Design limitations of using magnetic inductive coupling to verify nasogastric tube placement

被引:3
|
作者
Silva, Nuno P. [1 ]
Elahi, Adnan [1 ]
O'Halloran, Martin [1 ]
O'Keeffe, Derek T. [2 ]
O'Loughlin, Declan [3 ,4 ]
机构
[1] Univ Galway, Translat Med Device Lab, Galway, Ireland
[2] Univ Galway, Dept Med, Div Endocrinol, Galway, Ireland
[3] Trinity Coll Dublin, Elect & Elect Engn, Dublin, Ireland
[4] Trinity Coll Dublin, Trinity Ctr Biomed Engn, Dublin, Ireland
基金
欧洲研究理事会;
关键词
electromagnetic coupling; electromagnetic induction; inductive power transmission; medical devices and systems; mutual coupling;
D O I
10.23919/EuCAP57121.2023.10133045
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Nasogastric tube insertion is a standard procedure but with potentially fatal consequences if misplaced. Many bedside methods have been proposed for identifying misplacement including magnetic inductive coupling. This study aims to identify the coil design limitations for nasogastric tube placement confirmation using magnetic inductive coupling. The fundamental limits on the coil sizes and detection distances are identified based on the anatomical understanding. Candidate designs are simulated and fabricated, and key parameters such as coil radius and penetration depth are examined. Specifically, the mutual inductance and coupling factor between the coils in the candidate designs are both calculated and measured with respected to distance. This work highlights that a small coil of up to 3 mm in diameter could be integrated into a nasogastric tube without impairing clinical function. These preliminary results suggest that magnetic inductive coupling is a promising approach for nasogastric tube localisation, however, optimisation of the detector coil size could improve localisation performance.
引用
收藏
页数:5
相关论文
共 50 条
  • [11] Noninvasive verification of nasogastric tube placement using a magnet-tracking system: A pilot study in healthy subjects
    Bercik, Premysl
    Schlageter, Vincent
    Mauro, Marina
    Rawlinson, John
    Kucera, Pavel
    Armstrong, David
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2005, 29 (04) : 305 - 310
  • [12] Conceptual design of inductive magnetic sensors using photolithography processes
    Testa, Duccio
    Cemes, Marcus
    Decaux, Eugenie
    Lavilla, Camille
    Tanguy, Florian
    Amor, Mehdi
    FUSION ENGINEERING AND DESIGN, 2025, 214
  • [13] Design of the inductive sensor and the intelligent coating pachometer using magnetic method
    Yao, Guangren
    Yibiao Jishu Yu Chuanganqi/Instrument Technique and Sensor, 1999, (11): : 19 - 21
  • [14] Diagnostic accuracy of using capnography in verification of nasogastric tube placement among adult patients in hospital settings: Protocol of a diagnostic study
    Chau, Janita Pak Chun
    Tong, Danny Wah Kun
    Lo, Suzanne Hoi Shan
    Sze, Savina Yee Man
    Kwok, Molly Lai Mei
    Lai, Peter Chi Keung
    Lam, Harris Kai Cheong
    Chung, Josephine Yuen Man
    Liu, Xu
    Chien, Wai Tong
    Choi, Kai Chow
    PLOS ONE, 2023, 18 (10):
  • [15] Nasogastric feeding tube placement using an ultrathin transnasal endoscopy in patients with esophageal cancer: Preliminary report of a one year experience
    Lin, CH
    Liu, NJ
    Lee, CS
    Tang, JH
    Wei, KL
    Chun, Y
    Sung, KF
    Cheng, CL
    Chin, CT
    Chen, PC
    GASTROINTESTINAL ENDOSCOPY, 2003, 57 (05) : AB145 - AB145
  • [16] Real-time image-guided nasogastric feeding tube placement: A case series using Kangaroo with IRIS Technology in an ICU
    Mizzi, Anna
    Cozzi, Silvano
    Beretta, Luigi
    Greco, Massimiliano
    Braga, Marco
    NUTRITION, 2017, 37 : 48 - 52
  • [17] Comparing insertion characteristics on nasogastric tube placement by using GlideScope™ visualization vs. MacIntosh laryngoscope assistance in anaesthetized and intubated patients
    Ibadullah, Wan Hafsah Wan
    Yahya, Nurlia
    Ghazali, Siti Salmah
    Kamaruzaman, Esa
    Yong, Liu Chian
    Dan, Adnan
    Zain, Jaafar Md
    REVISTA BRASILEIRA DE ANESTESIOLOGIA, 2016, 66 (04): : 363 - 368
  • [18] Limitations of Routine Verification of Nasogastric Tube Insertion Using X-Ray and Auscultation: Two Case Reports of Life-Threatening Complications
    Nejo, Takahide
    Oya, Soichi
    Tsukasa, Tsuchiya
    Yamaguchi, Naomi
    Matsui, Toru
    NUTRITION IN CLINICAL PRACTICE, 2016, 31 (06) : 780 - 784
  • [19] Design Guidelines of Inductive Coils Using a Polymer Bonded Magnetic Composite for Inductive Power Transfer Systems in Electric Vehicles
    Delgado, Alberto
    Schoenberger, Derek
    Oliver, Jesus Angel
    Alou, Pedro
    Cobos, Jose Antonio
    IEEE TRANSACTIONS ON POWER ELECTRONICS, 2020, 35 (08) : 7884 - 7893
  • [20] Design and experimental verification of CMOS magnetic-based microbead detection using an asynchronous intra-chip inductive-coupling transceiver
    Niitsu, Kiichi
    Kobayashi, Atsuki
    Yoshida, Kohei
    Nakazato, Kazuo
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (01)