplacement including magnetic inductive coupling. This study aims to identify the coil design limitations for nasogastric tube placement confirmation using magnetic inductive coupling. The fundamental limits on the coil sizes and detection distances are identified based on the anatomical understanding. Candidate designs are simulated and fabricated, and key parameters such as coil radius and penetration depth are examined. Specifically, the mutual inductance and coupling factor between the coils in the candidate designs are both calculated and measured with respected to distance. This work highlights that a small coil of up to 3 mm in diameter could be integrated into a nasogastric tube without impairing clinical function. These preliminary results suggest that magnetic inductive coupling is a promising approach for nasogastric tube localisation, however, optimisation of the detector coil size could improve localisation performance.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] An Inductive-Coupling Bus with Collision Detection Scheme Using Magnetic Field Variation for 3-D Network-on-Chips
    Kadomoto, Junichiro
    Miyata, Tomoki
    Amano, Hideharu
    Kuroda, Tadahiro
    2016 IEEE ASIAN SOLID-STATE CIRCUITS CONFERENCE (A-SSCC), 2016, : 41 - 44
  • [42] Increasing the Safety of Blind Gastric Tube Placement in Pediatric Patients: The Design and Testing of a Procedure Using a Carbon Dioxide Detection Device
    Gilbert, Rebecca T.
    Burns, Suzanne M.
    JOURNAL OF PEDIATRIC NURSING-NURSING CARE OF CHILDREN & FAMILIES, 2012, 27 (05): : 528 - 532
  • [43] Techniques for dual-band LNA design using cascode switching and inductor magnetic coupling
    Martins, Miguel A.
    Fernandes, Jorge R.
    Silva, Manuel M.
    2007 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOLS 1-11, 2007, : 1449 - 1452
  • [44] DESIGN OF WIRELESS POWER TRANSFER SYSTEMS USING MAGNETIC RESONANCE COUPLING FOR IMPLANTABLE MEDICAL DEVICES
    Jolani, Farid
    Mehta, Jeetkumar
    Yu, Yiqiang
    Chen, Zhizhang
    PROGRESS IN ELECTROMAGNETICS RESEARCH LETTERS, 2013, 40 (40): : 141 - 151
  • [45] Optimal Design of Magnetorheological Valve Using the Coupling of FE Magnetic Analysis and MOGA Optimization for Prosthetic Ankle
    Kumar, Sachin
    Chandramohan, Sujatha
    Sujatha, S.
    JOURNAL OF VIBRATION ENGINEERING & TECHNOLOGIES, 2024, 12 (02) : 2205 - 2217
  • [46] Optimal Design of Magnetorheological Valve Using the Coupling of FE Magnetic Analysis and MOGA Optimization for Prosthetic Ankle
    Sachin Kumar
    Sujatha Chandramohan
    S. Sujatha
    Journal of Vibration Engineering & Technologies, 2024, 12 : 2205 - 2217
  • [47] Study on Efficiency Maximization Design Principles for Wireless Power Transfer System Using Magnetic Resonant Coupling
    Li, Hongchang
    Yang, Xu
    Wang, Kangping
    Dong, Xiaoshuai
    2013 IEEE ECCE ASIA DOWNUNDER (ECCE ASIA), 2013, : 888 - 892
  • [48] Coupling modeling and bandpass filter design using electric and magnetic couplings on CMOS 65 nm technology
    Mou, Jin-Chao
    Wu, Di-Si
    Li, Yuan Chun
    MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2020, 62 (07) : 2458 - 2462
  • [49] Design of a 24-GHz VCO Using Magnetic Coupling of Gate-Drain Through a Transformer
    Lee, Milim
    Lee, Changhyun
    Park, Changkun
    2015 ASIA-PACIFIC MICROWAVE CONFERENCE (APMC), VOLS 1-3, 2015,
  • [50] Evaluation and Optimization of Operating Parameters Affecting the Davis Tube Magnetic Tester by Using 2nd Factorial Design
    Haffez, Carnal S. Abdel
    Abu-Ali, Mohamed Hussein
    MATERIALS TESTING, 2012, 54 (06) : 437 - 443