The development of a clearance level radioactive waste measurement system using an array of NaI(Tl) detectors

被引:1
|
作者
Li, C. G. [1 ]
Gao, F. [1 ]
Sun, H. M. [1 ]
Zhao, D. S. [1 ]
Su, M. [1 ]
Bai, H. Y. [1 ]
Xiong, Z. H. [1 ]
Mo, Z. H. [1 ]
机构
[1] China Acad Engn Phys, Inst Mat, Mianyang 621908, Peoples R China
基金
中国国家自然科学基金;
关键词
Clearance level radioactive waste; NaI(Tl) detector array; Detection efficiency; Minimum detectable activity; EFFICIENCY;
D O I
10.1016/j.nima.2023.168748
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In order to meet the needs of rapid classification and measurement of the radioactive waste, a clearance level radioactive waste measurement system has been successfully developed and applied, achieving the goal of waste minimization. The measurement system consisted of 34 NaI(Tl) detectors taking into account the detector layout, detection efficiency and shielding chamber (to reduce the background of the environment). The NaI(Tl) detector array was built by a geometric condition of spatially symmetric to measure gamma rays. The shielding chamber composed of 2 mm copper and 10 mm lead is used as an passive shielding, and the radioactive waste is placed inside the shielding chamber, which suppress the gamma rays from the decay of 40K, 208Tl and other nuclides in the environment. A method to determine the counting of energy interval for the gamma rays is proposed to obtain the radioactivity of the radioactive waste package. The performance testing results of the measurement system show that the energy resolution and efficiency for 661.7 keV (137Cs) are 7.84% and 69.49% respectively, and the minimum detectable activity of 137Cs is 24.52 Bq. All these indicate that the indices of the clearance level radioactive waste measurement system have reached the international recommended level.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Development of a radioactive waste measurement system using arrayed NaI(Tl) detectors
    Yang, Guang
    Shi, Rui
    Wang, Zhou
    Tuo, Xianguo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2024, 85 (11) : 883 - 889
  • [2] New analytical approach to calibrate the NaI (Tl) detectors using spherical radioactive sources
    El-Khatib, Ahmed M.
    Gouda, Mona M.
    Badawi, Mohamed S.
    Nafee, Sherif S.
    El-Mallah, Ekram A.
    RADIATION PROTECTION DOSIMETRY, 2013, 156 (01) : 109 - 117
  • [3] Development of pixelated NaI(Tl) detectors for PET
    Surti, S
    Karp, JS
    Muehllehner, G
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1093 - 1098
  • [4] Validation of measurement comparability of NaI(Tl) scintillation detectors for radioactive cesium in brown rice sample by interlaboratory comparison
    Tsutomu Miura
    Mayumi Hachinohe
    Akira Yunoki
    Shioka Hamamatsu
    Yasuhiro Unno
    Journal of Radioanalytical and Nuclear Chemistry, 2020, 326 : 1225 - 1231
  • [5] Validation of measurement comparability of NaI(Tl) scintillation detectors for radioactive cesium in brown rice sample by interlaboratory comparison
    Miura, Tsutomu
    Hachinohe, Mayumi
    Yunoki, Akira
    Hamamatsu, Shioka
    Unno, Yasuhiro
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2020, 326 (02) : 1225 - 1231
  • [6] A Bayesian Approach for Remote Depth Estimation of Buried Low-Level Radioactive Waste with a NaI(Tl) Detector
    Kim, Jinhwan
    Lim, Kyung Taek
    Park, Kyeongjin
    Cho, Gyuseong
    SENSORS, 2019, 19 (24)
  • [7] Gamma-Ray Source Positioning Using Array NaI(Tl) Detectors in the Radiation Portal Monitors
    Qibiao Wang
    Chao Deng
    Jianbo Yang
    Rui Shi
    Songbai Zhang
    Xianguo Tuo
    Journal of the Korean Physical Society, 2019, 75 : 196 - 201
  • [8] Gamma-Ray Source Positioning Using Array NaI(Tl) Detectors in the Radiation Portal Monitors
    Wang, Qibiao
    Deng, Chao
    Yang, Jianbo
    Shi, Rui
    Zhang, Songbai
    Tuo, Xianguo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2019, 75 (03) : 196 - 201
  • [9] MEASUREMENT OF RADIOACTIVE-TRACER MICROSPHERE BLOOD-FLOW WITH NAI(TL)-WELL AND GE-WELL TYPE DETECTORS
    WINKLER, B
    STAMMLER, G
    SCHAPER, W
    BASIC RESEARCH IN CARDIOLOGY, 1982, 77 (03) : 292 - 300
  • [10] Development of a large NaI(Tl) detection system
    Tang, Mengjiao
    Zhang, Lianjun
    He, Gaokui
    Zhao, Jiangbin
    Liu, Yang
    Wang, Chao
    Que, Ziang
    JOURNAL OF INSTRUMENTATION, 2024, 19 (10):